Overview: The flexiTOC is a high-sensitivity Total Organic Carbon (TOC) analyzer designed for applications ranging from tap water to ultra-pure water. It utilizes UV oxidation to convert organic carbon compounds in aqueous samples to carbon dioxide (CO₂), which is then quantified using a sensitive NDIR detector.
Technical Specifications & Requirements:
  • Method & Detection: UV oxidation with CO₂ detection (NDIR).
  • Type of Sample: Tap to ultra-pure water.
  • Sample Volume: 5 – 30 mL, with automatic addition of reagent or dilution water.
  • Period of Measurement: Sample interval of 5 - 10 minutes, with user-defined interruptions up to 24 hours.
  • Measurement Range: 0.05 to 20 ppm, up to 100 ppm in diluting mode.
  • Detection Accuracy: Less than 500 ppb ± 5 ppb, greater than 500 ppb ± 1% within non-diluting mode.
  • Ports: 1 for sample, 1 for dilution water, 1 for reagent; 1 for sample, 3 for SST.
  • Related Norms: USP <643>, EP 2.2.44, DIN EN 1484.
  • Ambient Parameters: Relative humidity: 10 to 95%, temperature: 10 to 45 °C.
  • Interface Ports: Analog (0 - 20 mA), according to Namur NE43.
  • Carrier Gas: N2, O2, compressed air, with gas pressure of 1 - 4 bar and consumption of approximately 12 L/h.
  • Inlet Pressure & Temperature: Less than 0.5 bar (pressure reducer needed if above 0.5 bar) & 50 °C.
Options & Features:
  • Multiline Mode: Supports up to 4 different inlets and sequential measurement of up to 3 loops.
  • Sample Flow Check: Initiates alarm if inlet flow equals zero, in combination with a pressure sensor.
  • 21 CFR Part 11 Software: Ensures secure data handling for pharma applications.
  • Heat Exchanger & Pressure Regulator: Required for medium temperatures above 50 °C and pressures exceeding 0.5 bar.
  • Sample Mode: Allows for offline measurement of separate samples with customizable sequences.
Calibration & System Verification: Enables reagent and water blank determination, multipoint calibration with internal standard dilution for enhanced accuracy.
Data Export: Measurement data can be exported to external storage devices in various formats and stored in the cloud for error diagnosis.
System Status: Provides a detailed status overview, including operating hours for sensors, pumps, and the UV lamp.
Protection & Performance: The flexiTOC features IP54 protection for its screen and fan, optimized UV reactor geometry, and harmonized gas flow for maximum oxidation efficiency.
See more

Catalog excerpts

Datasheet flexiTOC-1

flexiTOC The next generation TOC analyzer for Tap Water to Ultra Pure Water The flexiTOC is a high-sensitivity Total Organic Carbon (TOC) analyzer designed for tap water to ultra pure water applications. The analytical method is based on UV oxidation of organic carbon compounds in aqueous samples to carbon dioxide (CO₂). The generated CO₂ is quantified using a highly sensitive NDIR detector, ensuring excellent detection limits and measurement stability. The system supports reagent-free operation for TOC levels below 2000 ppb and operation with reagent for higher TOC concentrations and less oxidative substances, depending on the application. Technical Speciifications & Requirements method & detection UV oxidation with CO₂ detection (NDIR) tap to ultra pure water sample volume 5 – 30 mL reagent or dilution water is added automatically Sample interval 5 -10 min, user defined interruptions up to 24h measurement range detection accuracy < 500 ppb ± 5 ppb, > 500 ppb ± 1 % within non diluting mode 1 for sample, 1 for dilution water, 1 for reagent | 1 for sample, 3 for SST related norms ambient parameters relative humidity: 10 to 95 %, temperature: 10 to 45 °C interface ports analog (0 - 20 mA), according to Namur NE43 N2, O2, compressed air, gas pressure: 1 - 4 bar, consumption: approx. 12 L/h inlet pressure & temperature < 0.5 bar (> 0.5 bar pressure reducer is needed)

 Open the catalog to page 1
Datasheet flexiTOC-2

Options & Features Multiline Mode The flexiTOC pro is providing 4 different inlets. The additional software upgrade allows the user to measure up to 3 loops sequentially, and configure the duration of each interval measurement separately. Sample mode A sample mode configuration is available, and used to measure separate samples offline. This sequence can be customized by the user. Enhanced hardware performance Optimized UV reactor geometry and harmonized gas flow maximize oxidation efficiency, supported by continuous monitoring with automatic zero-flow alarm. Check of sample flow Initiates alarm...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.