Overview: The Nice 4020 Radiometer is designed to accurately measure the intensity of phototherapy systems, aiding compliance with the American Academy of Pediatrics (AAP) guidelines for managing hyperbilirubinemia in newborns.
Key Features:
  • Measures light intensity in the blue spectrum to ensure effective phototherapy treatment.
  • Can be used for both overhead and under-baby systems, with a diffuser on the sensor to minimize angular sensitivity.
  • Standard units of measurement include µW/cm2/nm and µW/cm2.
  • Auto-ranging capability from 0.1 to 199.9 μW/cm2/nm.
  • Microprocessor electronics enhance linearity and reliability.
  • Precision optical system ensures reliable and reproducible measurements.
Technical Specifications:
  • Spectral Response Range: 400 to 520 nm.
  • Centre Wavelength: 450 nm.
  • Bandwidth: 60 nm.
  • Measuring Functions: Spectral Irradiance in µW/cm2/nm and µW/cm2.
  • Measuring Range: 0.1 ~ 199.9 µW/cm2/nm and 1 ~ 8540 µW/cm2.
  • Hold Function: Allows holding of the measured value.
  • Alarm Condition: Displays "OvLd" with a continuous beep if irradiance exceeds peak rating.
Compliance and Safety:
  • IEC 60601-1 Specifications: Class 1 protection against electric shock, Type B degree of protection.
  • Mode of Operation: Short term usage.
  • Not protected against hazardous explosions.
  • Protection against liquid ingress: IPX1.
  • Product Classification: Class IIa.
Company Information: Neotech is an ISO 13485:2016 certified company with design capabilities, located in Chennai, India.
See more

Catalog excerpts

Phototherapy Radiometer-1

Ray nice 4020 Helps you comply with AAP Guidelines by providing accurate readings in blue spectrum1 To accurately measure the intensity of Phototherapy Systems 1 Subcommittee on Hyperbilirubinemia. American Academy of Pediatrics clinical practice guideline: Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics. 2004; 114(1):297-316

 Open the catalog to page 1
Phototherapy Radiometer-2

Spectral Response Range : 400 to 520 nm Measuring Functions :Spectral Irradiance: ^W/cm2/nm Measuring Range: Spectral Irradiance : 0.1 ~ 199.9 ^W/cm2/nm Spectral Irradiance: 1 ~ 8540 ^W/cm2 Hold Function:Hold the measured value. Alarm Condition High Irradiance:If the Irradiance level exceeds peak rating then display will indicate as “OvLd” (Over Load) with continuous beep. IEC 60601-1 Specifications : Class 1 : Type B : Not protected : IPX1 : Class IIa. Type of Protection against electric Shock Degree of Protection against electric Shock Mode of Operation : Short term usage Protection against...

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.