Overview
The Quick Ship Provider Plus® 6540-AC is a UL-listed, surface-mounted, horizontal modular headwall designed for rapid capacity expansion in acute care settings. It features integral horizontal channels with a clear anodized aluminum finish, facilitating easy cleaning and accommodating various hospital accessories.

Unpacking and Inspection
Inspect shipping containers for damage upon receipt and document any issues. Unpack and check units for defects before installation, reporting any problems to Modular Services immediately.

Preparation
Review approved shop drawings and submittal booklet for technical details, including wiring diagrams, equipment types, and room locations.

Unit Installation
Remove top and bottom covers by unscrewing and rotating them away from the unit. Attach hanging brackets 9.8 inches below the device console centerline using wall anchors. Position the unit over the brackets and secure it by drilling through the top extrusion and fastening with anchors.

Electrical Service Connections
Electrical connections can be made on either end of the unit using hospital-grade cables. Knock-outs are provided for field connections.

Care and Maintenance
Clean with mild detergent and warm water, avoiding lubricants or excessive moisture. Disinfect with EPA-approved products and use household cleansers for tough stains. Contact sales representatives for parts or repairs.

Installation Terms and Conditions
Units are pre-wired for normal and emergency use. Modular Services is not liable for delays or consequential damages and provides no warranties beyond those stated.

Warranty
Equipment is warranted to be free from defects for 12 months from acceptance or service start date. The warranty excludes lamps and non-furnished materials and is voided by unauthorized modifications.
See more

Catalog excerpts

Provider Plus® 6540-AC Installation Manual-1

INSTALLATION MANUAL ≈QUICK SHIP PROVIDER PLUS® SURFACE-MOUNTED HEADWALL Model #6540-AC

 Open the catalog to page 1
Provider Plus® 6540-AC Installation Manual-2

OVERVIEW The Quick Ship Provider Plus® 6540-AC is a UL-listed, surface-mounted, horizontal modular headwall assembly that has been specifically designed to rapidly add capacity to an acute care space for surge type scenarios. Horizontal channels are an integral, non-removable part of the horizontal tier assembly and are flush with the front face of the unit. The lower channel surface is flat and smooth to facilitate ease of cleaning. The horizontal equipment channels have a clear anodized aluminum finish. Horizontal channel accessories are available in a sufficient variety to meet the needs of...

 Open the catalog to page 2
Provider Plus® 6540-AC Installation Manual-3

To attach hanging brackets, measure 9.8" below the device console centerline to locate bottom of hanger brackets. Mark a level horizontal line at this location for use in aligning the brackets. Mount two hanger brackets (Figure 2, item 1) with 3/8" Rawl hollow wall anchors (item 2) or equivalent (by others) into wall studs only. Position hanger brackets so that both are within the length of the unit and will engage one stud each. Drill 25/64" holes through drywall and metal studs. Fasten with anchors after leveling hangers and tighten fully. Figure 2. Device Console Centerline Hanger Bracket...

 Open the catalog to page 3
Provider Plus® 6540-AC Installation Manual-4

Once unit is aligned with all services, locate position of all wall studs behind the unit. Using a 25/64" drill bit, drill one hole per stud through detent in top extrusion (Figure 4). Each anchor must pass through unit, drywall and wall stud. Fasten with 3/8" Rawl hollow wall anchors or equivalent (by others). Level unit and fully tighten all anchors. Detent in top extrusion for upper mounting anchors ELECTRICAL SERVICE CONNECTIONS Electrical connections can be made on either end of the unit. Field conditions will determine which end is more convenient. Power is to be connected to the unit via...

 Open the catalog to page 4
Provider Plus® 6540-AC Installation Manual-5

Clean with mild detergent and warm water. Do not use lubricants or oils on unit. Avoid excessive moisture build-up, as it can damage mechanisms in the unit. Disinfect as required with disinfectant approved by the Environmental Protection Agency. To remove difficult spots or stains, use standard household cleansers and a soft-bristled brush. For parts or repairs, contact your sales representative. INSTALLATION TERMS AND CONDITIONS Each Modular Services unit, or unit section, shall be completely pre-wired for normal and emergency. Modular Services Company shall have no responsibility or liability...

 Open the catalog to page 5
Provider Plus® 6540-AC Installation Manual-6

WARRANTY Modular Services Company warrants that all equipment assemblies shall be free from defects in material and workmanship for a period of 12 months from date of the owner’s acceptance to the installing contractor or the date the equipment is put into service, whichever comes first. Warranty excludes electric lamps and/or any material not furnished by Modular Services. Warranty does not cover damage due to improper installation and/or abuse. It is the responsibility of the customer to report any noted product deficiencies to Modular Services immediately upon discovery. It is the responsibility...

 Open the catalog to page 6
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.