Overview: This document is a self-study guide for the O-Two eSeries Transport Ventilator, focusing on abbreviations, alarms, and battery status. It provides essential information for understanding the ventilator's functions and alerts.
Abbreviations: The guide lists and explains various terms related to ventilation modes and parameters, such as Assist Control Ventilation (A/C V), Breathing Circuit Integrity (BCI), Continuous Positive Airway Pressure (CPAP), and others. These terms are crucial for understanding the ventilator's operation and settings.
Alarms: The document details the visual and audible alarms associated with different conditions, such as patient circuit disconnect, low/high airway pressure, and apnea. Each alarm is categorized by priority, with specific symbols and LED indications for high-priority alarms, such as flashing red symbols and multiple audible pulses.
Battery Status: The guide explains the battery status indicators, which show both discharging and charging states. It highlights the critical warning that the ventilator will not start or will shut down at approximately 2% battery capacity. The status is visually represented on the display, with color changes indicating different levels of battery charge.
Contact Information: The document provides contact details for O-Two Medical Technologies Inc., including phone numbers, email, and website for further assistance or inquiries.
See more

Catalog excerpts

eSeries Transport Ventilator Abbreviations, Alarms and Battery Status-1

O-Two Self-Study Guide eSeries Transport Ventilator Abbreviations, Alarms and Battery Status

 Open the catalog to page 1
eSeries Transport Ventilator Abbreviations, Alarms and Battery Status-2

e7Series Abbreviation, Alarms and Battery Status Abbreviations Term A/C V BCI BiLVL CPAP CPR Rate Termination I:E LED Mve Mv O2 (%) Paw (AV) Paw (peak) PCV PEEP Pi Pmax P min PSV SIMV T APNEA Te TFT Ti Trig. Vte Vt Explanation Assist Control Ventilation (see ventilation modes self-study guide) Breathing Circuit Integrity (Patient circuit disconnect) Bi-phasic Positive Airway Pressure Ventilation (see ventilation modes self-study guide) Continuous Positive Airway Pressure (see ventilation modes selfstudy guide) Cardio Pulmonary Resuscitation (see ventilation modes self-study guide) Ventilation...

 Open the catalog to page 2
eSeries Transport Ventilator Abbreviations, Alarms and Battery Status-3

e7Series Abbreviation, Alarms and Battery Status Alarms Symbol Patient Circuit disconnect (Breathing circuit Integrity) Visual Alarm Audible Alarm Alarm symbol High after 15 seconds Flashing symbol Flashing Yellow for 15 seconds then Red Flashing symbol High Airway Pressure Flashing symbol Flashing symbol High Minute Volume Flashing symbol Blocked airway Flashing symbol Two bursts with five pulses Flashing symbol Two bursts with five pulses Flashing symbol Two bursts with five pulses Flashing symbol Two bursts with five pulses Flashing symbol One burst with three pulses High input pressure ≥...

 Open the catalog to page 3
eSeries Transport Ventilator Abbreviations, Alarms and Battery Status-4

e7Series Abbreviation, Alarms and Battery Status Battery Status Battery status will be displayed in section 1 of the display. There are two different status indicators showing battery discharging and charging status respectively. Battery discharging status Full Capacity Approx. 75% of full capacity Approx. 50% of full capacity Approx. 25% of full capacity On screen symbol change to Yellow color and flashing Approx. 5% of full capacity On screen symbol change to Red color and flashing with associated Red color warning LED. Battery charging status 1 Full Capacity At approximately 2% of full battery...

 Open the catalog to page 4
eSeries Transport Ventilator Abbreviations, Alarms and Battery Status-5

e7Series Abbreviation, Alarms and Battery Status O -Two Medical Technologies Inc. 45A, Armthorpe Road, Brampton, Ontario, Canada L6T 5M4 Toll Free: (800) 387 3405 Tel: (905) 792 OTWO (6896) Fax: (905) 799 1339 E-mail: [email protected] Website: www.otwo.com NOTE: Refer to the product manual for further definitions and descriptions of function.

 Open the catalog to page 5

All O-Two Medical Technologies catalogs and technical brochures

  1. Statvac II

    2  Pages

  2. CAREvent PAR

    2  Pages

  3. CAREvent ALS+

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.