O-Two Self-Study Guide eSeries Transport Ventilator Control and Function Keys and Connections

Open the catalog to page 1
eSeries Transport Ventilator Control and Function Keys and Connections - 2

eSeries Control and Function Keys and Connections Control and Display Layout N M B – Battery operation indicator C – External power indicator D – Battery charging indicator E – Manual / Hold button I – Lock button J – Waveform selection button K – Cancel button L – Confirmation indicator light F – Set screen brightness button G – Alarm Silence button M – Control Selection Knob N – ON/OFF button NOTE: Refer to the product manual for further definitions and descriptions of fun

Open the catalog to page 2
eSeries Transport Ventilator Control and Function Keys and Connections - 3

eSeries Control and Function Keys and Connections Function Keys While all ventilation parameter settings are controlled by the Control Selection Knob (M) in Figure 1, there are a number of key membrane buttons which control additional ventilator functions: 3.2.1 ON/OFF To turn on the ventilator, Press the ON/OFF button (N) in Figure 1 for one second, during that second the associated green LED will start flashing at a high frequency. After 1 second the ventilator will turn on but without ventilation until the appropriate patient size symbol is selected. If the button is pressed and released...

Open the catalog to page 3
eSeries Transport Ventilator Control and Function Keys and Connections - 4

eSeries Control and Function Keys and Connections Key Membrane Controls 3.2.3 Lock Lock function will disable all buttons and the Control Selection Knob except for the ON/OFF, Alarm Silence and Day/Night buttons, which are enabled at all times. To lock the key membrane or cancel the lock function: 1- Press Lock button (I) in Figure 1. The Lock symbol will be displayed on the screen. 2- To cancel the Lock function, press the Lock button (I) again. Note: During lock function, if any locked button is pressed, the Lock symbol will flash while the locked button is pressed. 3.2.4 Alarm Silence...

Open the catalog to page 4
eSeries Transport Ventilator Control and Function Keys and Connections - 5

eSeries Control and Function Keys and Connections 3.2.7 Cancel The Cancel button (K) in Figure 1 allows the operator to return to the previous settings if the last unconfirmed changes in settings are not required. 3.2.8 Pause/ Resume During activation of the Pause/ Resume button (H) in Figure 1, the ventilator will stop ventilating with all buttons kept active (if they are not locked) except Manual/ Hold button. To activate pause function, proceed as follows: 1. Press Pause/ Resume button (H). The Pause symbol will flash on the screen with the confirmation symbol and the Confirmation...

Open the catalog to page 5
eSeries Transport Ventilator Control and Function Keys and Connections - 6

eSeries Control and Function Keys and Connections 3.2.9 Manual/ Hold During exhalation phase, if Manual/ Hold button (E) in Figure 1 is pressed, a mandatory breath will be initiated and either the flow rate or set pressure control parameter will be delivered as long as the Manual/ Hold button is pressed or until I-time setting is achieved. After I-time if the button still pressed, the ventilator will switch to inspiratory hold function in which the ventilator will cut the flow but will keep the exhalation port closed in order to block exhaled gas from going to ambient resulting in the...

Open the catalog to page 6
eSeries Transport Ventilator Control and Function Keys and Connections - 7

eSeries Control and Function Keys and Connections O-Two Medical Technologies Inc. 45A, Armthorpe Road, Brampton, Ontario, Canada L6T 5M4 Toll Free: (800) 387 3405 Tel: (905) 792 OTWO (6896) Fax: (905) 799 1339 E-mail: resuscitation@otwo.com Website: www.otwo.com NOTE: Refer to the product manual for further definitions and descriptions of function. Rev 1 28_02_2018