Overview: The YP-90AC Infant Incubator is designed to provide a safe and controlled environment for newborns. It features dual control modes (Air and Baby) managed by a micro-computer, separate displays for set, air, and baby temperatures, and a humidity concentration system.
Specifications: The incubator operates within a temperature range of 20℃ to 30℃, with a set temperature requirement of 3℃ above ambient. It includes a low noise DC motor, RS-232 data output, and an oxygen input interface. The power requirement is AC220V-230V/50Hz or AC110-120V/50-60Hz, 1000VA.
Temperature Control: Air temperature control ranges from 25℃ to 37℃, with an optional override mode up to 39℃. Baby temperature control ranges from 34℃ to 37℃, with an optional override mode up to 38℃. The skin temperature sensor accuracy is ±0.2℃, with temperature variability ≤0.5℃ and uniformity ≤0.8℃.
Humidity and Noise: Humidity control ranges from 20% to 90% RH, with a display range of 0%RH to 99%RH. Noise inside the hood is ≤45dB(A).
Alarms and Safety: The incubator includes multiple failure alarms such as over temperature, deviation, sensor failure, fan motor, power failure, and system alarms.
Phototherapy Equipment: Configured with fixed Neonate Bilirubin Phototherapy Equipment, it offers options for blue fluorescent lamps or blue LEDs, with irradiance values of ≥640uW/cm2/nm and ≥1300uW/cm2/nm respectively.
Optional Features: Includes a vertical height adjustment stand, weighing system, and disposable skin temperature sensor.
Dimensions and Packaging: The main body measures 1085mm x 640mm x 910mm, and the cabinet measures 1125mm x 665mm x 725mm. The gross weight is 78Kg for the main body and 68Kg for the cabinet. Both are packed separately.
Contact Information: Ningbo David Medical Device Co., Ltd., located at Bldg No. 5 Ningbo Smart Park, #98 Chuangyuan Rd, Hi-tech Zone, Ningbo, China. Tel: 0086-574-8780 0008 or 8780 0007, Fax: 0086-574-8780 3714, Email:
[email protected], Website: www.nbdavid.com