AUTOLITH® (Legacy) Quick Reference Guide

AUTOLITH® (Legacy) Quick Reference Guide

AUTOLITH® (Legacy) Quick Reference Guide

Product catalog summary
Getting Started
To begin using the AUTOLITH System, ensure it is plugged into a proper power outlet and turn on the power switch. Connect the foot-pedal to the front panel and place the sterilized EHL Probe and Extender Cable onto the sterile field using proper sterile techniques.

Connecting the EHL Probe
Attach the EHL probe to the extender cable, ensuring the connection is opposite the square black ferrite. The remaining end of the extender cable should be passed to the O.R. circulator.

EHL Probe Recognition
The O.R. circulating nurse connects the extender cable to the front panel "PROBE" port. The system will recognize the probe size and adjust the energy delivery accordingly.

The Endoscope and EHL Probe
The doctor should insert the EHL probe into the endoscope's working channel to prepare for stone fragmentation. Use normal physiological saline for irrigation.

Power Settings
Start with a low power setting and increase as needed. The power range is 10% to 100%, with a suggested starting point of 30%-40%.

BiPolar Electro-Hydraulic Lithotripsy
Indicated for urinary and biliary stones. Align the green lines to connect.

Ordering Information
Details on disposable BiPolar EHL Probes and accessories are provided, including sizes and part numbers.

Operational Notes
  • The OUTPUT light flashes with each shot.
  • Higher power settings may reduce probe life.
  • Optimal stone fragmentation occurs with the probe 1-2mm from the stone.
  • Use the Standby button to prevent unintentional firings.
  • Check the total shots discharged using the Total Shots button.

Frequency and Number of Shots
Adjust the frequency of shots based on stone movement. The highest frequency is 30 shots per second. For stones that move, use fewer shots at a higher frequency. The maximum is 60 shots over 2 seconds, with a suggested setting of 3-5 shots per foot-pedal depression.
See more

Catalog excerpts

AUTOLITH® (Legacy) Quick Reference Guide-1

The AUTOLITH ® System Quick Reference and Information Guide Indicated for: Urinary & Biliary Stones BiPolar Electro-Hydraulic Lithotripsy Getting STARTED ® Turn on the power “ON” switch after plugging the AUTOLITH System into a proper power outlet. Attach the foot-pedal to the front panel foot-pedal port and place the sterilized EHL Probe and Extender Cable onto the sterile field using proper sterile technique. Align the green lines to connect Connecting the EHL Probe Connect the EHL probe to the end of the extender cable opposite the square black ferrite. Pass the remaining end of the extender cable to the O.R. circulator. EHL Probe Recognition The O.R. circulating nurse should then connect the remaining end of the extender cable to the front panel “PROBE” port. Notice the “Connect Probe” message change to indicate the size of the EHL probe that has ® been connected, i.e. 1.9Fr. (.63mm). The AUTOLITH System will now deliver the precise amount of energy specific to the size of the connected EHL probe. The Endoscope and EHL Probe The doctor should now prepare to fragment the stone. Careful insertion of the EHL probe into the appropriate endoscope working channel is now accomplished and treatment of the stone is ready to start. Using direct visualization, the EHL probe tip and the stone are sighted. Normal physiological SALINE (premixed bags of .9% normal saline) should be used for irrigation at all times. Power Start LOW and increase the power if needed for effective stone fragmentation. The total power range is 10% to 100% Suggested to start: 3

 Open the catalog to page 1
AUTOLITH® (Legacy) Quick Reference Guide-2

Frequency of Shots This adjusts how fast the selected number of shots are delivered. For stones that are free to move or bounce away, select fewer shots at a higher frequency. A rate of 30 shots per second is the highest (fast) frequency setting. A rate of 1 shot per second is the slowest frequency setting. Suggested setting: 30/sec. Number of Shots Again, if the stone can move or bounce away, select fewer shots. A maximum of 60 shots delivered over 2 seconds is the maximum number of shots that can be fired with 1 foot-pedal depression. Suggested setting: 3-5 shots per foot-pedal depression....