Objective
The study aimed to evaluate the effectiveness of a new ultraviolet ray device developed by Nuova ASAV snc in reducing microbial and fungal loads in the air of a closed environment. The tests were conducted in a microbiology laboratory, a critical environment for hygienic-sanitary evaluation.

Test Protocol
The tests were conducted in a laboratory room with specific dimensions and equipment. The air sampling was performed using a SAS Surface Air System sampler at two distinct points in the room. Initial sampling was done without the device to establish a baseline, followed by sampling at various intervals after the device was activated.

Results
The results showed significant reductions in microbial contamination levels over time. Key findings include:
  • Colony count at 30°C: Over 80% reduction after 60 minutes, reaching 100% in Point 2 after 240 minutes.
  • Coagulase-positive staphylococci: Over 80% reduction after 50 minutes, exceeding 90% after 60 minutes.
  • Coliforms: Over 80-90% reduction after 20 minutes, reaching 100% after 30 minutes.
  • Moulds: Over 80% reduction after 50 minutes.
  • Yeasts: 100% reduction in Point 1 after 50 minutes.


Conclusions
The study concluded that the ultraviolet device effectively reduces microbial loads, particularly Staphylococci and Coliforms, within 50-60 minutes of operation. The device's continuous use offers protection for staff in environments with potential air contamination sources, without the need for chemicals or harmful by-products.
See more

Catalog excerpts

STUDY OF DISINFECTANT ACTION SANITIZING TOOL FOR AIR MODEL DNA-1

STUDY OF DISINFECTANT ACTION SANITIZING TOOL FOR AIR MODEL DNA C.A. FRANCO LEONI NUOVA ASAV SNC VIA DELL’INDUSTRIA, 44, 42025 CAVRIAGO (RE) – PIVA E CF: 0074191035

 Open the catalog to page 1
STUDY OF DISINFECTANT ACTION SANITIZING TOOL FOR AIR MODEL DNA-2

Ali Lab srl Via del Quaresimo, 6 42123 Reggio Emilia – ITALIA Microbiology and Quality manager: Dr.ssa Miriam Fornaciari Ali Lab srl – Via del Quaresimo 6, 42123 Codemondo (RE) – P.IVA 02777690351 Telefono: 0522-372363 – email: info@alil

 Open the catalog to page 2
STUDY OF DISINFECTANT ACTION SANITIZING TOOL FOR AIR MODEL DNA-3

OBJECTIVE By carrying out the following study, we wanted to test the reduction of microbial and fungi load in the air of a closed environment by a new model of ultraviolet ray device developed by the Nuova ASAV snc company. All tests were carried out in a Microbiology Laboratory. The chosen environment is a place where there is the possibility of contracting infections, a risk mainly due to exposure to a wide range of types of agents, from bacteria to fungi. It was deliberately decided to go and evaluate the action of the machine during a period of work within a very critical environment from...

 Open the catalog to page 3
STUDY OF DISINFECTANT ACTION SANITIZING TOOL FOR AIR MODEL DNA-4

The use of the SAS provides for the positioning of the Contact petri dishes in the special housing positioned under a perforated cover (disinfected each time with sodium hypochlorite). 180 liters of air were sucked onto each plate and the results obtained from the sampling were brought back to 1 m3. The air samplings were carried out in two different areas of the laboratory: • Point 1: Half room near window side; • Point 2: Near door side. The samplings were always carried out with operators present during their work activities. The first sampling was carried out without the use of the DNA tool...

 Open the catalog to page 4
STUDY OF DISINFECTANT ACTION SANITIZING TOOL FOR AIR MODEL DNA-5

RESULTS The results of the different air samples are summarized in Table 1, where are reported the UFC / m3 and the percentage reduction related to the microbiological parameters detected in the Laboratory air at the two different sampling points. Table 2 indicates the average of the percentage reduction for the various microbiological parameters considered. Table 1. Microbial contamination levels and % reduction for Points 1 and 2. The percentage reduction varies differently depending on the parameter considered: Colony count at 30°C A reduction of more than 80% is achieved after 60 minutes...

 Open the catalog to page 5
STUDY OF DISINFECTANT ACTION SANITIZING TOOL FOR AIR MODEL DNA-6

Table 2 . Medium % reduction. The table does not include data regarding the spores of the reducing sulfite Clostridia as they were not detected in the environment as early as the beginning of the sampling, i.e. with the tool off. CONCLUSIONS First, it is interesting to note the initial situation of the Laboratory air. The sampling showed a limited contamination of the environment mainly by Colony count at 30°C, Moulds and Staphylococci. The charges of Coliforms and Yeasts were low, and the Sulfite-reducing bacteria were completely absent. In Table 3 it is possible to see the % reduction relating...

 Open the catalog to page 6
STUDY OF DISINFECTANT ACTION SANITIZING TOOL FOR AIR MODEL DNA-7

Table 4. Average % reduction trend. Both in Table 3 and in Table 4 it is noted that from 50 minutes onwards the blast chilling power of the machine begins to become predominant, with percentages ≥ 80%. If the starting concentrations proved to be very high (such as molds and total microbial load) the abatement power is greater than 90% after 180 minutes. The results show how DNA, already after 50-60 minutes of ignition, manages to completely break down Staphylococci and Coliforms and begins to work consistently also on Colony count at 30°C and Moulds. The performance of the yeasts is not very...

 Open the catalog to page 7