Formomat PL
6Pages

{{requestButtons}}

Catalog excerpts

Formomat PL - 1

Formomat PL Reliable and economical. • Reproducible and validatable sterilization cycles. • Perfect sterilization of temperature sensitive items even with long lumens. • Convincingly economical sterilization method in compliance with European Standard EN 14180 MMM. Protecting human health.

Open the catalog to page 1
Formomat PL - 2

Formomat PL Reliable and economical. Convincing technology. Every thing has its form and every thing needs its time. Particularly heat sensitive instruments with specific sizes and design need highly qualified processing ensuring gentle care of instruments and perfect sterilization at the same time – also for items with long lumens. The new MMM sterilizer series of Formomat features an optimized LTSF-method including economical, technological and organizational advantages. Examples: variable temperature process (55 - 75 °C), simple operation thanks to a new design of the MMM panel, clearly...

Open the catalog to page 2
Formomat PL - 3

The new MMM Formomat features a low temperature LTSF sterilization method at 55°C, 60°C or 75°C which is suitable for all kinds of temperature sensitive objects. Even extremely long and narrow lumens can be safely sterilized inside. The MMM Formomat has been officially certified to sterilize lumens of 1.5 m length with inner diameter of 2 mm by standard cycles, even at 55°C. Biological testing is made with the same indicators as used in steam sterilization i.e. Bac. stearothermophilus. Due to its unique construction including external chamber and door heating by electrical heating mats,...

Open the catalog to page 3
Formomat PL - 4

Powerful features: • Approved sterilization method according to latest European standards for LTSF sterilization EN 14180. • Fully automatic cycle run. Cycle documentation by integrated printer. Optional cycle documentation via PC and A4 print-out. • Low sterilization temperatures for heat sensitive instruments, as of 55°C, 60°C and optional 75°C. • All cycles suitable for long and narrow lumens. Standardized PTFE test device with 1.5 m length, inner diameter 2 mm. • Biological testing with Bac. stearothermophilus indicators. • Even temperature and media distribution due to chamber and door...

Open the catalog to page 4
Formomat PL - 5

Fully developed details. CSSD integration: Program configuration: As the Formomat sterilizer meets highest safety standards, it can be integrated into a CSSD without any problem, which is a further advantage: optically uniform sterilizer front, easy to clean and easily integrated in a network. Thus a clear separation of the clean and the sterile area. The control system of the Formomat can be configured in such a way that the operation requires user identification first. To add further process safety, any illegal manipulation of parameters is prevented by access protection. Flow evaporator:...

Open the catalog to page 5
Formomat PL - 6

MMM Münchener Medizin Mechanik GmbH Semmelweisstraße 6 D-82152 Planegg / Munich Phone: +49 89 89918-0 Fax: +49 89 89918-118 www.mmmgroup.com we manufacture products that meet the highest demands of our customers world wide. The depth and precision of production standards at both plants ensure that we accomplish the rigorous quality requirements of medical engineering. 900 competent employees work together as a committed and enthusiastic team, dedicated to achieving the mission of the MMM Group. Protecting human health. As one of the world’s leading suppliers of sterile processing systems,...

Open the catalog to page 6