1. Catalogs
  2. Northern Meditec
  3. ICU Ventilator Crius V6

ICU Ventilator Crius V6

ICU Ventilator Crius V6

Product catalog summary
Contact Information:
Address: 4th Floor, Building C, Jin Wei Yuan Industrial Park, Julongshan Area, Kengzi St, Pingshan District, Shenzhen City, P.R.China
Tel: +86 755 29970566
Email: [email protected]
Website: northernmeditec.com
Fax: +86 755 86528647
Product Overview:
Product Name: Crius V6 ICU Ventilator
Display: 15.6-inch diagonal, 18.5-inch optional, Active matrix color TFT LCD, SVGA resolution 1366 x 768
Ventilation Type: Blower
Ventilation Modes: A/C-VCV, A/C-PCV, A/C-PRVC, SIMV-VC, SIMV-PC, SIMV-PRVC, BIPAP, APRV, CPAP/PSV, CPAP-VG, NIV, IV
Special Functions: Manual breath, O2 therapy, standby, screen-lock, inspiratory hold, expiratory hold, sigh, ATC, RSB, Compliance, p0.1, calibrate, nebulizer, O2 suction
Patient Types: Adult/Pediatric/Neonate
Patient Height: Adjustable
Monitoring and Parameters:
Real-time Waveforms: Paw, Flow, Volume, EtCO2, SPO2
Optional Parameters: CO2, SPO2
Loops: P-V, V-Flow, Flow-P
Trends: 1s, 5s, 1min, 10min, 30min, 60min
Tidal Volume: 5 to 2000mL (2 to 2000ml optional)
Rate: 1 to 100bpm
PEEP/CPAP: 0 to 50 cmH2O
Oxygen: 21 to 100%
I:E Ratio: 1:10 to 4:1
Inspiratory Time (TI): 0.2 to 5 s
Flow Trigger: 0.2 to 15 l/min
Pressure Trigger: -0.5 to -20cmH2O
Pressure Control: 5 to 90 cmH2O
Pressure Support: 0 to 90 cmH2O, added to PEEP/CPAP
Pressure Ramp: 0 to 2s
Apnea Absence: 10 to 30s
Expiratory Trigger Sensitivity (ETS): OFF, 5 to 90 % of inspiratory peak flow
Peak Flow Spontaneous: > 210 L/min
Electrical and Gas Supplies:
Input Voltage: 110 to 240V ~ ±10 %, 50 / 60 Hz AC or 24 VDC±10 %
Power Consumption: 50 W typical
Protection Class: Class I
Backup Battery: 2 batteries (5200mAh each), hot swappable
Oxygen Supply: 280 to 600 kPa (41 to 87 psi), 120 L / min
Low Pressure: Oxygen ≤15 L / min, max. 600 hPa for low pressure
Alarms and Safety:
Operator-adjustable: Low/high minute volume, low/high pressure, low/high tidal volume, low/high rate, apnea time, low/high oxygen
Optional Alarms: Low/high CO2, low/high SPO2
Special Alarms: Oxygen concentration, disconnection, loss of PEEP, exhalation obstruction, flow sensor, power supply batteries, gas supply
Loudness Adjustable: 1 to 9
Event Log: Storage and display of up to 1000 events with date and time stamp
Physical Dimensions:
Weight: < 45.3 kg
Main Unit Dimensions (WxHxD): Approx. 600x402x415mm
See more

Catalog excerpts

ICU Ventilator Crius V6-1

I Electrial interface Nurse call, USB, NET, RS232, VGA interface Type Main patient outlet:ISO 22M/15F Oxygen inlet:DISS or NIST male Display Display size: 15.6 in diagonal, 18.5"optional Display type: Active matrix color TFT LCD Display resolution: SVGA resolution, 1366 x 768 ontrols Fully electronic control Ventilation type: Blower Ventilation modes: A/C-VCV, A/C-PCV, A/C-PRVC, SIMV-VC, SIMV-PC, SIMV-PRVC, BIPAP, APRV, CPAP/PSV, CPAP-VG, NIV, IV. Special functions: manual breath, O2 therapy, standby, screen-lock, inspiratory hold, expiratory hold, sigh, ATC, RSB, Compliance, p0.1, calibrate, nebulizer, O2 suction. Patient types: Adult/Pediatric/Neonate Patient height: Adjustable Flow trigger: Pressure trigger: Pressure control: Pressure support: Pressure ramp: Apnea absence: Monitoring Real-time waveforms: Paw, Flow, Volume, EtCO2, SPO2 Optional: CO2, SPO2 (related parameters) Loops: P-V, V-Flow, Flow- P Tidal volume: 5 to 2000mL (2 to 2000ml optional) Expiratory trigger sensitivity (ETS): OFF, 5 to 90 % of inspiratory peak flow Peak flow Spontaneous: > 210 L/min Electrical and gas supplies Input voltage: 110 to 240V ~ ±10 %, 50 / 60 Hz AC or 24 VDC±10 % Power consumption: 50 W typical Protection class: Class I Backup battery: 2 batteries(5200mAh each)/ hot swappable Oxygen supply: 280 to 600 kPa (41 to 87 psi), 120 L / min Low pressure: oxygen <15 L / min, max. 600 hPa for low pressure Operator-adjustable: Low/high minute volume, low/high pressure, low/high tidal volume, low/high rate, apnea time, low/high oxygen Optional: low/high CO2, low/high SPO2 Special alarms: Oxygen concentration, disconnection, loss of PEEP, exhalation obstruction, flow sensor, power supply batteries, gas supply Loudness Adjustable: 1 to 9 Event log: Storage and display of up to 1000 events with date and time stamp Main unit dimensions(WxHxD): Approx. 600x402x415mm Northern Meditec Limited Add: 4th Floor, Building C, Jin Wei Yuan Industrial Park, Julongshan Area, Kengzi St, Pingshan District, Shenzhen City, P.R.China Tel: +86 755 29970566 Fax: +86 755 86528647 Email: [email protected] Website: http://www.northernmeditec.com

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.