Product catalog summary
Flow Rate Specifications: The syringe infusion pumps support flow rates from 0.1 to 2000 ml/h depending on syringe size, with increments varying by rate. For example, 0.01 ml/h increments are used up to 99.99 ml/h, 0.1 ml/h from 100 to 999.9 ml/h, and 1 ml/h above 1000 ml/h.
Infusion Volume and Time: The pumps can handle infusion volumes from 0.1 to 10000 ml, with time settings ranging from 1 minute to 200 hours.
Bolus and Induction Dose: Bolus rates can reach up to 2000 ml/h for 50 ml syringes, with programmable volume, time, or flow rate options. Induction doses follow similar specifications.
Keep Open Rate (KOR): Adjustable from 0 to 5 ml/h in 0.1 ml/h increments.
Accuracy and Syringe Compatibility: Flow rate accuracy is ±2% as per EN 60601-2-24 standards. The pumps automatically recognize syringe volumes from 5 to 50/60 ml.
Occlusion Pressure and Operation Modes: There are 11 occlusion pressure levels ranging from 50 to 1000 mmHg. Various operation modes are available, including continuous, intermittent, and PCA.
Additional Functions: Features include automatic syringe fixation, a drug library for up to 120 protocols, password protection, alarm customization, night mode, and connectivity options via MedimaNet.
Alarms and Event Log: The system includes numerous alarms for various conditions and can log a minimum of 2000 events.
Power Supply and Battery: The pumps operate on 100-230 VAC or 12-16 VDC, with a NiMH battery providing up to 20 hours of operation at 5 ml/h.
Interface and Connectivity: RS 232C and IR connections are available, along with nurse call connectivity.
Dimensions and Weight: The pumps measure 355 x 115 x 166 mm and weigh less than 2.5 kg.
Operating and Transport Conditions: Suitable for temperatures from +5 to +40°C and humidity up to 90% during operation, and -20 to +50°C with 95% humidity during transport.
Safety and Compliance: The pumps comply with various safety standards, including EN 60601-1 and MDD 93/42/EEC - II b.
Docking Stations and Accessories: Designed for multiple pumps, with optional LAN connectivity. Accessories include cords and software tools for configuration and monitoring.
Software Tools: Medima offers software for infusion monitoring, drug library management, and pump configuration.
See more

