Processing (Cleaning, Disinfection and Sterilisation) of the Medijet® 1010/ 1020 medin Medical Innovations GmbH Adam-Geisler-Str. 1 D-82140 Olching +49 8142-44846-0 www.medingmbh.com info@medingmbh.com

Open the catalog to page 1
OP_Medijet_1010_1020 - 2

Operating, cleaning and maintenance instructions General Principles All instruments must be cleaned, disinfected and sterilised before each use; this also applies in particular to initial use after delivery, since all instruments are delivered non-sterile (cleaning and disinfection after removal from protective transport packaging; sterilisation after packaging). Effective cleaning and disinfection is an indispensable prerequisite for effective sterilisation. Within the scope of your responsibility for sterility of the instruments during use, please ensure that in principle, only...

Open the catalog to page 2
OP_Medijet_1010_1020 - 3

Operating, cleaning and maintenance instructions 2.3 Manual Cleaning and Disinfection When selecting the cleaning agent and disinfectant used, the following should be borne in mind: that these are fundamentally suitable for the cleaning and disinfection of instruments made of metals and plastics, that the cleaning agent – if applicable – is suitable for ultrasound cleaning (no development of foam) that a disinfectant whose efficacy has been certified (e.g. VAH/DGHM or FDA/EPA approval or CE mark) is used and that this disinfectant is compatible with the cleaning agent used, and that the...

Open the catalog to page 3
OP_Medijet_1010_1020 - 4

Operating, cleaning and maintenance instructions Proof of fundamental suitability of the instruments for effective steam sterilisation was provided by an independent, accredited testing laboratory using the steam steriliser Systec V-150 (Systec GmbH Labor-Systemtechnik, Wettenberg) and the fractionated vacuum method as well as the gravity displacement method. In doing so, typical conditions in the hospital and medical practice as well as the method described above were taken into account. The flash sterilisation method is fundamentally not permissible. In addition, do not use any hot air...

Open the catalog to page 4

All Medin Medical Innovations GmbH catalogs and technical brochures

  1. Nuflow®

    2 Pages

  2. Pb_1000_long

    3 Pages

  3. OP_Medijet

    103 Pages

  4. Pb_4000_long

    3 Pages

  5. OP_Miniflow

    103 Pages

  6. OP Bubble

    2 Pages

  7. Processing

    4 Pages

  8. Pb_1207_long

    2 Pages

  9. PB_1207MKI

    2 Pages

  10. PB_1207

    2 Pages

Archived catalogs

  1. MEDINsindi

    2 Pages

  2. Medin_Catalog

    16 Pages

  3. OP_Blender

    18 Pages

  4. OP_Medijet_1000

    104 Pages

  5. products_medin

    16 Pages