VIVA™ XTCRT-D
12Pages

{{requestButtons}}

Catalog excerpts

VIVA™ XTCRT-D - 1

Product specifications Physical characteristics Volume3 35 cm3 Surface area of device can 57 cm2 Radiopaque IDc PXT Materials in contact with Titanium, polyurethane, human tissued silicone rubber Hybrid CFx lithium/silver vanadium oxide a Volume with connector ports unplugged. b Grommets may protrude slightly beyond the can surface. c The radiopaque ID, which includes a Medtronic-identifier symbol, can be viewed in a fluoroscopic image of the device. d These materials have been successfully tested for the ability to avoid biological incompatibility. The device does not produce an injurious temperature in the surrounding tissue during normal operation. Replacement indicators Recommended < 2.73 V on 3 consecutive daily Replacement Time (RRT) automatic measurements End of Service (EOS) 3 months after RRT Maximum energy levels and typical full energy charge times Maximum programmed energy 35 J Maximum delivered energy^ 36 J Typical charge time at Beginning of Service (BOS)d Typical charge time at Recommended 0 Replacement Time (RRT)d . s a Energy delivered at connector block into a 50 Q load. b For 35 J programmed energy, delivered energy exceeds 35 J. c Energy stored at charge end on capacitor. d Charge time during a nonwireless telemetry session may be slightly higher. PhysioCurve™ Design AdaptivCRT™ Algorithm CardioSync™ Optimization SmartShock™ Technology OptiVol™ 2.0 Fluid Status Monitoring Complete Capture Management™

Open the catalog to page 1
VIVA™ XTCRT-D - 2

Device parameters Tachyarrhythmia detection parameters Parameter Programmable values Other enhancements Stabilitya Onset AT/AF Detection Onset Percent AT/AF Interval (Rate)a High Rate Timeout Fast AT/AF Interval (Rate)a T-Wave RV Lead Noise Timeout Sensitivity Atrialc'd FVT Detection FVT Interval (Rate)a VT Detection VT Interval (Rate)a VT Initial Beats to Detect VT Beats to Redetect VT Monitor VT Monitor Interval (Rate)a Monitored VT Beats to Detect PR Logic™/Wavelet Template [date] a The measured intervals are truncated to a 10 ms multiple (for example, 457 ms becomes 450 ms). The device...

Open the catalog to page 2
VIVA™ XTCRT-D - 3

Atrial tachyarrhythmia therapy parameters, continued Programmable values Automatic CV AT/AF Automatic CV Status Pathwaya Fast AT/AF Automatic CV Status Interval Decrement Stop Atrial Rx after (Shared) Rx/Lead Suspect Pathwaya Shared CV Minimum R-R Intervalb Disable Atrial ATP if it accelerates V. rate? Active Can™/SVC Coilc Disable all atrial therapies if atrial lead position is suspect? Yes 3>; No (Atrial Lead Position Check) Automatic CV Limits Delivery Window 00:00; 01:00; 02:00; 03:00 $> Duration to stop 12; 24; 48 <0>; 72 hr; None Episode Duration before Rx Delivery Episode Duration...

Open the catalog to page 3
VIVA™ XTCRT-D - 4

Ventricular tachyarrhythmia therapy parameters Parameter Programmable values VF Therapy parameters VF Therapy Status On <0>; Off Ramp therapy parameters Initial # Pulses 1; 2 ... 8 <6> ... 15 1; 2 ... 10 # Sequences VT Therapies: 3 <0>; FVT Therapies: AX>B; B>AX Rx1-Rx4: B>AX<8>; Rx5-Rx6: AX>B <6> During Charging O; Before Charging; Off Smart Modeb Ramp+ therapy parameters Initial # Pulses Therapy Type Burst <0>; Ramp; Ramp+ Switch when number of consecutive ATP successes equals Smart Mode 1; 2 ... 10 # Sequences VT Therapies: 3 <§>; FVT Therapies: VT/FVT Therapy parameters VT Therapy...

