Introduction
ZONARE Transducers are designed for use with ZONARE Ultrasound Systems. This document provides guidelines for cleaning and disinfecting these transducers to ensure safety and functionality.

Intended Use
The transducers are intended for medical use with ZONARE Ultrasound Systems. Users should refer to the specific system manuals for detailed intended use.

Cautions and Warnings
- Disinfectants are recommended based on chemical compatibility, not biological effectiveness.
- Do not immerse transducers longer than 45 minutes to avoid damage.
- Avoid using non-recommended cleaning solutions or incorrect solution strengths.
- Never immerse beyond 10 mm below the top edge of the transducer housing.
- Always check for damage before use and contact ZONARE if any is found.
- Disconnect transducers before cleaning and avoid electrical shock by turning off the system.

General Cleaning
Transducers are not pre-cleaned or disinfected. Users must clean and disinfect them before use, considering the level of disinfection required based on intended use and contamination risk.

Cleaning & Disinfecting Methods
Wipe Method:
1. Disconnect and remove any covers.
2. Clean with a soft cloth dampened with mild soap and water.
3. Inspect for damage.
4. Use a recommended disinfectant solution.
5. Air-dry or towel dry.

Immersion Method:
1. Disconnect and remove any covers.
2. Clean with mild soap and water.
3. Inspect for damage.
4. Mix disinfectant as per instructions and immerse transducer.
5. Wipe with soap and water, then rinse with fresh water.
6. Air-dry or towel dry.

High-Level Disinfection
Follow similar steps as above, ensuring to use high-level disinfectants approved by ZONARE.

Approved Disinfectants
Various immersion and wipe solutions are approved for different transducers, including CIDEX, Anioxyide 1000, and T-SPRAY. Specific solutions are recommended for the P8-3TEE transducer.

Additional Systems
The Nanosonics Trophon® EPR System is approved for high-level disinfection of the E9-4 Transducer with specific software versions.

Technical Support
Contact information for technical support in North America, Asia, and Europe is provided for further assistance.
See more

Catalog excerpts

Transducer_Cleaning_&_Disinfection-1

ZONARE Transducers ZONARE ZONARE Transducers Cleaning & Disinfection Cleaning & Disinfection of ZONARE Transducers Q00066 Rev H

 Open the catalog to page 1
Transducer_Cleaning_&_Disinfection-2

ZONARE Transducers ZONARE Medical Systems Inc. 420 N. Bernardo Avenue Mountain View, CA 94043 USA MedPass International Ltd. Windsor House Barnett Way, Barnwood Gloucester GL4 3RT United Kingdom Technical Support North America and Asia: Phone support: 877-913-9663 or 650-316-3199 Email: [email protected] Sales support: 1-877-966-2731, [email protected] www.zonare.com Europe: Phone support: +49 (913) 1974-9412 Email: [email protected] www.zonare.de CAUTION: Federal Law restricts this device to sale by or on the order of a licensed healthcare practitioner (USA). Part Number Q00066-00F...

 Open the catalog to page 2
Transducer_Cleaning_&_Disinfection-3

ZONARE Transducers Draws attention to information that may relate to safety of the patient, the operator, or the equipment Waste Electrical & Electronic Equipment Standard. ZONARE meets the WEEE Standard. For more information, contact ZONARE Inc. Refers to a type BF patient-applied part (B= body, F= floating applied part) Refers to a type CF patient-applied part (C= cardiac, F= floating applied part) Consult Operator Manual Authorized Representative in the European Community Catalog Number of Item Shipping: Temperature Limits Caution: Federal law restricts this device to sale by or on the order...

 Open the catalog to page 3
Transducer_Cleaning_&_Disinfection-4

ZONARE Transducers Refer to the ZONARE Ultrasound System Safety Manual, Operator’s Manual or Instructions for Use for a description of the system including compatible transducers and operating instructions. Intended Use ZONARE Transducers are intended for use with ZONARE’s Ultrasound Systems. Refer to the corresponding ZONARE Ultrasound System Manual(s) for the Intended Use. Cautions and Warnings Refer to the corresponding ZONARE Ultrasound System Manual(s) or Instructions for Use for a complete list of Warnings and Cautions associated with the use of the ZONARE Transducers and Ultrasound System....

 Open the catalog to page 4
Transducer_Cleaning_&_Disinfection-5

ZONARE Transducers Cleaning & Disinfecting Note: The instructions below should be used in conjunction with the Cleaning and/or Disinfection agent manufacturer’s instruction for use. These provide guidelines to ensure ZONARE transducers are not damaged during the cleaning and/or disinfection process. Wipe Method 1. 2. 3. 4. Disconnect the transducer from the Cart or Scan Engine/Scan Module. If present, remove the transducer cover/sheath and discard. Use a soft cloth dampened with a mild soap (Liquinox) and water solution to remove particulate, ultrasound coupling gel, or body fluids. Examine the...

 Open the catalog to page 5
Transducer_Cleaning_&_Disinfection-6

ZONARE Transducers Figure 1: Immersion Level Limits for Transducers Disconnect the transducer from the Cart or Scan Engine/Scan Module. If present, remove the transducer cover/sheath and discard. Use a soft cloth dampened with a mild soap (Liquinox) and water solution to remove particulate, ultrasound coupling gel, or body fluids. Examine the transducer for damage, such as cracks, holes, or fluid leaks. If damage is evident discontinue use of the transducer and contact ZONARE Technical Support. Follow disinfectant manufacturer’s instructions. Cleaning & Disinfection of ZONARE Transducers Q00066...

 Open the catalog to page 6
Transducer_Cleaning_&_Disinfection-7

ZONARE Transducers ZONARE-Approved Transducer Disinfectants NOTE: All transducers may be immersed except for AUX CW probes. The following disinfectants are approved for all ZONARE transducers except the P8-3TEE: Immersion Solutions CIDEX CIDEX Plus Wipe Solutions T-SPRAY Sani-Cloth PLUS Disinfectants Approved for P8-3TEE Transducer Immersion Solutions Cidex OPA (non glutaraldehyde) Cidex Plus® Metricide Plus 30® Steranios 2% Gigasept FF Gigasept PPA Wipe Solutions T-Spray Enzol/ Cidezyme Additional High-Level Disinfection System for E9-4 Transducer ONLY: ® Nanosonics Trophon EPR System ® Nanosonics’...

 Open the catalog to page 7

All Mindray North America catalogs and technical brochures

  1. BC-5390

    4  Pages

  2. A8/A9 Brochure

    12  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.