alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -1

Alexandrite Long Pulse Laser 755nm

 Open the catalog to page 1
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -4

www.miramhoo.com Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO 1. Appearance & Components 1. Appearance & Components (Noted: This photo is only considered as a reference; please reply with the real laser which may has differences at some details from the photo showed above. ) 1. Caster: to support and transport the equipment. 2. Handrail: 3. Right panel. 4. Monitor:

 Open the catalog to page 4
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -5

Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO to catch the headpiece. to conduct laser energy. to hold t the wires and cooling tubes. 11. Front panel. 12. Finger switch.

 Open the catalog to page 5
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -6

www.miramhoo.com Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO 2. Packing list Headpiece with cold wind cooling, spot-size 6-18mm

 Open the catalog to page 6
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -7

www.miramhoo.com Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO 2. Safety This article shows the details of safety when operation. The operator and technician must be certified after professional training of safety, know well the requirements of safety and steps to operate the laser. 1. Electric safety 1. Make sure the voltage of laser requiring will be adapted by local electricity supply. 2. 10 minutes after switching off and plug off the cable, only the maintenance can be started. No leave the laser without anyone taking care of by power on, or else injure will be happened to human or laser...

 Open the catalog to page 7
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -8

www.miramhoo.com Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO Water tank label Caution, no remove the covers.

 Open the catalog to page 8
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -9

www.miramhoo.com Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO 3. Specifications 1. Machine specification

 Open the catalog to page 9
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -10

www.miramhoo.com Alexandrite Long Pulse Laser 755nm-GENTAL MAX PRO Headpiece Candela Gentlemax Pro Control Monitor 10 inches Digital Real Color LCD Control system Single chip Headpiece Headpiece/spotsize 6-18mm spotsize with cold air coolilng Cooling Cooling system Water cooling water temperature Skin cooling internal air cooling machine optional cold air cooler Pulse width Pulse frequency Power system Power Capacitance capacity Laser board Lamp Fiber Size & Weight Size

 Open the catalog to page 10
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -11

1. Air quality: The system should work in a non-corrosive environment. Corrosive substances, such as acids, etc., may damage the surface of the wires and electronic equipment. Dust in the air should be kept to the minimum. Metal particles have a corrosive effect on the electronic device. 2. Temperature and humidity Working temperature: 10-30oC keep temperature: Working Relative Humidity:max 80% Storage relative humidity: Max 90% 3. Atmospheric pressure: 860hPA—1060hPA 4. Environmental requirements: Avoid the sunshine

 Open the catalog to page 11
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -12

1. Fiber support arm installation 1. Find the fiber support arm. 2. Fix arm tightly with screws. 3. Adjust the angle and screw it.

 Open the catalog to page 12
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -13

2. Fix fiber tightly with screw. insert into the fiber port 7. Fix fiber tightly with screw. 3. Prepare the finger switch and insert into the port. 8. Prepare the cooling tube and insert into the port. 9. Insert into the port deeply.

 Open the catalog to page 13
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -14

1. Find the water from water inlet 4.Connect foot-pedal and electric plug. 3. Open the overfall at the same time

 Open the catalog to page 14
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -15

1. Prepare the electric plug. 4.2. Water-cooling system Due to the generated huge heat of the xenon lamp when do treatment, and the treatment handle cooling the skin produce a lot of heat, so the machine needs to add a certain amount of pure water before work. Water added standard : Distilled water or purified water ( avoid to use tap water or mineral water, otherwise it will seriously affect the working life of the machine and the cooling effect.) 1. 2. Water tank capacity:15L. the shortest period of replace cooling water: three months. Cooling water filling (see 4.1.3 step): a. Water monitoring...

 Open the catalog to page 15
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -16

operating room should install air-conditioning equipment to ensure the temperature about 25oC before using the machine; otherwise the water temperature will be a sharp rise in a short time, which will not all shortens the working time of the machine , but also affect the service life of the machine. d. Water quality monitoring. Machine Built-in cooling water quality monitoring module. if the water is impurities, and water quality is bad, the machine will no longer monitor the water flow, suggesting that "water is not circulating." It should replace the cooling water in this time. e. water level...

 Open the catalog to page 16
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -17

5. Monitor Operations Guide This chapter describe the treatment handle of Alex. And Nd Yag laser system, the proposed treatment parameters and operation method. Please make sure to read in detail before using the product. 1. System operation 1.1. Start the machine。 1. Make sure the power cord is connected, insert the key, screw to the right, the start device inside the machine will start the machine, the user interface will light. 1. At this page, please select the wavelength related for treatment. YAG is 1064nm, Alex is 755nm. Now we click “YAG” for example.

 Open the catalog to page 17
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -18

4. The below page shows the main operating page for parameters adjustment. ( Figure 5.1.4).

 Open the catalog to page 18
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -19

a. press this button back to wavelength selection page. b. Press this button for applications reference, The page will show the page as below, At this page, the YAG Applications can be selected by, Hair Removal, Diffuse Redness, Vascular Lesions, and Facial Treatment. Or press page. c. Energy selection, At this page, the energy can be selected from

 Open the catalog to page 19
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -20

d. Pulse width selection, By this page, the pulse width can be selected from 10-100ms. e. Frequency selection, By this page, the frequency can be selected from 0.5-2.0HZ. f. By this page, the spot-size can be selected by 6mm(diameter), 8mm, 10mm, 12mm, 15mm, 18mm.you can adjust the the spot size directily on the handle. Press 5 seconds to enter into the engineer mode by password. Cold air blowing Pulse duration and Pulse interval selection. The pulse duration and pulse interval can be selected by 0ms, 50ms, 100ms, 500ms, 1000ms. If Pulse duration =1000ms, and Pulse Interval=0ms, it means the...

 Open the catalog to page 20
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -21

Press to shift wavelength from YAG to Alex. 4. “Alex” page. The alex page shows following,

 Open the catalog to page 21
alexandrite 755 nm nd yag 1064 nm hair removal Gentle Alex -22

The alex same as YAG. So here we do not explain again for details. 2. Treatment parameter Check the “Application” page. 3. Treatment progress Take the755nm wavelength as an example. 1064nm wavelength is the same, which will not be illustrated here again. The steps are as following: 1. Make sure the machine is installed complete; 2. Make sure the water is full filled; 3. Position the machine correctly beside the treatment table; clean the treatment area; 4. Start the machine; 5. Enter the welcome page; 6. Enter the Alex page; 7. Set up the parameters as required; 8. Press “standby” key to ready...

 Open the catalog to page 22

All Miramhoo Aesthetics catalogs and technical brochures

  1. Morpheus8 RF-2

    32  Pages

  2. DL4

    6  Pages

  3. Miramhoo-HRL2

    6  Pages

  4. Miramhoo-DL5

    6  Pages

  5. Miramhoo-DL4

    6  Pages

  6. 22D HIFU-1

    9  Pages

  7. Shockwave ESWT

    19  Pages

  8. Skin analysis-1

    11  Pages

  9. Miramhoo-HL2

    5  Pages

  10. truSculpt ID

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.