Foresight SGL Power Head

Foresight SGL Power Head

Foresight SGL Power Head

Product catalog summary
Introduction
This document provides instructions for the use of the Foresight Power Head Multiflex and Fixed Back End dental devices. It is intended for licensed professionals and should be kept for future reference.

Warnings
Usage is restricted to licensed professionals as per Federal (US) law.

Specifications
  • Operating Air Pressure: 35-40 psi
  • Speed: 330,000 rpm Min.
  • Weight: Multiflex - 93.0g, Fixed Back End - 71.0g
  • Head Height: 14.0mm
  • Head Diameter: 12.3mm
  • Bur Type: Standard Length
  • Coupling Type: Multiflex or Fixed Back End
  • LED Power: Self-Generated

Installation
Follow the provided figures for proper installation of the handpiece.

Changing Bur
  • To remove, push the end cap and remove the bur.
  • To mount, insert the bur fully into the head and ensure it is secure.

Cartridge Replacement
  • Remove the cartridge by turning the head cap counter-clockwise and pushing the bur.
  • Mount the cartridge by inserting it into the head cap and securing it with the cap wrench.

Cleaning & Lubrication
  • Remove the handpiece and clean with a soft cloth and soapy solution.
  • Use a needle to clear debris from the spray hole.
  • Lubricate using the Ispray can and Multiflex nozzle.
  • Operate the handpiece to purge debris and excess cleaner.

Sterilization
  • Remove the bur and wipe the surface.
  • Seal in a sterilization pouch and autoclave for 15 minutes at 132ºC.
  • Allow a minimum of 20 minutes for drying.

Warranty
The product is warranted against manufacturing defects for one year. The warranty covers repair or replacement but excludes damages from misuse or unauthorized alterations.

Repair
For repairs, provide contact information and nature of repair to Medidenta.

Replacement Parts
  • T-FSS - Foresight Power Head Cartridge
  • P-FSSCW - Foresight Power Cap Wrench
  • P-FSSHC - Foresight Power Head Cap
  • ISPRAY – Handpiece Cleaner / Lubricant
See more

Catalog excerpts

Foresight SGL Power Head-1

M MedidentaSGL By Medidenta INSTRUCTION FOR USE Read this manual before use. Keep it in a safe place for future reference. A WARNING Federal (US) law restricts the use of this device to licensed professionals. 1. Only use burs of reputable manufacturers with a shaft diameter of 1.59/1.60mm {.0625/.0629"), a total length of max. 26mm (1.02"), and head diameters of up to 2mm (.0787"). Burs which deviate from the stated measurements must not be used 2. Before each use, make sure that the bur is not bent or damaged. 3. Before each use, make sure that the bur is correctly seated and locked in the chuck. 4. Do not run the handpiece without a bur or bur blank locked in the chuck. 5. When handpiece becomes noisy or sluggish after extended length of use, replace with a new turbine. 6. Do not push the head cap button during rotation. 1. Installation Follow (Fig. 1 & Fig. 2) below to install the handpiece. HANDPIECE MULTIPLEX COUPLER AIR SUPPLY HANDPIECE AIR SUPPLY Fig. 2 2. Changing Bur Bur removal Push on end cap with thumb to click, and remove the bur (Fig. 3) Push on end cap and insert the bur into the head until it stops (5mm deep approx.) release pressure on button bur should be secure. Firmly tug on the bur to insure that it is properly secured. Caution: Always insert the bur full length of the chuck. If you attempt to extend it from the chuck, the bur may dislodge during use and cause injury to you or your patients. 3. Cartridge Replacement Removing cartridge (Fig. 4&5) Mount a bur in the chuck. Turn the head cap counter-clockwise with the cap wrench (Fig. 4), and remove the head cap. Push on the bur and the cartridge is easily removed from the back of the head (Fig. 5). Clean inside the handpiece head with spray lubricant. Mounting Cartridge (Fig. 6&7) Check for all components inside head (Fig. 6) Gently insert the rear end of Cartridge into the Head-Cap, then hold the Cartridge by the Head-Cap, bring Head of handpiece down over the top of the chuck pin and thread the Head-Cap into the Head of handpiece, as shown on (Fig. 7) 4. Cleaning & Lubrication 1. Remove the handpiece from the coupling. 2. Insert bur into the chuck of the handpiece and secure.

 Open the catalog to page 1
Foresight SGL Power Head-2

3. Remove any adherent material using only a soft cloth, soft brush, gauze, or cotton swab that has been immersed in a soapy solution. Do not use a wire brush. Wipe and dry thoroughly. Do not use ultrasonic bath, nor immerse the handpiece in water or any solution. 4. Use needle to remove any debris that may clog the spray hole (Fig. 8) Fig. 8 5. Remove bur, insert the Multiflex nozzle tip into Ispray lubrication can (Fig. 9) then insert back end of handpiece into the Multiflex nozzle (Fig. 10) and spray for approximately 3 seconds. 6. Remove Multiflex nozzle from Ispray can and install nozzle...

 Open the catalog to page 2

All Medidenta catalogs and technical brochures

  1. TT Care 100+

    2  Pages

  2. PRODUCT CATALOG

    82  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.