1. Catalogs
  2. Medima
  3. Family of syringe infusion pumps

Family of syringe infusion pumps

Family of syringe infusion pumps

Family of syringe infusion pumps

Product catalog summary
Flow Rate Specifications: The syringe infusion pumps support flow rates ranging from 0.1 ml/h to 2000 ml/h depending on the syringe size, with increments varying based on the flow rate range. For example, increments are 0.01 ml/h up to 99.99 ml/h, 0.1 ml/h from 100 ml/h to 999.9 ml/h, and 1 ml/h above 1000 ml/h.
Infusion Volume and Time: The pumps can handle infusion volumes from 0.1 ml to 10000 ml, with increments of 0.01 ml up to 999.99 ml and 1 ml above 1000 ml. Infusion time ranges from 1 minute to 200 hours.
Bolus and Induction Dose: Bolus rates can reach up to 2000 ml/h for 50 ml syringes and up to 400 ml/h for 5 ml syringes. Both bolus and induction doses are programmable by volume, time, or flow rate.
Keep Open Rate (KOR): The KOR is adjustable from 0 to 5 ml/h in 0.1 ml/h increments.
Accuracy and Syringe Compatibility: Flow rate accuracy is ±2% as per EN 60601-2-24 standards. The pumps are compatible with syringes from 5 ml to 50/60 ml of well-known brands, with automatic volume recognition.
Operation Modes: Various models offer different operation modes, including continuous, intermittent, and PCA modes. Advanced features include automatic syringe fixation, drug library storage, and password protection for programmed parameters.
Alarms and Event Logging: The system includes a comprehensive alarm system for various conditions and can log a minimum of 2000 operational activities or alarms with timestamps.
Power Supply and Battery: The pumps operate on a power supply of 100-230 VAC or 12-16 VDC and include a NiMH battery with a working time of 20 hours at 5 ml/h.
Connectivity and Interface: The pumps feature RS 232C and IR connections, with options for nurse call connections and integration with hospital systems via MedimaNet.
Dimensions and Weight: The pumps measure 355 x 115 x 166 mm and weigh below 2.5 kg.
Environmental Conditions: Operating temperatures range from +5 to +40°C with a maximum humidity of 90%. Transport conditions allow for temperatures from -20 to +50°C and humidity up to 95%.
Safety and Compliance: The pumps comply with various safety standards, including EN 60601-1 and MDD 93/42/EEC - II b.
Docking Stations and Accessories: Docking stations accommodate multiple pumps and facilitate easy mounting and connection. Accessories include cords and software tools for configuration and monitoring.
Software and Monitoring: Medima offers software solutions for infusion monitoring, drug library management, and system configuration, enhancing safety and precision in infusion therapy.
See more