Catalog excerpts

Medima S-1

Infusion monitoring and archiving network system. Safe and precise infusion Medima Sp. z o.o. Al. Jerozolimskie 200, 02-486 Warsaw, Poland tel. +48 (22) 313 22 66 fax. +48 (22) 313 22 69 email: [email protected] Family of syringe infusion pumps Technical specifications Flow rate - Ward monitor - Infusion progress monitoring - Infusion progress archiving - Reports - Automatic transfer of drug libraries to pumps Infusion volume Infusion time Bolus Medima User ToolBox Software supporting infusion safety and precision. Induction dose Medima Drug Editor KOR (KVO) Flow rate accuracy Syringes types Occlusion pressure Operation modes Additional functions Convenient and safe way of drug libraries creation in Windows environment. Enables fast transfer of drug libraries to pumps. Medima History File Event log capacity Power supply Battery Interface Nurse call connection Dimensions Weight Operating conditions Transport conditions Classi�cation Safety standards compliance Docking stations Accessories *** Software to archive and review infusions. Graphic presentation of infusion progress, detailed event log, history of infusion parameters modifications. 0.1 - 2000 ml/h for 50 ml syringe 0.1 - 1400 ml/h for 30 ml syringe 0.1 - 1000 ml/h for 20 ml syringe 0.1 - 600 ml/h for 10 ml syringe 0.1 - 400 ml/h for 5 ml syringe increments: 0.01 ml/h up to 99.99 ml/h, 0.1 ml/h from 100 ml/h to 999.9 ml/h, 1 ml/h above 1000 ml/h 0.1 - 10000 ml, increments: 0.01 ml up to 999.99 ml, 1 ml above 1000 ml 1 min - 200 h Up to 2000 ml/h for 50 ml syringes Up to 400 ml/h for 5 ml syringes Increments: 0.01 ml/h up to 99.99 ml/h, 0.1 ml/h from 100 ml/h to 999.9 ml/h, 1 ml/h above 1000 ml/h Programmable volume (dose), time or �ow rate. Automatic and manual bolus. Up to 2000 ml/h for 50 ml syringes Up to 400 ml/h for 5 ml syringes Increments: 0.01 ml/h up to 99.99 ml/h, 0.1 ml/h from 100 ml/h to 999.9 ml/h, 1 ml/h above 1000 ml/h Programmable volume (dose), time or �ow rate. 0 - 5 ml/h, 0.1 ml/h increments ±2% in accordance with EN 60601-2-24* 5 - 50/60 ml of world known brands**, automatic volume recognition 11 levels, 50 - 1000 mmHg, automatic reduction of occlusion bolus Model S ml/min, ml/h, ml/24h, volume and time Model S1, S2, S-PCA ml/min, ml/h, ml/24h, volume and time ng, μg, mg, μU, mU, U, kU, nmol, μmol, mmol, mol/kg, --/min, h, 24h intermittent infusion pro�le - 24 cycles Model S-PCA PCA mode Automatic syringe �xation. Drug library with possibility to save up to 120 drugs dosing protocols (models S1, S2, S-PCA) with soft and hard limits for all infusion parameters (models S2, S-PCA) Drug library and con�guration transfer from pump to pump Programmed parameters and chosen functions protected by password Alarm volume adjustment and choice of sound type Night mode: tuning the display’s brightness and alarm loudness, programmable time of start/end Simple �xing to Medima docking station without removing the �xing clamp and without interruption of infusion Embeded power supply Possibility to connect to hospital system through the MedimaNet system Open system: software update through RS232C No mains, battery low, battery empty, 1 - 30 min. to infusion end, infusion end, 1 - 30 min. to syringe empty, syringe empty, syringe, high pressure, occlusion, line disconnected, improper �xing, drug not active, error xx, infusion broken, autoinfusion block the �ow, battery disconnected, timer S-PCA : cover open, lock open, bolus button, bolus blocked dose limit Min. 2000 records of operational activities or alarms with date and time of the event 100 (-15%) - 230 (+10%) VAC, 50/60 Hz, max 20 VA, 12 - 16 VDC, max 1 A NiMH, working time: 20 h / 5 ml/h. Recharging time to 100% - max. 6 h RS 232C and IR connection to the docking station 24V, 1 A, NO / NC 355 x 115 x 166 mm (W x H x D) - excluding pole clamp Below 2.5 kg Temperature: +5 ÷ +40oC, humidity max. 90% Temperature: -20 ÷ +50oC, humidity max. 95% Class I, type CF, de�brillation proof, IP22 EN 60601-1, EN 60601-1-2, EN 60601-2-24, EN-1789, MDD 93/42/EEC - II b Designed to accommodate 2, 4, 6 or 8 infusion pumps and a docking station designed for ambulances RS232C cord, DC power supply cord, Nurse call cord Infusion stand for Medima infusion pumps and Medima docking stations �xing Duo holder for �xing and carrying two Medima infusion pumps Software packets: Medima User ToolBox and Medima Service ToolBox, MedimaNet network software * Technical accuracy is 0.5%. Flow rate accuracy depends on the syringe quality. ** List of syringes available on request. There is a possibility to calibrate other syringes according to customer needs. *** Accessories are not standard equipment of pumps In pursuit of quality improvement and due to changing standards the information in this lea�et may be subject to change. Contact Medima Sp. z o. o. to update the information. Medima Service ToolBox Contains Medima Loader and Medima Configurator. Enables pumps upgrade as well as setting and uploading configurations to pumps adjusted to user’s needs.

 Open the catalog to page 1
Medima S-2

Safe infusion systems Large and Easy-to-Read Graphic Display Shows infusion rate, syringe type, a drug name, power supply status, infusion pressure and other information. Data are readable from a considerable distance. Wide Range of Syringes Syringes of all world known brands from 5ml to 50/60ml. Automatic Syringe Loading and Recognition Facilitates and speeds up pump operation, reduces errors of initial infusion (very important in neonatology). Simple Programming and Easy Operation Quick data input from the numeric keyboard. Parameters can be changed without infusion interruption. Infusion Programming...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.