1. Catalogs
  2. Nihon Kohden Europe
  3. iNIBP Case Report – Experiences in OR
video corpo

iNIBP Case Report – Experiences in OR
1 /12Pages

iNIBP Case Report – Experiences in OR

iNIBP Case Report – Experiences in OR
1 /12Pages

Catalog excerpts

iNIBP Case Report – Experiences in OR-1

Linear inflation Technology Non-invasive Blood Pressure Measurement Fighting Disease with Electronics

 Open the catalog to page 1
iNIBP Case Report – Experiences in OR-2

C ONTENTS Introduction……………………………………………………………………………… 2 Technological Description of iNIBP iNIBP advantages………………………………………………………………………… 2 iNIBP technology adapts to each patient situation and condition ………………… 4 [Case 1] Laparoscopic assisted distal gastrectomy……………… 5 [Case 2] Pelvic lymphadenectomy, para-aortic lymphadenectomy [Case 3] Laparoscopic resection of descending c

 Open the catalog to page 2
iNIBP Case Report – Experiences in OR-3

iNIBP Case Report – Experiences in OR Introduction iNIBP is Nihon Kohden's new non-invasive blood pressure measurement algorithm using linear inflation technology. This case report, which describes the perioperative use of iNIBP, is designed to help clinicians and healthcare professionals understand and appreciate the performance characteristics of iNIBP, including the advantages and limitations of the measurement. Technology Description of iNIBP iNIBP advantages Non-invasive blood pressure (NIBP) is commonly used as an important vital sign to evaluate patients' hemodynamic status. In the traditional...

 Open the catalog to page 3
iNIBP Case Report – Experiences in OR-4

Cuff 300 pressure (mmHg) 250 iNIBP can complete the measurement in a shorter amount of time as it inflates the cuff while sensing blood pressure variation and detecting oscillations, whereas the deflation method needs to repeat the inflation-deflation cycle when the target inflation pressure, based on previous systolic pressure measurements are comparatively low. iNIBP Deflation method Deflation measurement completes Measurement completes as soon as SYS is determined Shorter measurement time Figure 3.a. iNIBP in blood pressure elevation Cuff 300 pressure (mmHg) 250 Measurement completes as soon...

 Open the catalog to page 4
iNIBP Case Report – Experiences in OR-5

iNIBP Case Report – Experiences in OR iNIBP technology adapts to each patient situation and condition iNIBP automatically switches its measurement mode to step-deflation in situations where it cannot determine an accurate blood pressure value with inflation method because of factors related to measurement environment and patient's physiology. Also, iNIBP automatically employs the deflation method when signal reliability of the blood pressure reading, based on the blood pressure determined while inflating, is poor (Figure 4). Examples of these situations and conditions are shown below. When these...

 Open the catalog to page 5
iNIBP Case Report – Experiences in OR-6

[Case 1] Laparoscopic assisted distal gastrectomy Disease : Stomach cancer Anesthesia : General anesthesia with epidural block (volatile anesthetic) Patient : 68 years old male, height 170 cm, weight 81 kg. Cuff on left arm Anesthesia time : 8:55 - 15:50 Surgical time : 9:37 - 15:25 in this case, a minimally-invasive procedure was performed. Blood pressure increased approximately 30 mmHg between the second (A) and third (B) NIBP measurement, then decreased 25 mmHg between the third rB) and fourth (C) measurement. in cases such as this where blood pressure varies significantly, the conventional...

 Open the catalog to page 6
iNIBP Case Report – Experiences in OR-7

iNIBP Case Report - Experiences in OR [Case 2] Pelvic lymphadenectomy, para-aortic lymphadenectomy Disease : Ovary cancer Anesthesia : General anesthesia with epidural block {volatile anesthetic} Patient : 47 years old female, height 161 cm, weight 56 kg. Cuff on left arm Anesthesia time : 10:55 - 17:06 Surgical time : 11:21 - 16:39 in the first half of the surgery (beginning of surgery until approximately 14:30), the cuff frequently detected noise caused by touching the cuff, so the iNiBP algorithm did not determine blood pressure while inflating the cuff and switched its measurement mode to...

 Open the catalog to page 7
iNIBP Case Report – Experiences in OR-8

Disease : Descending colon cancer Anesthesia : General anesthesia with epidural block {volatile anesthetic} Patient : 72 years female, height 151 cm, weight 74.5 kg. Cuff on left arm Anesthesia time : 9:05 - 15:49 Surgical time : 10:00 - 15:03 Deflation measurement Average Minimum Maximum The patient had low amplitude pulse wave and her blood pressure was relatively low (SYS 60-90/DIA 30-50 mmHg) during surgery. When the amplitude of oscillations is lower than a threshold, where it cannot determine an accurate blood pressure value with inflation mode, iNIBP determines NIBP value with the deflation...

 Open the catalog to page 8
iNIBP Case Report – Experiences in OR-9

iNIBP Case Report - Experiences in OR [Case 4] Radical prostatectomy Disease : Prostate cancer Anesthesia : General anesthesia with epidural block (volatile anesthetic) Patient : 65 years old male, height 168 cm, weight 68 kg. Cuff on left arm Anesthesia time : 11:50 - 17:25 Surgical time : 12:14 - 16:52 Deflation measurement -Anesthesia starts -Anesthesia ends ---Surgery starts -Surgery ends Measurement time Average Minimum in this case, all measurements throughout the surgery, from the anesthesia Induction through patient awakening, were performed with the inflation measurement mode under the...

 Open the catalog to page 9
iNIBP Case Report – Experiences in OR-10

[Case 5] Laparoscopic uterine myomectomy Disease : Uterine myoma Anesthesia : General anesthesia (volatile anesthetic) Patient : 37 years old female, height 151 cm, weight 53 kg. Cuff on left arm Anesthesia time : 8:38 – 12:43 Surgical time : 9:08 – 12:25 Deflation measurement Inflation measurement Anesthesia starts Anesthesia ends Surgery starts Surgery ends After the start of the surgery, the deflation method was used for the third ( A ) and fourth ( B ) measurement because of noise caused by patient's body motion. Subsequently, iNIBP completed all measurements with the inflation measurement...

 Open the catalog to page 10
iNIBP Case Report – Experiences in OR-11

iNIBP Case Report – Experiences in OR Discussion Nihon Kohden's Linear Inflation Technology, Non-invasive Blood Pressure Measurement,“iNIBP”determines blood pressure by , detecting oscillations during the initial inflation of the cuff. Compared to the conventional method, iNIBP can provide accurate blood pressure readings in less time without applying excessive pressure to the patient, leading to maximized comfort for patients. iNIBP allows clinicians to accurately measure blood pressure immediately and when this critical vital sign is needed the most, especially during dynamic blood pressure...

 Open the catalog to page 11

Archived catalogs

  1. OLG-2800K

    2  Pages

  2. YAWARA CUFF 2

    2  Pages

  3. MEE-1000A

    8  Pages

  4. PMS-Brochure

    8  Pages

  5. ECG-1950 VET

    2  Pages

  6. EEG

    4  Pages

  7. EMG

    4  Pages

  8. aEEG

    2  Pages

  9. BluPRO

    2  Pages

  10. Cardiofax C

    4  Pages

  11. Life scope G9

    4  Pages

  12. ViTrac

    4  Pages

  13. Life_Scope_VS

    4  Pages

  14. Cardiofax_S

    4  Pages

  15. esCCO_8755

    4  Pages

  16. INBP

    12  Pages

  17. YZ-0194

    3  Pages

  18. CAP-One

    2  Pages

  19. SpO2 Probes

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.