1. Catalogs
  2. Metrohm
  3. NIRS DS2500 Analyzer
video corpo

NIRS DS2500 Analyzer

NIRS DS2500 Analyzer

NIRS DS2500 Analyzer

Product catalog summary
Introduction
The NIRS DS2500 Analyzer represents a significant advancement in near-infrared (NIR) analysis, offering exceptional accuracy across a broad wavelength range of 400 to 2,500 nm. It is designed for use in the receiving area or at-line in the nutraceutical industry, ideal for rapid identification of raw materials, routine production control, and final product monitoring.
Key Features
  • Dedicated reflectance and interactance analyzer with high optical performance.
  • Consistent results in harsh environments and suitable for LAN networking.
  • Equipped with a new predispersive monochromator.
  • Optional sampling accessories available.
Technology
The NIRS DS2500 uses a predispersive monochromator for unmatched optical performance, ensuring versatility and stability across the full spectral range. It provides high accuracy with a high signal-to-noise ratio and includes internal standards for light intensity, bandwidth, and wavelength position control. The device is designed for minimal recalibration needs, with options for automatic recalibration and quality control.
Calibration Compatibility
The analyzer is backwards compatible with System II and Metrohm XDS analyzers, allowing easy data migration without performance loss.
Software
The Metrohm DS2500 operates on Vision software, supporting the latest calibration technologies with features like automatic database storage, supported regression methods (PLS and MLR), real-time analysis, and LIMS compatibility.
Durability
Built for high performance in harsh conditions, the NIRS DS2500 is robust, easy to use, IP65 certified, and compliant with pharma regulations, making it suitable for at-line use in production plants.
Support Program
Metrohm offers a comprehensive support program including on-site maintenance, software updates, calibration support, and training.
Legal Compliance
The equipment complies with several directives including EMC, LVD, RoHS, and WEEE.
Technical Specifications
  • Dimensions: 14.8 × 19.3 × 11.8 inches; Weight: 27 kg; Degree of protection: IP 65.
  • Measurement mode: Reflectance or transflectance; Wavelength range: 400-2,500 nm.
  • Detector: Silicon (400-1,100 nm), Lead Sulfide (1,100-2,500 nm).
  • Optical bandwidth: 8.75 ± 0.1 nm; Spectral resolution: 0.5 nm.
  • Analysis time: < 1 minute; Wavelength accuracy: < 0.05 nm.
Ordering Information
Includes various software options and certified reflectance standards.
See more

Catalog excerpts

NIRS DS2500 Analyzer-1

The next step in NIR analysis

 Open the catalog to page 1
NIRS DS2500 Analyzer-2

Near infrared analysis just keeps on getting better. With the NIRS DS2500, near infrared analysis has taken yet another step into the future of routine analysis. This de­ ­ dicated analyzer offers exceptional accuracy across a broad wavelength range of 400 to 2’500 nm and is de­ ­ signed for use in the receiving area or at-line of nutraceu­ tical industry the NIRS DS2500 is ideal for: • Rapid identification of incoming raw materials • Routine production control for improved efficiency and cost savings • Final product monitoring NIRS DS2500 Analyzer key features • Dedicated reflectance and interactance...

 Open the catalog to page 2
NIRS DS2500 Analyzer-3

Technology Unmatched optical performance The new predispersive monochromator technology used in the NIRS DS2500 ensures versatility and stability across the full spectral range from 400 to 2’500 nm. In less than a minute, with the highest possible signal-to-noise ratio, the NIRS DS2500 can effortlessly manage analysis of de­ ­ manding and low level parameters in need of high accu­ racy. With this high performing instrument on your team, you can be sure that you are getting the highest level of quality control at all stages of production. Part of a complete high performing solution Whether you...

 Open the catalog to page 3
NIRS DS2500 Analyzer-4

Specifications Feature Specification Dimensions (W × D × H) 14.8 × 19.3 × 11.8 inches (375 × 490 × 300 mm) Weight 27 kg Degree of protection IP 65 Measurement mode Reflectance or transflectance (for liquids) Wavelength range 400-2’500 nm Silicon (400-1’100 nm), Detector Lead Sulfide (1’100-2’500 nm) Optical bandwidth 8.75 ± 0.1 nm Spectral resolution 0.5 nm Number of data points 4’200 Absorbance range Up to 2 AU Analysis time < 1 minute* Wavelength accuracy < 0.05 nm Wavelength precision < 0.005 nm (Based on a single analyser) Wavelength precision instrument-to-instrument < 0.02 nm (Based on...