Product catalog summary
Instructions for Use
This document provides detailed instructions for the use of the FORESIGHT 45 Fixed Back End dental handpiece, item code FS-45S. It is intended for licensed professionals only.

Warnings and Precautions
- Use only burs from reputable manufacturers with specified dimensions.
- Inspect burs for damage before use and ensure they are correctly seated.
- Do not operate the handpiece without a bur.
- Replace the turbine if the handpiece becomes noisy or sluggish.
- Avoid pressing the head cap button during rotation.

Specifications
- Model: FORESIGHT 45 Fixed Back End
- Operating Air Pressure: 35-40 psi
- Speed: 400,000 rpm Min.
- Weight: 69.8g
- Head Height: 13.6mm
- Head Diameter: 10.61mm
- Bur Type: Surgical Length
- Coupling Type: Fixed Back End
- LED Power: Self-Generated

Installation and Operation
- Follow the provided figure for installation.
- For changing burs, use the end cap to secure or remove the bur.
- Ensure the bur is fully inserted to prevent dislodgement.

Cartridge Replacement
- Use a cap wrench to remove the head cap and replace the cartridge.
- Ensure all components are in place before reassembly.

Cleaning and Lubrication
- Remove the handpiece from the coupling and clean with a soft cloth.
- Avoid using ultrasonic baths or immersing in water.
- Use a needle to clear debris from the spray hole.
- Lubricate and operate the handpiece to purge debris.

Sterilization
- Remove the bur and place the handpiece in a sterilization pouch.
- Steam autoclave at 132ºC for 15 minutes, followed by a 20-minute drying cycle.

Warranty
- The product is warranted for 1 year against manufacturing defects.
- The warranty does not cover damage from misuse or unauthorized alterations.

Repair and Replacement Parts
- For repairs, provide contact information and nature of repair.
- Replacement parts include cartridges, cap wrenches, head caps, and cleaners.

Contact Medidenta for repairs or parts at 3325 W Ali Baba Lane, Suite 603, Las Vegas, NV 89118, Tel: 800-221-0750.
See more

Catalog excerpts

Foresight 45 SGL-1

1. Installation Follow (Fig. 1) below to install the handpiece. Push on end cap with thumb to click, and remove the bur (Fig. 2) PUSH ON END CAP WITH THUMB Fig. 2 Bur mounting Push on end cap and insert the bur into the head until it stops (5mm deep approx.) release pressure on button bur should be secure. Firmly tug on the bur to insure that it is properly secured. Caution: Always insert the bur full length of the chuck. If you attempt to extend it from the chuck, the bur may dislodge during use and cause injury to you or your patients. Federal (US) law restricts the use of this device to licensed professionals. 1. Only use burs of reputable manufacturers with a shaft diameter of 1.59/1.60mm (.0625/.0629”), a total length of max. 26mm (1.02”), and head diameters of up to 2mm (.0787”). Burs which deviate from the stated measurements must not be used 2. Before each use, make sure that the bur is not bent or damaged. 3. Before each use, make sure that the bur is correctly seated and locked in the chuck. 4. Do not run the handpiece without a bur or bur blank locked in the chuck. 5. When handpiece becomes noisy or sluggish after extended length of use, replace with a new turbine. 6. Do not push the head cap button during rotation. 3. Cartridge Replacement Removing cartridge (Fig. 3&4) Mount a bur in the chuck. Turn the head cap counter-clockwise with the cap wrench (Fig. 3), and remove the head cap. Push on the bur and the cartridge is easily removed from the back of the head (Fig. 4). Clean inside the handpiece head with spray lubricant. Continue to hand-thread the Head-Cap until almost flush to the last thread. Then, pull and wiggle the chuck pin into the bottom of the Head to properly seat the cartridge. (Fig.7). further tighten the Head-Cap with Cap-Wrench 4. Cleaning & Lubrication 1. Remove the handpiece from the coupling. 2. Insert bur into the chuck of the handpiece and secure. 3. Remove any adherent material using only a soft cloth, soft brush, gauze, or cotton swab that has been immersed in a soapy solution. Do not use a wire brush. Wipe and dry thoroughly. Do not use ultrasonic bath, nor immerse the handpiece in water or any solution. 4. Use needle to remove any debris that may clog the spray hole (Fig. 8)

 Open the catalog to page 1
Foresight 45 SGL-2

5. For Fixed Back End Handpiece: Insert nozzle tip into air drive tube on the back end of the handpiece and spray for approximately 3 seconds. (Fig. 9) 6. Fully depress the button of the end cap and hold the collet open then plunge the bur in and out several times to distribute the cleaner into the chucking mechanism. Fully insert the bur, release the pressure on end cap button to secure the bur. Rotate the bur back and forth between your fingers to further distribute the cleaner & lubricant in the head of the handpiece to help loosen any debris. 7. Attach handpiece to an air supply to operate...

 Open the catalog to page 2

All Medidenta catalogs and technical brochures

  1. TT Care 100+

    2  Pages

  2. PRODUCT CATALOG

    82  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.