TOF-1

Eyes that never miss a twitch objective neuromuscular transmission monitor

 Open the catalog to page 1
TOF-2

The TOF-Watch—more effective than subjective monitoring at detecting residual blockade Patients monitored objectively with a TOF-Watch experienced a signicantly lower incidence of residual blockade (train-of-four [TOF] ≤0.9) when compared with patients monitored subjectively 1,2 by clinical exam alone or with a peripheral nerve stimulator lacking an objective display. Incidence of residual blockade in PACU*1 40 4.5% monitored with the TOF-Watch monitored with a peripheral nerve stimulator * Postanesthesia care unit. The limitations of subjective monitoring Research has also shown that visual...

 Open the catalog to page 2
TOF-3

Acceleromyography—the most accurate and reliable method of objectively measuring neuromuscular blockade 4 Acceleromyography (AMG) is a technique used by the TOF-Watch for recording evoked muscle responses. The TOF-Watch performs AMG by using a small piezoelectric transducer to convert measured acceleration into electrical signals, which are then processed and presented as clear and accurate measures 4,5 of neuromuscular transmission. The advantages of the TOF-Watch for performing AMG include4,6: Accuracy Simplicity Suitability for routine use during surgery and in the PACU A versatile partner...

 Open the catalog to page 3
TOF-4

What can your TOF-Watch do for you? The TOF-Watch series consists of 3 monitors: TOF-Watch, TOF-Watch® S, and TOF-Watch® SX. All 3 monitors accurately measure the depth of neuromuscular blockade to help assess1,8: Onset time of neuromuscular blocking agents (NMBAs) The need for the administration of maintenance doses of NMBAs When to administer a reversal agent When to safely extubate While all 3 monitors work to help ensure your patient maintains the appropriate level of blockade during surgery and reaches a TOF ratio ≥0.9 postsurgery, each TOF-Watch model differs in the total features provided.1...

 Open the catalog to page 4
TOF-5

TOF-Watch SX The most sophisticated instrument in the TOF-Watch series, the TOF-Watch SX incorporates all the features offered by the TOF-Watch and TOF-Watch S along with added features such as a surface sensor for measuring skin temperature and the ability to connect with Philips Monitors. ® TOF-Watch SX Monitor Software The TOF-Watch SX is also compatible with monitor software that can be purchased separately and downloaded to a desktop or laptop computer. This software can be used to easily add comments to a patient’s measured data recordings. These comments can later be edited, with changes...

 Open the catalog to page 5
TOF-6

Key features of the TOF-Watch TOF-Watch Stimulation patterns Train-of-four (TOF) Posttetanic count (PTC) Single twitch (ST) 1 or 0.1 Hz Double-burst stimulation (DBS) 3.3 or 3.2 Tetanus (TET) 50 or 100 Hz Slow TOF (TOFs) programmable from 1 to 60 min Stimulation current (0-60 mA at impedance ≤5 kOhm) Stimulation pulse width monophasic 200 µs Stimulation pulse width monophasic 300 µs Calibration with user-set current Calibration with automatically set supramaximal current Manual transducer sensitivity setting User-programmable upper and lower TOF/TOFs alarms (OFF, count, or % TOF) Automatic power...

 Open the catalog to page 6
TOF-7

TOF ratio ≥0.9—the gold standard Clinical studies are frequently dening recovery from neuromuscular blockade as a TOF ratio ≥0.9, due to an increased recognition of the risks associated with a TOF ratio <0.9.10-14 Hypoxemia A study using pancuronium showed that 60% of patients with a TOF ratio <0.7 experienced hypoxemia versus 10% of patients with a TOF ratio ≥0.710 A separate study showed that patients with a TOF ratio <0.9 were signicantly more likely to develop postoperative hypoxemia than those with a TOF ratio >0.9 (24 of 39 vs 7 of 30, P=0.003)11 Pharyngeal dysfunction In a study using...

 Open the catalog to page 7
TOF-8

Never miss a twitch with the TOF-Watch The TOF-Watch series offers an accurate and reliable method of objective acceleromyographic monitoring of neuromuscular blockade that overcomes the limitations of subjective monitoring.4 Reliable detection of residual blockade 4.5% incidence in the PACU for patients monitored with the TOF-Watch, versus 30% incidence for patients monitored subjectively 1,2 Avoids the risks associated with a TOF <0.9 Recovery to a TOF <0.9 has been associated with a wide range of side effects10-14 The TOF-Watch helps ensure patients are recovered to a TOF >0.9 before extubation1...

 Open the catalog to page 8

All MIPM Mammendorfer Institut für Physik und Medizin catalogs and technical brochures

  1. (CPMS)

    2  Pages

  2. Tesla DUO

    2  Pages

  3. TeslaM3

    4  Pages

  4. MRI-BOSS

    4  Pages

  5. Tesla ONE

    2  Pages

  6. Tesla NIBP

    2  Pages

  7. Tesla Guard

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.