Catalog excerpts

Family of syringe infusion pumps-1

MedimaNet Infusion monitoring and archiving network system. Technical specifications Safe and precise infusion Family of syringe infusion pumps - Ward monitor - Infusion progress monitoring - Infusion progress archiving - Reports - Automatic transfer of drug libraries to pumps Medima User ToolBox Software supporting infusion safety and precision. Medima Drug Editor 7WK1 Convenient and safe way of drug libraries creation in Windows environment. Enables fast transfer of drug libraries to pumps. Medima History File Software to archive and review infusions. Graphic presentation of infusion progress, detailed event log, history of infusion parameters modifications. Flow rate 0.1 - 2000 ml/h for 50 ml syringe 0.1 - 1400 ml/h for 30 ml syringe 0.1 - 1000 ml/h for 20 ml syringe 0.1 - 600 ml/h for 10 ml syringe 0.1 - 400 ml/h for 5 ml syringe increments: 0.01 ml/h up to 99.99 ml/h, 0.1 ml/h from 100 ml/h to 999.9 ml/h, 1 ml/h above 1000 ml/h Infusion volume0.1 - 10000 ml, increments: 0.01 ml up to 999.99 ml, 1 ml above 1000 ml Infusion time1 min - 200 h Bolus Up to 2000 ml/h for 50 ml syringes Up to 400 ml/h for 5 ml syringes Increments: 0.01 ml/h up to 99.99 ml/h, 0.1 ml/h from 100 ml/h to 999.9 ml/h, 1 ml/h above 1000 ml/h Programmable volume (dose), time or flow rate. Automatic and manual bolus. Induction dose Up to 2000 ml/h for 50 ml syringes Up to 400 ml/h for 5 ml syringes Increments: 0.01 ml/h up to 99.99 ml/h, 0.1 ml/h from 100 ml/h to 999.9 ml/h, 1 ml/h above 1000 ml/h Programmable volume (dose), time or flow rate. KOR (KVO)0 - 5 ml/h, 0.1 ml/h increments Flow rate accuracy±2% in accordance with EN 60601-2-24* Syringes types5 - 50/60 ml of world known brands**, automatic volume recognition Occlusion pressure11 levels, 50 - 1000 mmHg, automatic reduction of occlusion bolus Operation modes Model Sml/min, ml/h, ml/24h, volume and time Model S1, S2, S-PCA ml/min, ml/h, ml/24h, volume and time ng, pg, mg, pU, mU, U, kU, nmol, pmol, mmol, mol/kg, --/min, h, 24h intermittent infusion profile - 24 cycles Model S-PCAPCA mode Additional functions Automatic syringe fixation. Drug library with possibility to save up to 120 drugs dosing protocols (models S1, S2, S-PCA) with soft and hard limits for all infusion parameters (models S2, S-PCA) Drug library and configuration transfer from pump to pump Programmed parameters and chosen functions protected by password Alarm volume adjustment and choice of sound type Night mode: tuning the display's brightness and alarm loudness, programmable time of start/end Simple fixing to Medima docking station without removing the fixing clamp and without interruption of infusion Embeded power supply Possibility to connect to hospital system through the MedimaNet system Open system: software update through RS232C Alarms No mains, battery low, battery empty, 1 - 30 min. to infusion end, infusion end, 1 - 30 min. to syringe empty, syringe empty, syringe, high pressure, occlusion, line disconnected, improper fixing, drug not active, error xx, infusion broken, autoinfusion block the flow, battery disconnected, timer S-PCA : cover open, lock open, bolus button, bolus blocked dose limit Event log capacityMin. 2000 records of operational activities or alarms with date and time of the event Power supply100 (-15%) - 230 (+10%) VAC, 50/60 Hz, max 20 VA, 12 - 16 VDC, max 1 A BatteryNiMH, working time: 20 h / 5 ml/h. Recharging time to 100% - max. 6 h InterfaceRS 232C and IR connection to the docking station Nurse call connection24V, 1 A, NO / NC Dimensions355 x 115 x 166 mm (W x H x D) - excluding pole clamp WeightBelow 2.5 kg Operating conditionsTemperature: +5 -e +40oC, humidity max. 90% Transport conditionsTemperature: -20 -e +50oC, humidity max. 95% ClassificationClass I, type CF, defibrillation proof, IP22 Safety standards compliance EN 60601-1, EN 60601-1-2, EN 60601-2-24, EN-1789, MDD 93/42/EEC - II b Docking stationsDesigned to accommodate 2, 4, 6 or 8 infusion pumps and a docking station designed for ambulances Accessories *** RS232C cord, DC power supply cord, Nurse call cord Infusion stand for Medima infusion pumps and Medima docking stations fixing Duo holder for fixing and carrying two Medima infusion pumps Software packets: Medima User ToolBox and Medima Service ToolBox, MedimaNet network software * Technical accuracy is 0.5%. Flow rate accuracy depends on the syringe quality. ** List of syringes available on request. There is a possibility to calibrate other syringes according to customer needs. *** Accessories are not standard equipment of pumps In pursuit of quality improvement and due to changing standards the information in this leaflet may be subject to change. Contact Medima Sp. z o. o. to update the information. Medima Service ToolBox Contains Medima Loader and Medima Configurator. Enables pumps upgrade as well as setting and uploading configurations to pumps adjusted to user's needs. Project co-financed by the European Union from European Regional Development Fund medima Free copy - DH-Sx-01-EN-09 2014.10.01

 Open the catalog to page 1
Family of syringe infusion pumps-2

Safe infusion systems Large and Easy-to-Read Graphic Display Shows infusion rate, syringe type, a drug name, power supply status, infusion pressure and other information. Data are readable from a considerable distance. Wide Range of Syringes Syringes of all world known brands from 5ml to 50/60ml. Automatic Syringe Loading and Recognition Facilitates and speeds up pump operation, reduces errors of initial infusion (very important in neonatology). Simple Programming and Easy Operation Quick data input from the numeric keyboard. Parameters can be changed without infusion interruption. Infusion Programming...

 Open the catalog to page 2

All Medima catalogs and technical brochures

  1. Medima P

    2  Pages

  2. Medima S

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.