Ultra Low Plasma freezer with backup
6Pages

{{requestButtons}}

Catalog excerpts

Ultra Low Plasma freezer with backup - 1

Environmentally friendly Ultra Deep Freezer, -86 degree powered by Meditech Technology for safe storage of Plasma Bags . Stable Temperature Control Temperature is controlled by two sensors located in the liquid-loaded monitor bottles, Two thermistor sensors constantly monitor the temperature in both the upper and the lower part of chamber. Microprocessor control ensures the most accurate temperature control available. Temperature-maintained defrost designed with thermal sensors and heaters on the evaporator, all under precise microprocessor control. Heated air Vent to prevent vacuum formation. Designed to conform to AABB criteria, the series ensures stable and reliable temperature control utilizing Meditech original technology. A special highly efficient compressor designed and used by Meditech provides rapid cooling and quiet performance for each model. Pull Down Time around 4.5 hours. Benchmark in industry Temperature Variations Prevented The Meditech MTUDF Series is designed to minimize cold air loss even with frequent door openings. Separated transparent inner doors minimize the chamber air leakage during door openings. Foamed-in-place insulation with vacuum insulated panels in the walls and magnetic sealed outer doors with prevent chamber air leakage and promote complete door closings User Friendly Design Selectable storage system LED interior lamp with ON/OFF switch Eye Level Digital display is easy to see, and is calibratable through the control panel Filterless construction eliminates bothersome filter cleaning Alarm and Safety Functions To ensure the safety of precious blood supply, the Meditech Series provides the following safety functions: Audible and flashing LED visual alarms with remote alarm contacts, in case of power failure, high or low temperature Door Alarm and key lock are standard features Built-in temperature recorder.

Open the catalog to page 1
Ultra Low Plasma freezer with backup - 2

2 Know your Machine Equipment meets electrical safety specification such as that of IEC (ClassI) The display is positioned at eye-level to make it easier to read and simplify operation. Digital PLC Controller CFC free refrigerant gas/green gas CFC free insulation External Corrosion Resistant, powder coated (CR at least 1mm thickness) Led Light along the length of the Refrigerator for uniform illumination through all the Compartment Refrigerator Body Glass & Solid door, Automatic closing of the front door below opening angle of 90º and opening angle limited to 110º.Insulation and gasket is...

Open the catalog to page 2
Ultra Low Plasma freezer with backup - 3

LED INDICATION 1. Line In 2. Power 3. Comp On 4. Heater On 5. Battery On 6. Battery Low 7. Temp High 8. Temp Low 9. Power Fail 10. Sensor Fail 11. Chart Change 12. System On Scrolling LIVE Data logger on LED screen Temperature Incoming Voltage Ambient Temperature Time in hours of revolution chart Current Date Current Time Battery Voltage User Friendly Settings 1. Date 2. Time 3. High Temp Alarm 4. Low Temp Alarm 5. Hysterisis 6. Compressor Delay

Open the catalog to page 3
Ultra Low Plasma freezer with backup - 4

7” Touch Screen Controller Scrolling LIVE Data logger on LED screen Special Features 1. large 7 inch touch screen with display of set and process value 2. Precise control of temperature by using micro controller based temperature controller. 3. LN2 controlling System. 4. Integrated digital temperature cum controller with inbuilt 7 days circular chart Recorder. PLC with battery back up of 8 hours to ensure display & recording of temperature even during power failures on touch screen. 5. Integrated data management system with storage efficiency of temperature records up to 1000 readings &...

Open the catalog to page 4
Ultra Low Plasma freezer with backup - 5

LN2 Back-up System The LN2 Back-up System is designed to provide a means of maintaining the contents of Ultra Low Temperature Freezer at a preset temperature should the power or the refrigeration system fail. The freezer is fitted with a platinum resistance thermometer (referred to as a temperature sensor or probe) to measure the internal temperature, a solenoid valve and a door switch. Transfer lines and connectors are supplied to connect between the solenoid valve and the LN2 low-pressure storage. The backup control option is activated on Touch Panel and a temperature selected between -50...

Open the catalog to page 5
Ultra Low Plasma freezer with backup - 6

MTUD Series Model 1. Meditech can manufacture freezers according to customer Specifications and size 2. Full SS Body with life time warranty available on customer demand 3. Customer should specify Touch Screen controlling options at the tome of ordering. Standard configuration is PLC Controller.

Open the catalog to page 6

All Meditech Technologies India catalogs and technical brochures

  1. walk in cooler

    11 Pages

  2. cold room

    10 Pages

  3. co2 incubator

    15 Pages

  4. BOD Incubator

    10 Pages

  5. Incubator

    2 Pages

  6. MTCR 60

    6 Pages

  7. MTPS432

    5 Pages

  8. MTPS360

    5 Pages

  9. MTPS288

    5 Pages

  10. MTPS216

    5 Pages

  11. MTPS144

    5 Pages

  12. MTPS72

    5 Pages

  13. MBDW80

    11 Pages

  14. MBDW120

    11 Pages

  15. MBDW40

    11 Pages

  16. Camping Cot

    1 Pages

  17. MTR30

    3 Pages

  18. MTR26

    3 Pages

  19. ME7-2

    3 Pages

  20. ME8

    8 Pages

  21. MTBDR16

    3 Pages

  22. MTAC60

    7 Pages

  23. MTAC40

    7 Pages

  24. MTAC26

    3 Pages

  25. MA1000

    3 Pages

  26. Spirometer

    1 Pages

  27. Cold rooms

    10 Pages

Archived catalogs