
Catalog excerpts

About Moor Instruments Moor Instruments, established in 1987, is a world leader in Tissue blood flow and temperature monitoring with moorVMS-LDFTM the design, manufacture and distribution of laser Doppler systems, used for the monitoring and imaging of blood flow in the microvasculature. First hand experience of laser Doppler research and development within Moor dates back to 1978 and with this we have the breadth of knowledge to help with your application and the enthusiasm to try to find answers to any of your questions. By giving priority to performance, quality and service we strive to be our customers number one choice. Our dedicated design team are involved with a number of development projects for other partners and moorVMS-LDF2 with optional moorVMS-LDF1 single channel module. manufacturers. Whatever your needs, as a researcher, clinician or manufacturer, Moor will work harder for you. Specifications: Quality Control Moor Instruments is certified to ISO 13485: 2003. The moorVMS-LDF is CE certified. Measurement Parameters Flux (tissue perfusion). Conc* (blood cell concentration). DC (mean intensity). Temperature. Measurement Channels moorVMS-LDF1 Single channel. moorVMS-LDF2 Dual channel. Reliability Assured by our industry leading 5 year manufacturers warranty for all the moorVMS-LDF modules. Optics Temperature stabilised output laser diode; 785nm. Maximum output power 2.5mW. Laser Safety Classification Class 1 per IEC 60825-1:2007. Class 1 per 21 CFR 1040.10 and 1040.11. Laser Doppler Signal Processing Bandwidth: High pass 20Hz. Low pass 3kHz*, 15kHz, 22kHz*. Flux smoothing time constants: 0.1s*, 0.5s*, 1.0s, 3.0s* and unfiltered*. Automatic gain control and zeroing. Temperature Measurement Range: 5ºC to 50ºC. Resolution: 0.1ºC, accuracy +/- 0.3ºC. Outputs LCD screen providing display of flux, DC and temperature. USB Interface for connection to PC. Analogue outputs: BNC sockets, 0-5V. moorVMS-LDF1 1x flux, 1x temperature/DC (selectable). moorVMS-LDF2 2x flux, 2x temperature/DC (selectable). All outputs have independent user selectable scaling. General Power source: Universal voltage, 100-230V AC, 30VA, 50 to 60Hz. Dimensions: W x H x D mm, Weight: kg moorVMS-LDF1: 235 x 60 x 200, 1.4kg. moorVMS-LDF2: 235 x 80 x 200, 1.5kg. Operating environment: Clinic or laboratory, excluding domestic. Operating temperature: 15-30ºC. Classification Medical devices directive 93/42/EEC: Class IIa, Active device for diagnosis. Laser classification: Class 1 per IEC 60825-1:2007. Type of protection against electric shock: Class I. Degree of protection against electric shock: Type BF applied parts. Protection against harmful ingress of water: IPX0 (not protected). Not suitable for use in an oxygen rich atmosphere. Not suitable for use in the presence of flammable anaesthetics. Safety Standards Complies with: IEC 60601-1:2005, IEC 60601-1-2:2007, IEC 60825-1:2007. ISO 10993-1:2003, Medical devices directive 93/42/EEC. FDA Laser Notice No. 50; dated July 26, 2001. References A.P. Shepherd, P. Å. Öberg, Laser-Doppler Blood Flowmetry, 1990, ISBN 0-7923-0508-6, Kluwer Academic Publishers. Moor Instruments reserves the right to change specifications without notice. *Feature only available whilst using the PC Windows software, please contact us for a demo version. innovation in microvascular assessment Moor Instruments Ltd Millwey Axminster Devon EX13 5HU UK tel +44 (0)1297 35715 fax +44 (0)1297 35716 email sales@moor.co.uk website www.moor.co.uk Moor Instruments Inc 501 Silverside Road Suite # 66 Wilmington DE 19809 USA tel (302) 798 7470 fax (302) 798 7299 email sales@moorinc.com innovation in microvascular assessment
Open the catalog to page 1
moorVMSTM - Vascular Monitoring System The moorVMS-LDF laser Doppler blood flow monitor is a high performance, medical grade module for clinic or laboratory. Use of DSP technology brings you a portable, lightweight module featuring uncompromised specifications and reliability at a breakthrough price. The features include; The laser Doppler technique Laser Doppler has become a gold standard for dynamic microvascular blood flow assessments. Optic probes are used to monitor blood flow in capillaries, venules and arterioles; vessels responsible for nutrition of the skin and thermoregulation....
Open the catalog to page 2
moorVMSTM - Vascular Monitoring System The moorVMS-LDF laser Doppler blood flow monitor is a high performance, medical grade module for clinic or laboratory. Use of DSP technology brings you a portable, lightweight module featuring uncompromised specifications and reliability at a breakthrough price. The features include; The laser Doppler technique Laser Doppler has become a gold standard for dynamic microvascular blood flow assessments. Optic probes are used to monitor blood flow in capillaries, venules and arterioles; vessels responsible for nutrition of the skin and thermoregulation....
Open the catalog to page 3
About Moor Instruments Moor Instruments, established in 1987, is a world leader in Tissue blood flow and temperature monitoring with moorVMS-LDFTM the design, manufacture and distribution of laser Doppler systems, used for the monitoring and imaging of blood flow in the microvasculature. First hand experience of laser Doppler research and development within Moor dates back to 1978 and with this we have the breadth of knowledge to help with your application and the enthusiasm to try to find answers to any of your questions. By giving priority to performance, quality and service we strive to...
Open the catalog to page 4All Moor Instruments catalogs and technical brochures
-
moorVMS-TREND
4 Pages
-
moorVMS-OXY_Brochure
4 Pages
-
moorVMS-VASC
4 Pages