VISIA AF™ VR Model DVAB1D1
7Pages

{{requestButtons}}

Catalog excerpts

VISIA AF™ VR Model DVAB1D1 - 1

Product specifications Physical characteristics Volumea Surface area of device can Materials in contact with human tissued Titanium, polyurethane, silicone rubber Battery chemistry Hybrid CFx lithium/silver vanadium oxide Volume with connector ports unplugged. Grommets may protrude slightly beyond the can surface. c The radiopaque ID and Medtronic radiopaque identifier can be viewed in a fluoroscopic image of the device. d These materials have been successfully tested for the ability to avoid biological incompatibility. The device does not produce an injurious temperature in the surrounding tissue during normal operation. a Replacement indicators < 2.73 V on 3 consecutive Recommended Replacement daily automatic Time (RRT) measurements End of Service (EOS) Maximum energy levels and typical full energy charge times Maximum programmed energy Maximum delivered energya,b 36 J Maximum stored energyc Typical charge time at Beginning of Service (BOS)d Typical charge time at Recommended Replacement Time (RRT)d a Energy delivered at connector block into a 50 Ω load. For 35 J programmed energy, delivered energy exceeds 35 J. c Energy stored at charge end on capacitor. d Charge time during a nonwireless telemetry session may be slightly higher. b § AF Diagnostics § PhysioCurve™ Design § SmartShock™ 2.0 Technology § OptiVol™ 2.0 Fluid Status Monitoring § Complete Capture Management

Open the catalog to page 1
VISIA AF™ VR Model DVAB1D1 - 2

Device parameters RV Lead Noise Timeout AF detection parameters Parameter Least sensitive; Less sensitive; Balanced sensitivity ; More sensitive; Most sensitive Ectopy Rejection Episodes ≥ 6 min ; Episiodes ≥ 10 min; Episodes ≥ 20 min; Episodes ≥ 30 min; Episodes ≥ 60 min Programmable values The measured intervals are truncated to a 10 ms multiple (for example, 457 ms becomes 450 ms). The device uses this truncated interval value when applying the programmed criteria and calculating interval averages. b The Wavelet feature is automatically set to On when VF Detection is set to On. c This...

Open the catalog to page 2
VISIA AF™ VR Model DVAB1D1 - 3

Ventricular tachyarrhythmia therapy parameters, cont. Programmable values Programmable values Smart Modeb Ramp therapy parameters … 15 Initial # Pulses Smart Modeb RV Capture Management™ parameters Ramp+ therapy parameters … 15 Initial # Pulses Smart Modeb V-V Minimum ATP Interval V. Amplitude Active Can™/SVC Coil RV Amplitude Safety Margin RV Minimum Adapted Amplitude RV Acute Phase Remaining Blanking periods Parameter Rate response pacing parameters If the Active Can/SVC Coil parameter is set to Can Off, the Active Can electrode is not used as part of the high-voltage delivery pathway. If...

Open the catalog to page 3
VISIA AF™ VR Model DVAB1D1 - 4

Ventricular rate stabilization parameters Parameter Programmable values Maximum Rate Interval Increment Avg. V. Rate During AF Alert Enablec Shared (Device Tone and Patient Home Monitor) AF Daily Burden Daily Burden for Avg. V. Rate Post VT/VF shock pacing parameters Parameter Number of Shocks Delivered in an Episoded Programmable values Device Tone Post VT/VF Shock Pacing Overdrive Rate Overdrive Duration Alert Enable – Urgency Patient Home Monitor Alert Enablec Shared (Device Tone and Patient Home Monitor) Post shock pacing parameters Parameter Programmable values Post Shock V. Amplitude...

Open the catalog to page 4
VISIA AF™ VR Model DVAB1D1 - 5

Lead/Device integrity alerts, cont. Parameter SVC Defibrillation Impedance Enableb,c Programmable values RVtip to RVcoil; RVtip to RVring Can to RVcoil ; Can to RVring; RVtip to RVcoil; RVtip to RVring; Can to SVCb,d; RVcoil to SVCb Shared (Device Tone and Patient Home Monitor) RV Pacing Impedance Less than RV Pacing Impedance Greater than RV Defibrillation Impedance Less than RVtip to RVcoil ; RVtip to RVring RV Defibrillation Impedance Greater than SVC Defibrillation Impedance Less than SVC Defibrillation Impedance Greater than EGM1 and EGM2 ; EGM1 and EGM3; EGM1 and LECG; EGM2 and EGM3;...

Open the catalog to page 5
VISIA AF™ VR Model DVAB1D1 - 6

Manual cardioversion parameters T-Shock induction parameters Selectable values Selectable values Enabled ; Disabled Enabled; Disabled AX>B; B>AX Monophasic ; Biphasic AX>B; B>AX If the Active Can/SVC Coil parameter is set to Can Off, the Active Can electrode is not used as part of the high-voltage delivery pathway. If the Active Can/SVC Coil parameter is set to SVC Off, the SVC Coil electrode is not used as part of the high-voltage delivery pathway. 50 Hz Burst induction parameters Selectable values Minimum Interval Pulse Width Manual Ramp therapy parameters Parameter Selectable values...

Open the catalog to page 6
VISIA AF™ VR Model DVAB1D1 - 7

Longevity Projected service life in years Projected service life in years Pacing Mode, percent pacing Pacing Amplitude The service life projections are based on the following assumptions: • Semi-annual maximum energy charging frequency • 1 hour of in-office wireless telemetry annually • Pre-arrhythmia EGM storage programmed to On for a • Typical shelf storage time before implant 6-month period (two 3-month follow-up intervals), Projected service life estimates are based on accelerated over the entire life of the device battery discharge data and device modeling as specified. • 3 hours of...

Open the catalog to page 7

All Medtronic catalogs and technical brochures

  1. EverFlex™

    4 Pages

  2. Micra™

    12 Pages

  3. Provia™

    4 Pages

  4. Intuition™

    3 Pages

  5. 5392

    2 Pages

  6. Evolut™ PRO+

    12 Pages

  7. SOLOIST™

    2 Pages

  8. BIS

    16 Pages

Archived catalogs

  1. MicroFrance®

    166 Pages

  2. HMS Plus

    6 Pages

  3. Endurant II

    8 Pages

  4. StealthViz

    4 Pages

  5. ENT product

    476 Pages

  6. Parts catalog

    15 Pages