Manual Disinfection Procedure:
  • Remove coarse contaminants from instruments and disassemble them as needed.
  • Action time begins when the last instrument is immersed.
  • Follow dosage and action time directions precisely.
  • Exercise caution when handling sharp instruments.
  • Wear protective gloves during disinfection.
  • 5 liters of concentrate can produce up to 500 liters of working solution.
  • Includes corrosion protection and tested material compatibility.
  • Effective against limited viruses (e.g., HBV/HCV/HIV, Herpes-simplex-Virus Type I), bacteria, fungi, and tuberculosis.
  • Suitable for all common instruments and burs; can be used in ultrasonic baths.
  • Recommended by Karl Storz Endoscopes and NTI Kahla.
  • Listed by DGHM/VAH, CE 0408 certified.
Product Information:
  • GREEN&CLEAN ID N: Concentrate for the disinfection of instruments and drills.
  • Dosage for 3 liters of solution: 30 ml for 1% dilution, 60 ml for 2% dilution, 225 ml for 7.5% dilution.
  • Fill tray with 3 liters of cold water and add the concentrate.
  • Immerse instruments completely, ensuring no air bubbles.
  • Cover the disinfection tray and leave for the specified time: 1% for 60 min, 2% for 15 min, 7.5% for 5 min.
  • Rinse instruments with water and dry them.
  • Recommendation: Use a 1% dilution with a 60-minute action time for manual disinfection.
See more

Catalog excerpts

GREEN&CLEAN ID N Disinfection of instruments-1

Dose the concentrate, e.g. for 3 litres 30 ml: 1% dilution 60 ml: 2% dilution 225 ml: 7,5% dilution Fill tray with 3 litres of cold water (indicator) and add the concentrate Immerse instruments completely and free of air bubbles into the solution Cover disinfection tray and leave: Rinse instruments with water and dry Manual disinfection procedure: GREEN&CLEAN ID N: Concentrate for the disinfection of instruments and drills Remove coarse contaminants from instruments and disassemble them as needed yy Action time starts when the last instrument is immersed yy Exactly follow dosage and action time directions yy Be careful when working with sharp instruments yy Wear protective gloves when disinfecting 5 l of concentrate result in up to 500 l of working solution* yy with corrosion protection, tested material compatibility yy limited virucidal (e.g. HBV/HCV/HIV, Herpes-simplex-Virus Type I), bactericide/fungicide, tuberculocide yy suitable for all common instruments and burs yy use in ultrasonic baths is also possible* yy recommended by Karl Storz Endoscopes and NTI Kahla yy DGHM/VAH listed, CE 0408 * Please note the manufacturer’s directions regarding action time and dosage! Recommendation: 1% dilution, 60 minutes action time (for manual disinfection).