Technical Specifications:
The AccuFill™ KFAS-L2 is designed for filling volumes over 1 liter. It operates on either a single-phase 110VAC, 50/60 Hz, 15A or a single-phase 220VAC, 50/60 Hz, 7.5A power supply. The fill measurement system utilizes a Coriolis Mass Flow Meter with a single-use fill line, ensuring non-contact filling. The system includes two manual loading fill stations and a radio-frequency (RF) fill tube seal system.

Fill Chamber:
The fill chamber is equipped with HEPA ISO 5 (Class 100) positive pressure controls, ensuring a clean environment. The chamber materials are made of 304/316 L stainless steel.

Pneumatic and Filling Requirements:
The system requires dry, filtered air at 90-120 psi and 15 SCFM. The suggested fill volume range is from 1 L to over 1000 L, with product filling pressures ranging from a minimum of 1.4 bar to a maximum of 3 bar. Fill times are variable, generally achieving 15 L/min, with filling accuracy of ± 2.5% at 1 L and ± 1% above 5 L.

Controls and Validation:
The system is controlled via a PLC with an embedded computer and a color touch screen. It includes a validated aseptic particle counting system for continuous real-time monitoring in the fill zone. Validation support is available.

Additional Features:
The system is CE marked and weighs approximately 1,082 lbs. Optional features include a transfer table and a swing arm control panel, with dimensions and specifications subject to change. Drawings illustrate these optional components.
See more

Catalog excerpts

AccuFill™ KFAS-L2 - Fills over 1 L-1

AccuFill™ KFAS-L2 - Fills over 1 L Optional Transfer Table and Swing Arm Control Panel Available – Dimensions and Specifications Subject To Change Drawings Illustrate Optional Swing Arm Control Panel and Large Bag Transfer Table SINGLE-USE NON-CONTACT ASEPTIC FILLER VALIDATED ASEPTIC Technical Specifications: AC Input Requirements Single Phase 110VAC, 50/60 Hz, 15A or Single Phase 220VAC, 50/60 Hz, 7.5A Pneumatic Requirements Product Filling Pressures Filling Stations Suggested Fill Volume Range Filling Accuracy 2 Manual - Loading Fill Stations 1 L - 1000 L + ± 2.5% @ 1 L, ± 1% above 5 L with feed control system Fill Measurement System Coriolis Mass Flow Meter with Single-Use Fill Line Single-Use Fill System Non Contact Filling using the Single-Use Fill Line Fill Times Fill Tube Seal System Controls Fill Chamber Materials Variable According to Feed Pressures, Generally 15 L/Min. Radio-Frequency (RF) PLC with Embedded Computer, Color Touch Screen 304/316 L Stainless Steel Fill Chamber HEPA ISO 5 (Class 100) with Positive Pressure Controls Particle Counting System Continuous real time particle counting in fill zone Weight Validated Aseptic CE Mark ~ 1,082 lbs Net Validation Support Available CE Marked 1001 Flynn Road • Camarillo, CA 93012 • USA +1.805.388.9911 • +1.805.388.5948 Fax www.meissner.c

 Open the catalog to page 1

All Meissner Filtration Products catalogs and technical brochures

  1. M-Flex

    4  Pages

  2. Duraclear

    4  Pages

  3. SepraPor®

    4  Pages

  4. ALpHA G

    4  Pages

  5. Small Flow

    12  Pages

  6. TepoFlex®

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.