Overview
NIRS Analyzer Pro is an advanced process analysis system utilizing high-resolution diode array technology for nondestructive analysis of various materials directly in the process line. It supports both reflectance and transmittance modes, offering real-time data for improved process control.

Features and Benefits
  • High-resolution diode array technology ensures accurate and continuous analysis.
  • Factory standardization allows for quick implementation of multiple analyzers.
  • Uptime protection systems reduce maintenance needs.
  • Dedicated sample interfaces enhance accuracy and speed of implementation.
  • Instant measurement across the complete wavelength range for fast-moving samples.
  • Dual lamp technology maintains accuracy with backup lamp activation.
  • Integrated Metrohm calibration tool, ISIcalTM, facilitates easy calibration development.
  • Metrohm standard OPC interface enables integration with local control systems for automatic process regulation.

System Description
The system is housed in a robust cabinet and can be mounted in production areas. It provides precise instrument matching, minimizing implementation efforts and ensuring calibration model transferability. Measurements are displayed in the control room and can be integrated into regulation systems for closed-loop control.

Technical Specifications
  • Temperature range: -5 to 40 °C, with purge up to 65 °C.
  • Pressure: Max 200 PSI, with specific limitations for Varinline access units.
  • Wavelength range: Reflectance (1,100-1,650 nm), Transmittance (850-1,050 nm).
  • Light source: Dual lamp system with a mean time between failures (MTBF) of 17,500 hours.
  • Protection: IP69K for dust and high-pressure water/steam cleaning.
  • Communication: Ethernet, OPC, RINA, FossCareTM.

Standards and Approvals
The NIRS Analyzer Pro complies with several directives including LVD, RoHS, WEEE, ATEX, and REACH. It is developed according to Metrohm ISO 9001 standards.

Ordering Information
  • Various models available including ContactReflection, FiberSystem, and DirectLight/NonContact.
  • Vision Software and certified process reflectance and wavelength standards are available for purchase.
See more

Catalog excerpts

NIRS Analyzer Pro-1

NIRS Analyzer Pro is a process analysis system based on high-resolution diode array technology. It provides nondestructive analysis of granules, powders, slurries or opaque samples directly in the process line without bypass – a true in-line system.

 Open the catalog to page 1
NIRS Analyzer Pro-2

• High resolution diode array technology for accurate and continuous analysis in reflectance or transmittance mode • Built-in instrument factory standardisation for quick and simple implementation of one or several analysers • Uptime protection systems for low maintenance • Dedicated sample interfaces provides accuracy and rapid implementation • Instant measurement of complete wavelength range for direct measurement of fast moving samples • Quantitative and qualitative data for better in-line process control • Dual lamp technology retaining accuracy when the backup lamp is activated providing...

 Open the catalog to page 2
NIRS Analyzer Pro-3

NIRS Analyzer Pro is a process analysis system based on high re­­ lution diode array technology. It provides non so­­ destruc­­ analysis of products such as flour, feeds, tive meals, ground meat and dairy products directly in the process line without bypass. The analyser is housed in a robust cabinet mounted at the relevant location in the production area. Measurements are displayed in the control room and results can be fed into a regulation system for closed-loop automa­­ tic control. The solution helps to optimise the use of raw materials and to consistently run production closer to target...

 Open the catalog to page 3
NIRS Analyzer Pro-4

Dedicated sample interfaces 04 Window reflectance In-line analysis of paste, granulates, powdered products etc in pipes or transport systems without bypass can be performed. The products pass over the interface window. The window reflection interface can easily be installed into the production line using standard GEA Tuchenhagen flowcells or welding an interface flange into the wall of the pipe/transport system. Temperature 150 °C (302 °F) Pressure Vacuum > 1 torr, Pressure < 3’000 PSI Lens Sapphire; Diameter 45 mm, thickness 12 mm, with food grade EPDM O-ring seal Hygiene USDA, Dairy Pipe flowcells...

 Open the catalog to page 4
NIRS Analyzer Pro-5

Powder probe In line analysis of fine powder like dairy powders etc. The powder probe interface can easily be installed into a hopper or pipe with free falling product. The probe has no moving parts. The probe is automatically cleaned with compressed air before each analysis. Material Diameter Length Fiber Air Temperature Installation Hygiene Cleaning Teflon (PCTFE) 1˝ 7˝ Steel armoured fiber bundle (1, 3, 5 or 10 meters) Clean compressed air 43-72 PSI Max 120 °C 1˝ Swagelok crimp fitting USDA, Dairy Clean with compressed air or wash in water, hot or cold depending on product

 Open the catalog to page 5
NIRS Analyzer Pro-6

Lateral transmittance In line analysis of slurries and viscous products such as WPC, Cream Cheese, Mozarella etc. The Lateral transmittance probe does not restrict the flow rate of the product and can easily be installed in the production line using a standard GEA Tuchenhagen flowcell for installation in a pipe or by welding an interface flange into the wall of a tank. Materials Lens Temperature Pressure Hygiene Optical fiber Pipe flowcells Stainless steel Sapphire, 5 mm thick, with food grade EPDM O-ring seal Max 150 °C (302 °F) Max 200 PSI (13.79 bars) – Warning! Varinline access units higher...

 Open the catalog to page 6
NIRS Analyzer Pro-7

Technical Specifications NIRS Analyzer Pro Reflection Analysis time 5 - 50 ms / integration time; Average time per result 3-15 seconds Measurement mode Reflectance (Window reflectance; Direct Light, Powder Probe) Wavelength range 1’100-1’650 nm Detector InGaAs Diode array Spectral dispersion 1.1 nm / pixel NIRS Analyzer Pro Transmittance Analysis time 5 ms-60 sec / integration time; Average time per result 3-15 seconds Measurement mode Transmittance (Lateral transmittance, transmittance probe pair) Wavelength range 850-1’050 nm Detector Si Diode array Spectral dispersion 1.0 nm / pixel General...

 Open the catalog to page 7
NIRS Analyzer Pro-8

NIRS Analyzer PRO – ContactReflection NIRS Analyzer PRO – FiberSystem NIRS Analyzer PRO – DirectLight/NonContact Software 6.6069.102 Vision Software Subject to modifications Layout by Ecknauer+Schoch ASW, printed in Switzerland by Metrohm AG, CH-9100 Herisau 8.928.5000EN – 2013-03 Standards 6.7450.070 NIRS Certified Process Reflectance and Wavelength Standard (1") 6.7450.080 NIRS Certified Process Reflectance and Wavelength Standard (0.5") 6.7450.090 NIRS Certified Process Reflectance and Wavelength Standard (1", 45 degrees) 6.7450.100 NIRS Certified System2 Reflectance Standards (set of 7 pc.)...