Chapter 1 - Operation Safety Precautions
Before using the CL10H Cold Air Skin Cooling System, familiarize yourself with the instructions to ensure safe and correct operation. Key precautions include:
  • Allow the equipment to stand upright for over 3 hours before use.
  • Ensure stable placement to minimize vibration.
  • Avoid tilting the device more than 30 degrees during handling.
Chapter 2 - Installation
Product Theory: The system uses a frequency conversion air compressor to deliver cold air, protecting the epidermis during laser treatments.
Function: It cools the skin before, during, and after laser treatments, reducing pain and preventing burns. Compatible with various laser machines.
Advantages:
  • Continuous cooling minimizes pain and thermal injury.
  • Adjustable wind speed and cooling time for patient comfort.
  • Cooling level can reach -20°C, allowing high-power laser use.
  • Features a 7-inch touch screen for easy operation.
Specifications:
  • Power: 220V/110V, 500W
  • Fan speed: 0-0.5m³/min, 9-speed settings
  • Noise: <60db
  • Temperature range: -20°C
  • Working time: 60-120 minutes adjustable
  • Net weight: 26kg
  • Dimensions: 320x450x350 mm
Chapter 3 - Operational Guidelines
Control Panel: Features a touch screen for adjusting settings.
Installation: Follow the provided diagram for setup.
Boot-up Process: Start the cold fan, adjust the air volume, and wait for the temperature to reach -17°C before use.
Adjust Wind Speed: Use the display screen to adjust up to 9 gears.
Defrosting: Initiate defrosting after 2 hours of continuous use when necessary.
Water Discharge: Release water from the storage tank when prompted by an alarm.
Chapter 4 - Use Precautions
  • Avoid prolonged exposure to the same skin area to prevent frostbite.
  • Maintain a distance of at least 8 cm from the skin.
  • Release water before transport and avoid tilting over 30 degrees.
  • Do not clean the panel with alcohol.
  • Allow a 3-minute interval between power cycles.
Chapter 5 - After-sales Service and Warranty Period
The product comes with a 12-month warranty in mainland China. Free repairs are provided for manufacturing defects. However, misuse, exceeding specifications, or natural disasters are not covered.
See more

Catalog excerpts

Cryo Zimmer skin Cooler-2-1

Cold air skin cooling system MODEL:CL10H

 Open the catalog to page 1
Cryo Zimmer skin Cooler-2-2

Chapter 1 - Operation Safety Precautions Before power-on, operation and maintenance inspection, it is necessary to familiarize yourself with the contents of the instructions in order to ensure correct use. When using, you must also be familiar with your equipment and all relevant safety precautions. ! Matters needing attention  After receiving the equipment, it should be placed upright for more than 3 hours before booting up  Place smoothly to reduce vibration.  Handling should prevent the fuselage from tilting more than 30 degrees, let alone horizontal or inverted. Chapter 2 - Installation...

 Open the catalog to page 2
Cryo Zimmer skin Cooler-2-3

2、Function The professional skin cooling machine is specially for cooling skin before, during and after using of non-invasive CO fractional laser, Q switch laser, IPL or diode laser treatment; pain release for patients to avoid any burn or uncomfortable treatment; Also efficient pain treatment, swelling reduction and muscle relaxation. It Can work with 808nm diode laser hair removal machine, co2 fractional laser machine, SHR machine, IPL machine and Nd yag laser machine...Suitable for all machines if patients feel pain and needs cooling. 3、Advantages 1. It can produce continuous cooling wind...

 Open the catalog to page 3
Cryo Zimmer skin Cooler-2-4

Laser source Power requirement Temperature range (ambient temperature 25℃,humidity 30%) Working Time: Dimension of system 320*450*350 mm Package size

 Open the catalog to page 4
Cryo Zimmer skin Cooler-2-7

After the installation is completed, the following figure is as follows 3-3 Boot-up proces Press the cold fan start button, the thermometer shows that the temperature gradually decreases, the air volume knob is selected to the position of 1gear, the frequency conversion air compressor starts, the temperature meter of the cold fan shows that the temperature continues to decrease, when the temperature reaches - 17 degrees 5-10 minutes, the cold fan can be used normally, according to the customer's tolerance to cold air and the need for cosmetic surgery to select the appropriate wind speed.

 Open the catalog to page 7
Cryo Zimmer skin Cooler-2-8

3-4 Adjust wind speed Adjust through the air volume adjustment button on the display screen, ,up to 9 gears it is in the fourth to 6-8 gear。The higher the gear, the larger the air volume, the lower the cooling feeling of customers. 3-4 Defrosting When the air cooler works for about 2 hours continuously (the humidity of air may change for a long time), it can defrost when the temperature is higher than - 35 degrees without changing the gear of the air cooler. Press the defrosting button, the air cooler will enter the defrosting state, and will no longer refrigerate. When defrosting, it is necessary...

 Open the catalog to page 8
Cryo Zimmer skin Cooler-2-9

Chapter 4.Use precautions 1. Do not blow the same part for a long time to avoid frostbite. 2. The height of the tuyere from the skin should be above 8 cm to prevent frostbite. 3. When transporting, water must be released, and the inclination angle is not more than 30 degrees. 4. Do not use alcohol to scrub the panel to prevent the handwriting from falling off. 5. Don't switch on frequently. The time interval between two boots should not be less than 3 minutes. Chapter 5. After-sales Service and Warranty Period Thank you very much for purchasing CL10 skin air cooler manufactured by Beijing Yirmei...

 Open the catalog to page 9
Cryo Zimmer skin Cooler-2-10

(1)When the use condition is not good, please read and check the instructions again. (2)In case of malfunction, please contact the seller or the after-sales service of our company recorded in the instructions. (3)During the warranty period, we will repair the faults caused by our company's manufacturing problems free of charge. 2、Defects Liability Period he warranty period is 12 months in mainland China. If normal use fails during the warranty period, our company will repair and replace damaged parts free of charge. However, if the failure is caused by the following reasons, repairs will be paid...

 Open the catalog to page 10

All Miramhoo Aesthetics catalogs and technical brochures

  1. Morpheus8 RF-2

    32  Pages

  2. DL4

    6  Pages

  3. Miramhoo-HRL2

    6  Pages

  4. Miramhoo-DL5

    6  Pages

  5. Miramhoo-DL4

    6  Pages

  6. 22D HIFU-1

    9  Pages

  7. Shockwave ESWT

    19  Pages

  8. Skin analysis-1

    11  Pages

  9. Miramhoo-HL2

    5  Pages

  10. truSculpt ID

    6  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.