1. Catalogs
  2. Northern Meditec
  3. Patient Monitor -Virgo

Patient Monitor -Virgo

Patient Monitor -Virgo

Product catalog summary
Overview: The document provides detailed specifications and features of the Virgo Patient Monitor by Northern Meditec Limited. It highlights the monitor's capabilities, optional configurations, and technical specifications.
Specifications: The Virgo Patient Monitor features a 12.1” color TFT LCD screen with a wide and flat design. It supports up to 8 waveform displays and 12-lead ECG analysis. The monitor includes powerful calculation capabilities for hemodynamic, dose, oxygenation, and ventilation, along with pacemaker detection and ST & arrhythmia analysis.
Optional Configurations: The monitor can be configured with Masimo SpO2, IBP, CO2, a thermal recorder, a touch screen, and a Li-ion battery. It supports 2-channel or 4-channel IBP with various measurement ranges and accuracies.
Measurement Capabilities:
  • SpO2: Measurement range of 0-100% with a resolution of 1% and accuracy of ±2% for adults and pediatrics, ±3% for neonates.
  • NIBP: Automatic oscillometric method with systolic, diastolic, and mean pressure measurements. It includes overpressure protection and a cuff pressure range of 0-280mmHg.
  • TEMP: Two-channel temperature measurement with a range of 5~50℃ and accuracy of ±0.1℃.
  • Pulse Rate: Range of 30~254 bpm with accuracy of ±2bpm (non-motion).
  • RESP: Thoracic electrical bioimpedance method with a range of 3~150 rpm.
  • ECG: Supports 3-lead, 5-lead, and 12-lead configurations with various display sensitivities and sweep speeds.
Operational Environment: The monitor operates on AC 100-250V, 50/60Hz, within a temperature range of 5-40℃ and humidity of less than 80%. It is suitable for adult, pediatric, and neonatal patients.
Additional Features: The monitor includes a graphical and tabular trend review for up to 120 hours, a rechargeable Lithium-Ion battery, and various mounting solutions. It supports wired/wireless CMS and HL7 protocol to HIS, with SpO2 pulse-tone modulation and MEWS (Modified Early Warning Score).
See more

Catalog excerpts

Patient Monitor -Virgo-1

Patient Monitor 12.1"color TFT LCD screen,wide and flat screen design,ecnomic and reliable Configuration Optional ECG + SpO2+NIBP+2TEMP+PR+RESP, Li-ion battery Masimo SpO2, IBP, CO2, Thermal recorder, Touch screen, CMS 12.1" color TFT LCD screen 8 waveform display,up to 12-lead ECG analysis Powerful calculation(Hemodynamic,Dose,Oxygenation,Ventilation) Pacemaker detection ST & arrhythmia analysis Various mounting solutions Wired/Wireless CMS, support HL7 protocol to HIS SpO2 pulse-tone modulation (Pitch Tone) MEWS(Modified Early Warning Score) Graphical & tabular trend review( 120 hours) OxyCRGs screen * Night mode, standby mode, venipucture mode Rechargeable Lithium-Ion Battery 48 Hours full disclosure waveforms display Graphical & Tabular Trend Display TEMP Masimo SET® SpO2(Optional) 12.1" TFT LCD screen Resolution: 800 x 600 Number of traces: 8 waveforms ECG Accuracy:±0.1°C or ±0.2°C °F Measurement range: 5~50°C (41~122°F) Channel:Two channels Resolution: 0.1°C Parameters:T1,T2 and TD Lead type :3-lead,5-lead,12-lead ECG waveform:2 channels,7 channels, 12 channels Display sensitivity: 2.5mm/mV (x0.25), 5mm/mV (x0.5), 10mm/mV (x1.0), 20mm/mV (x2.0) Wave sweep speed: 6.25mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s Bandwidth Diagnostic mode: 0.05Hz~100Hz Monitor mode: 0.5Hz~40Hz Surgery mode: 1Hz~20Hz Strong filter mode: 5Hz~20Hz CMRR>106 dB Notch: 50/60Hz notch filter can be set to on or off Differential input impedance>5MQ Electrode polarization voltage range: ±400mV HR range: 15 - 350 bpm Baseline recovery time<3s after defibrillation (in monitor and surgery mode) Calibration signal:1mV (peak - peak), accuracy ±3% RESP Measurement method : Thoracic electrical bioimpedance Measuring lead: Lead I, II Wave gain: x0.25, x0.5', x1, x2 Measurement range: 3~150 rpm Bandwidth: 0.3 to 2 Hz Respiratory impedance range: 0.5-5Q Baseline impedance: 500-4000Q Gain: 10 grades Scan speed: 6.25mm/s, 12.5 mm/s, 25mm/s and 50mm/s Pulse Rate Range: 30~254 bpm Resolution: 1bpm Accuracy: ±2bpm (non-motion) ±5bpm (motion) Refreshing rate: 1s Northern Meditec Limited NIBP Measurement method : Automatic oscillometric method Operating mode:Manual, automatic, continuous Measurement unit: mmHg/kPa selectable Typical measurement time: 20~40s Measurement type: Systolic, Diastolic,Mean Measurement range (mmHg) Range of Systolic pressure: Adult 40-270 Range of Diastolic pressure: Adult 10-210 Range of Mean pressure: Adult 20-230 Measurement accuracy Maximum average error: ±5mmHg Maximum standard deviation: 8mmHg Resolution: 1mmHg Interval:1,2,3,4,5,10,15,30,60,90,120,180,240,480minutes Overpressure protection: Software and hardware, double safety protection Cuff pressure range: 0-280mmHg Measurement range : 0-100% Resolution: 1% Accuracy: ±2%(70-100%, Adult/Pediatric,non-motion, low perfusion); ±3% (70-100%, Neonate,non-motion); ±3% (70-100%, motion); 0-69%,unspecified Refreshing Rate: 1s Channel:2-channel or 4-channel ART: 0 to 300 mmHg PA: -6 to 120 mmHg CVP/RAP/LAP/ICP : -10 to 40 mmHg Measurement range: P1/P2 -50 to 300 mmHg Resolution:1mmHg Accuracy: ±2% or ±1mmHg, whichever is greater(without sensor) Sensitivity: 5uV/mmHg/V Impedance range: 300 to 3000Q Recorder (Optional) Built-in, Thermal dot array Horizontal resolution :16 dots/mm (25 mm/s paper speed) Vertical resolution:8 dots/mm Paper speed:25 mm/s, 50 mm/s Number of waveform channels:3 Operation Environment SpO2 Measurement range : 0-100% Resolution: 1% Perfusion index display Accuracy: ±2%(70-100%, Adult/Pediatric); ±3% (70-100%, Neonate); 0-69%,unspecified Refreshing Rate: 1s Power: AC 100-250V, 50/60Hz Temperature: 5-40C Humidity: <80% Patient Range: Adult, Pediatric, Neonate Add: 4th Floor, Building C, Jin Wei Yuan Industrial Park, Julongshan Area, Kengzi St, Pingshan District, Shenzhen, P.R.China Tel: +86 755 23090266 Fax: +86 755 23010273 Email: [email protected] Website: http://www.northernmeditec.com

 Open the catalog to page 1
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.