USER MANUAL MPW-260 Refrigerated laboratory centrifuge MPW-260R Laboratory centrifuge Serial number of the centrifuge: ………………………………………………………. For centrifuges with serial no (SN):

Open the catalog to page 1
User Manuel MPW-260 - 2

This manual was prepared with special care. MPW MED. INSTRUMENTS may change the manual at any time and without notice because of improvements, typographical errors or improvements to facilities All rights reserved. No part of this User Manual may be modified, distributed, published or reproduced without the prior permission of MPW MED. INSTRUMENTS. You can find the current version of the user manual on our website under: mpw.pl/en – DOWNLOAD section (one can choose demanded language version of website). Warning signs: WARNING! Warning of potential injury or health risk. DANGER! Risk of...

Open the catalog to page 2
User Manuel MPW-260 - 4

B. DECLARATION OF CONFORMITY (CE, ROHS 2) C. DECLARATION OF DECONTAMINATION (repair / return)

Open the catalog to page 4
User Manuel MPW-260 - 5

1 Application The MPW-260/R centrifuges (MPW-260 – ventilated, MPW-260R – with cooling) belong to the family MPW-260/R/RH of the table top laboratory centrifuges for in vitro diagnostic (IVD). Devices are used for separation samples taken from people's, animal’s and plant’s components of different densities, under the influence of the centrifugal force, to provide information about their biological state. Its construction ensures easy operation, safe work and wide range of applications at laboratories engaged in routine medical analyses, biochemical research works etc. In the centrifuge, it...

Open the catalog to page 5
User Manuel MPW-260 - 6

2 Technical specification "MPW MED. INSTRUMENTS" SPÓŁDZIELNIA PRACY, Boremlowska 46 Street, 04-347 Warsaw MPW - 260 MPW - 260R manufacturer type mains voltage (L1+N+PE) mains frequency, current protection [A] cooling medium running time 500 ml 90  18000 rpm (step 1 rpm) 24270 x g (step 1 x g) 8800 Nm 00:00:01 ÷ 99:59:59 – [hours, min., sec] (step 1s) time counting since start button is pressed / since preselected speed is reached capacity (max.) speed – RPM force – RCF kinetic energy (max.) short-time operation mode – SHORT continuous operation mode – HOLD user programs adjustable...

Open the catalog to page 6
User Manuel MPW-260 - 7

3 Installation Open the package. Remove the box containing the accessories. Take out centrifuge from the container. Keep the box and packing materials in case of service shipping. Content of the package name complete clamp spanner for the rotor spanner for emergency opening of the cover power cord – 230V / 120V fuse WTA T 4A – 230V / WTA T 6,3A – 120V (for MPW-260) fuse WTA T 10A – 230V / 120V (for MPW-260R) vaseline 20ml USB A-A cable user manual Ensure safe location. The centrifuge shall not be located near source of heat and shall not be subjected to direct sunlight. Centrifuge should be...

Open the catalog to page 7
User Manuel MPW-260 - 8

Do not position the centrifuge so that it is difficult to operate the power switch. Supply voltage given on the name plate has to be consistent with local supply voltage. MPW MED INSTRUMENTS laboratory centrifuges are 1st safety class devices and they are provided with the three-core cable with the plug resistant to dynamic loadings. Mains socket have to be provided with the safety pin (protective earth (PE)). It is recommended to install emergency cut-out that shall be located far from the centrifuge, near the exit or beyond the room. Before switching on, check whether the centrifuge is...

Open the catalog to page 8
User Manuel MPW-260 - 9

Fix the rotor on the motor axis firmly. Avoid unbalance. Load opposite buckets with the same accessories. Centrifugation of the test tubes of different sizes: ▪ There is a possibility to centrifuge test tubes of different sizes; however, it is absolutely necessary in such cases that opposite buckets and round carriers be the same. Mass of different containers with test tubes spun at the same time has to be comparable. Swing-out rotors must be equipped with all ( two or four –dependent to type of rotor) buckets. Lubricate the swing-out rotor journal pins. It is necessary to insert test tubes...

Open the catalog to page 9
User Manuel MPW-260 - 10

Fill test tubes outside the centrifuge. Please pay special attention to the quality and proper thickness of the glass test tubes walls. Those shall be test tubes for centrifuges. Fill the rotor equipment to the same weight to prevent the centrifuge from being unbalanced. Safety hints ROTORS MAINTENANCE ▪ Lubricate the swing-out rotor journal pins. Use only accessories in good condition. Protect equipment against corrosion using accurate preventive maintenance. Make sure that rubber O-rings are lightly coated with silicone grease. Use high vacuum grease, e.g. type „C” by LUBRINA. MPW...

Open the catalog to page 10
User Manuel MPW-260 - 11

Maintenance conditions START-UP ▪ Prior to switching the centrifuge on, one shall read carefully all sections of this user manual in order to ensure smooth operation and avoid damages of this device or its accessories. In order to protect the centrifuge against unbalance, fill in the test tubes up to the same weight. Centrifuge must not be transported with the rotor mounted on the shaft.. One must use original rotors, tubes and spare parts only. In case of faulty operation of the centrifuge one shall ask for assistance of service of MPW MED. INSTRUMENTS company or its authorized...

Open the catalog to page 11
User Manuel MPW-260 - 12

The centrifuge must be inspected by an authorized service provider at least once a year (after the warranty period). For example, the corrosive environment may be the reason for more frequent control. The tests shall be completed by issuing a validation protocol specifying checks on the technical condition of the laboratory centrifuge. It is recommended that you set up a document that records any repairs and inspections. Both of these documents should be stored at the place of use of the centrifuge. INSPECTION PROCEDURES CARRIED OUT BY THE OPERATOR Operator has to pay special attention to...

Open the catalog to page 12
User Manuel MPW-260 - 13

It is not allowed to use the rotors and round carriers with signs of corrosion or other mechanical defects. It is not allowed to centrifuge highly corrosive substances which may cause material impairment and lower mechanical properties of rotor and round carriers. It is not allowed to use rotors and accessories not admitted by the manufacturer. Let to use commercial glass and plastic test tubes, which are destined to centrifuging in this laboratory centrifuge. One should absolutely not use poor quality elements. Cracking of glass vessels and test tubes could result in dangerous vibration of...

Open the catalog to page 13