Open the catalog to page 4
VIVA™ XTCRT-D - 5

Pacing parameters Modes, rates, and intervals Parameter Programmable values .. , DDDR; DDD <6>; DDIR; DDI; AAIR; AAI; Mode VVIR; VVI; DOO; AOO; VOO; ODO Mode Switch On <6>; Off Minimum PVARP 150; 160 ... 250 O... 500 ms A. Refractory Period 150; 160 ... 310 <6> ... 500 ms a The corresponding Lower Rate Interval can be calculated as follows: Lower Rate Interval (ms) = 60,000/Lower Rate. b If CRT is adaptive, Paced AV and Sensed AV cannot be selected or programmed. CRT pacing parameters Parameter Programmable values Adaptive Bi-V and LV <0>; Adaptive Bi-V; Nonadaptive CRT Maximum Rate Atrial...

Open the catalog to page 5
VIVA™ XTCRT-D - 6

Rate Profile Optimization ADL Response Exertion Response Activity Threshold Activity Acceleration Activity Deceleration ADL Set Point UR Set Point Rate response pacing parameters Parameter Programmable values Conducted AF response parameters Parameter Programmable values Conducted AF Response On <$>; Off Response Level Low; Medium O; High Ventricular rate stabilization parameters Parameter Programmable values Post VT/VF shock pacing parameters Parameter Programmable values Rate adaptive AV parameters Programmable values Post VT/VF Shock Pacing Overdrive Rate Start Rate Overdrive Duration...

Open the catalog to page 6
VIVA™ XTCRT-D - 7

Sleep parameters Parameter Programmable values Number of Shocks Delivered in an Episoded Device Tone Alert Enable - Urgency Off ft; On-Low; On-High Patient Home Monitor Alert Enableb Shared (Device Tone and Patient Home Monitor) Non-Competitive Atrial Pacing (NCAP) parameters Number of Shocks Thresholda NCAP Interval Programmable values Non-Comp Atrial Pacing Additional pacing features Parameter Programmable values a Delivered as LV pacing when the AdaptivCRT™ operating value is LV. Otherwise, delivered as RV pacing. All Therapies in a Zone Exhausted for an Episode Device Tone Alert Enable...

Open the catalog to page 7
VIVA™ XTCRT-D - 8

Lead/Device integrity alerts, continued Parameter A. Pacing Impedance Enablec RV Pacing Impedance Enablec LV Pacing Impedance Enablec Programmable values Shared parameters Parameter Programmable values Patient Home Monitor Yes; No <6> a This parameter is displayed only if an associated alert has been enabled. Data collection parameters RV Defibrillation Impedance Enablec SVC Defibrillation Impedance Enableb,c Data collection parameters Parameter Programmable values LECG Source Can to SVC Ob,c; RVcoil to Aring; (Leadless ECG)a Can to Aring Shared (Device Tone and Patient Home Monitor) A....

Open the catalog to page 8
VIVA™ XTCRT-D - 9

System test parameters EP study parameters System test parameters Parameter Selectable values Pacing Threshold Test parameters Amplitude; Pulse Width; Test Type (LV Test) Phrenic Nerve Stim - Amplitude; Phrenic Nerve Stim - Pulse Width Test Type (Atrium or RV Test) Amplitude; Pulse Width T-Shock induction parameters Parameter Selectable valuesChamber1 RV <©>; RV+LV; LV Resume at Deliver Enabled <6>; Disabled Decrement after Pace Polarity (RV) Bipolar; Tip to Coil LVtip to RVcoil; LVring to RVcoil; LVtip to LVring; LVring to LVtip Modea (RV or LV test) VVI; VOO; DDI; DDD; DOO Modea (Atrium...

Open the catalog to page 9

All Medtronic catalogs and technical brochures

  1. EverFlex™

    4 Pages

  2. Micra™

    12 Pages

  3. Provia™

    4 Pages

  4. Intuition™

    3 Pages

  5. 5392

    2 Pages

  6. Evolut™ PRO+

    12 Pages

  7. SOLOIST™

    2 Pages

  8. BIS

    16 Pages

Archived catalogs

  1. MicroFrance®

    166 Pages

  2. HMS Plus

    6 Pages

  3. Endurant II

    8 Pages

  4. StealthViz

    4 Pages

  5. ENT product

    476 Pages

  6. Parts catalog

    15 Pages