Class II, Type A2 Biological Safety Cabinets The CellGard Class II Biological Safety Cabinet can be either exhausted back into the room or connected to a facility HVAC system. The cabinet's airflow is 30% exhausted / 70% recirculated to minimize cross-contamination of low to moderate risk biologicals in the absence of volatile toxic chemicals. Features To Increase Your Productivity Zero Leak Airflow System Unidirectional Airflow Large HEPA Filters Paper-Catch Screen Shell / Gaskets Under Negative Pressure Front Filter Removal Metal Diffuser Over Supply Filter Energy Saver DC ECM Motor...
Open the catalog to page 2
Improved Principles of Biological Safety Cabinet Design ^. HEPA FILTERED AIR ». CONTAMINATED WORK SURFACE AIR ► ROOM AIR CONTAMINATED EXHAUST AIR True Laminar Airflow Unidirectional airflow moving along parallel flow lines at a constant velocity minimizes air turbulence within the work zone. HEPEX™ Zero Leak Airflow System The HEPEX™ Zero Leak Airflow System eliminates the possibility of cabinet and gasket leaks. Greater Volume of Air Creates a Stronger Air Barrier Air exchanges within the Class II Biological Safety Cabinet occur Larger, Long Lasting HEPA Filters NuAire's metal framed HEPA...
Open the catalog to page 3
Cabinet Control Systems Features TouchLink NU-480 • Blower Motor Controlled Via Solid State DC Motor Controller • onitor, Display and Control Downflow Via Digital Dual M Thermistor Airflow Sensor • larm Setpoints, High/Low for Error Conditions A (Downflow and Exhaust Flow ) • Remote Alarm Contacts • Date/Clock Display, and Timer Function • Control Lights Via Solid State Switch • Control Outlets Via Solid State Switch • Perform Complete Diagnostic Functions • Password Protection nitecare™- A unique system initiated by the window closure, will reduce motor / blower operational airflow to...
Open the catalog to page 4
CellGard Class II Biological Safety Cabinet Front HEPA Filter Removal Exhaust HEPA Filter TOUCHLINK System Sliding Tempered Safety Glass Window (Closes Completely) Frameless Polished Edge Window Ergonomic Armrest Base Support Stand External Fluorescent Lighting Glass Side Walls (Optional) Removable Solid SST Work Tray Drain Valve Base Stand Options Motorized Adjustable Base Stand Base Stand with Storage Cabinet For more information contact NuAire, Inc. at 1.800.328.3352 or www.nuaire.com
Open the catalog to page 5
Critical Cleaning Components Equipment Design Features critical cleaning Critical Cleaning Components means the combination of high-quality equipment, proper procedures, and appropriate cleaning aids which combine to help ensure a laboratory environment free from contamination. Cleaning Aids Cleaning aids such as the Mini AlphaMop™ Isolator Cleaning Tool, 12" x 12" Sterile Wipe" and 9" x 9" AlphaWipe® Sterile, Low-Particulate wipes help personnel clean effectively. Hinged Wing Window Design 100% Stainless Steel Interior 90% Stainless Steel Overall Construction Stainless Steel Grills and...
Open the catalog to page 6
The best fit for your lab • Experience Ergonomics The cabinet design maximizes knee/thigh clearance to improve posture and the adjustable base stand allows optimization for leg and forearm support. i Forearm Support for Comfort and Safety ■ • Improves forearm support, keeps arms off front air grill * Closed cell foam, easy to clean • Closed cell non-absorbing foam • Disposable, easy application and removal (10 pads / package) Expanded Vision Zone Reduces Awkward Postures Reduces Glare Cabinets incorporate cool white lighting and frameless polished edge windows to allow for greater...
Open the catalog to page 7
Energy Saver Technology Motor Efficiency Higher RPMs represent filter loading from particulates in the laboratory. Filter Loading Capacity DC ECM technology increases filter life from the NSF minimum requirement of 50% or an approximate 3 year equivalent life to 10 years or more. Percent increase in total load capacity*: *Percent increase testing based on NSF/ANSI 49, ANNEX A.12 motor/blower performance test methods. 4 Foot Type A2 BSC that runs 24/7 (8736 hours per year) plus the energy required to control the laboratory ventilation by adding the rejected heat. Experience energy savings...
Open the catalog to page 8
CellGard ES offers an exclusive system to aid in current decontamination techniques such as Formaldehyde, H202, or CD systems. The window sash closes completely with the simple removal of the armrest allowing the window to rest on a gasket installed on the inside of the window frame. Window clamps fit securely on the outside of the window to provide the necessary pressure against the gasket for a tight seal. A supply inlet port is provided on top of the cabinet near the blower for the decontaminant introduction. An exhaust seal plate installs over the exhaust filter housing and is locked in...
Open the catalog to page 9
Style of Cabinet (All Sizes) All Models: Bench Top / Console with Base Stand / Storage Cabinet Cabinet Construction (All Sizes) All Models: All Welded 16 GA. Type 304 Stainless Steel, Pressure Tight Design Diffuser for Air Supply (All Sizes) All Models: Non-Flammable (Metal) HEPA Filter Seal Type Supply Filter 99.995% Eff. on .03 microns Exhaust Filter 99.995% Eff. on .03 microns HEPEX™ Seal Neoprene, Springloaded Performance Specifications Standard Services Service Coupling (3/8 inch NPT) Optional Services Remote Controlled Valves* Ultraviolet Light Left or Right Work Surface Cabinet Size...
Open the catalog to page 10
Options, Accessories, and Sales Variable Flow Exhaust Transition Exhaust Transition Butterfly Valves Flex Duct Exhaust Airflow Monitors Round Transition To Locate a NuAire Sales Team Member Please visit NuAire's Web site at: www.nuaire.com and select the 'Sales Team' menu item. Information can be found there for both United States and international distributors. Other NuAire Products Animal Handling Products For more information contact NuAire, Inc. at 1.800.328.3352 or www.nuaire.com
Open the catalog to page 11All NuAire Laboratory Equipment catalogs and technical brochures
-
PharmaGard Compounding Isolators
20 Pages
-
LabGard ES LP NU-640 AH BSC
12 Pages
-
LabGard ES NU-677
8 Pages
-
HEPA versus ULPA Filters
7 Pages
-
NU-543
20 Pages
-
NU-540
20 Pages
-
NuWind Centrifuge Tube Chart
3 Pages
-
NU-5700 CO2 Incubator
2 Pages
-
FumeGard NU-156
2 Pages
Archived catalogs
-
General Purpose Centrifuge
24 Pages
-
FumeGard NU-164
4 Pages
-
FumeGard NU-162
4 Pages
-
ADVANCED SAFETY ENGINEERING
6 Pages
-
General_Overview_Brochure
4 Pages