ENVIRONMENTAL TEST CHAMBERS
Open the catalog to page 1YOU SET THE LIMITS. WE HOLD THE CONDITIONS. TESTA-CT and TESTA-TT environmental test chambers expand the Memmert portfolio with solutions for precise environmental simulation. Featuring a wide temperature range from –52 / –75 °C up to +180 °C, controlled humidity from 10 to 98% RH (CT models), and temperature change rates of up to 10 K/min on request, they ensure reliable and reproducible testing conditions.
Open the catalog to page 2Benefit from the combined strengths of two industry leaders. Memmert, a pioneer in climate and temperature control since 1933, and Aralab, known for its innovation in climate chambers since 1985, bring decades of specialized knowledge to the table. This collaboration ensures that customers receive the highest quality solutions, backed by the expertise of two of the most trusted names in the industry. OFFER HIGHLY PRECISE AND REPRODUCIBLE CONDITIONS FOR CLIMATIC AND TEMPERATURE TESTING IN MANY INDUSTRIES. • ENVIRONMENTAL TESTING • ELECTRONICS, AUTOMOTIVE, AEROSPACE, • BUILDING MATERIALS, MATERIALS...
Open the catalog to page 3TESTA CHAMBERS - MODELS AND REFERENCESTESTA TT CHAMBERS - TEMPERATURE ONLY TESTA TT CHAMBERS TEMPERATURE RANGE HUMIDITY RANGE TESTA TT E50 TESTA CT CHAMBERS - TEMPERATURE AND HUMIDITY TESTA CT CHAMBERS TEMPERATURE RANGE HUMIDITY RANGE TESTA CT EP50, EC50 or ECP50 EC - models with Capacitive humidity sensor EP - models with Psychrometric humidity sensor ECP - models with both Capacitive and Psychrometric humidity sensors. Please consult Aralab if in doubt about the type of sensor to chose RANGES FOR CLIMATIC AND TEMPERATURE TESTINGTESTA CT TESTING CHAMBERS Performance in CLIMATIC testing range...
Open the catalog to page 4SYSTEM STRUCTURE 6. Observation window with working light (optional) 8. Touch screen controller
Open the catalog to page 5TESTA - ENVIRONMENTAL TEST CHAMBERS TESTA 300 - PERFORMANCES, DIMENSIONS AND DRAWINGS TESTA TT / TESTA CT 300 EXTERNAL DIMENSIONS (HxWxD) (mm) INTERNAL DIMENSIONS (HxWxD) (mm) Standard refrigeration system in -52 ºC models is air cooled; water cooling Supply power TESTA 300 TT/CT 75: is optional. For -75 ºC models, see point 4. Services hub installation needs: Electrical protection: Circuit breaker 3 x 32A + N with 300mA differential ¾” demineralized water supply 3-Phase electrical cable RV-K 5G10 on the top RJ45 communications port 20mm water drain at floor level female connection Water cooled...
Open the catalog to page 6TESTA CHAMBERS PERFORMANCE units PERFORMANCE IN TEMPERATURE TESTING Temperature range in Space @ low temp. point °C in Space @ high temp point °C Temperature fluctuation in time °C Temperature change rate (2a) (2b) PERFORMANCE IN HUMIDITY TESTING Humidity range Fluctuation in time DIMENSIONS Test space volume number of shelves included (more can be added) maximum weight load per shelf Entry ports Included as standard (more can be added) Diameter (other diameters available) Weight (approximately) POWER & REFRIGERATION Supply voltage Nominal Power Type of Refrigeration (3c) (air or water cooled...
Open the catalog to page 7TESTA - ENVIRONMENTAL TEST CHAMBERS EXTERNAL DIMENSIONS (HxWxD) (mm)INTERNAL DIMENSIONS (HxWxD) (mm) 1. Standard refrigeration system in -52 °C models is air cooled; water cooling is optional. For -75 °C models, see point 4. 2. Services hub installation needs: 0 %” demineralized water supply Condutivity: <50pS/cm , TDS <35PPM # 20mm water drain at floor level female connection 3. # Electrical cabinet installation needs: Supply power TESTA 1000 TT/CT 50: Electrical protection: Circuit breaker 3 x 32A + N with 300mA differential Supply power TESTA 1000 TT/CT 75 and 10K model: 400VAC, 50Hz, 50A...
Open the catalog to page 8TESTA CHAMBERS PERFORMANCE Testa TT 1.000 Testa CT 1.000 Testa TT 1.000 Testa CT 1.000 Testa TT 1.000 Testa CT 1.000 PERFORMANCE IN TEMPERATURE TESTING Temperature range in Space @ low temp. point °C ± 0,7 ± 0,7 ± 0,7 ± 0,7 ± 1,2 ± 1,2 in Space @ high temp point °C ± 1,4 ± 1,5 ± 1,5 ± 1,5 ± 1,5 ± 1,5 Temperature fluctuation in time °C ± 0,1°C to ± 0,3°C ± 0,1°C to ± 0,3°C PERFORMANCE IN HUMIDITY TESTING Humidity range Test space volume number of shelves included (more can be added) maximum weight load per shelf Entry ports Included as standard (more can be added) Diameter (other diameters available)...
Open the catalog to page 9EQUIPMENT DESCRIPTION TEMPERATURETEMPERATURE SENSORS • One (1) PT 100 Class A • One (1) PT 100 Class A, movable sensors for flexible placing inside chamber • By stainless steel electric heaters located in the air treatment tunnel • Air cooled hermetic compressor group (low noise and high efficiency) with enforced ventilation and without CFC's. Water-cooled condensers are also available as standard in -75°C models or an option for models with temperature cooling rate upgrades. • Safety thermostat with High / Low temperature configuration, with automatic stop of all thermic systems. • High / Low...
Open the catalog to page 10SECURITY Automatic stop function in case of water failure, with indication on the controller; High / Low Temperature alarms; High / Low humidity alarms. CONSTRUCTION • Interior: AISI 304 hermetical welded, vapour tight, stainless steel • Exterior: Zinc mild steel with epoxy coating finish (color RAL 7035) • Insulation: Rock Wool • Interior illumination: Included with Optional Observation Window • Door: Double silicone joints and anti-condensation heating frames (optional window) AIR FLOW / VENTILATION • Air Flow: Forced through ventilators/fans (300 model has one ventilator/fan, 1.000 model has...
Open the catalog to page 11• Programmable PLC exclusively developed for Test chambers. • Easy to use coloured Touch-Screen Display Interface. • Resolution of 0,1DC for Temperature and 0,1% for Relative Humidity. • High performance temperature and humidity control with value correction in all ranges. • Capability for creating unlimited programs and segments. • Internal non volatile memory for storing test data. • Automatic restart of tests due to power failure, without losing data and restarting test where it was interrupted. • Real-time monitoring of all functions and control of equipment. • Manage control settings via...