Brief Introduction:The TI-2000 Transport Incubator is designed for use in ambulances, compliant with EN 1789 standards. It features dual temperature control modes (Air and Baby) managed by a micro-computer, and can operate on both AC and DC power sources.
Specifications:- Temperature Control: Air mode (25℃-37℃, Override 37.1℃-38℃) and Baby mode (34℃-37℃, Override 37.1℃-37.5℃).
- Power Requirements: AC220V-230V, 50Hz or AC110-120V, 50/60Hz, 400VA; DC12V/10A or DC24V/6A.
- Temperature Sensor: Display range 5℃-65℃, accuracy ±0.2℃.
- Noise Level: ≤52 dB(A) inside hood.
- Battery: 90 minutes continuous operation.
- Dimensions: Main body 1090mm x 655mm x 760mm, Cabinet 1380mm x 610mm x 400mm.
- Weight: Main body 80Kg, Cabinet 53Kg.
Features:- Displays for set temperature, air temperature, baby temperature, and battery capacity.
- Multiple failure alarms including over-temperature, sensor, and power failure alarms.
- Double wall acrylic glass hood with side door for easy access.
- Passive humidification and adjustable height.
- Integrated oxygen supply system and observation lamp.
Optional Configuration:- ">37℃" temperature set function.
- Disposable skin temperature sensor.
- Ambulance trolley.
Environmental and Packaging:- Operating range: 10℃~30℃ (set temperature should be 3℃ higher than ambient temperature).
- Ambient air velocity: >0.3 m/s, <1.0 m/s.
- Packaged in two cartons for main body and cabinet.
Contact Information:Ningbo David Medical Device Co., Ltd, located at Bldg No. 5 Ningbo Smart Park, #98 Chuangyuan Rd, Hi-tech Zone, Ningbo, China. For inquiries, contact via fax at 0086-574-8780 3714 or phone at 0086-574-8780 0008/0007. Visit their website at www.nbdavid.com/en/ or email
[email protected].