Overview: The O-Two Statvac® II is an oxygen-powered aspirator designed for emergency situations where there is a risk of airway compromise due to fluids or foreign materials. It effectively suctions excessive fluids or particulates from the oro/naso-pharynx.
Key Features:- Fast and effective fluid and foreign material removal.
- Unique ON/OFF shuttle valve for quick operation.
- Lightweight and durable with an anodized aluminum body.
- Available with 8 oz or 12 oz collection jars.
Specifications:- Supply Pressure: 50 PSI (3.4 BAR)
- Vacuum Pressure: 300mm Hg +/- 10%
- Vacuum Free Airflow: 30 L/min +/- 15%
- Maximum Gas Consumption: 20 L/min
- Operating Temperature: -18 to 50°C (0 to 122°F)
- Storage Temperature: -40 to 60°C (-40 to 140°F)
- Weight: 5.6 oz (160 gms)
- Dimensions: Diameter 3.0 in. (76mm) x Height 7.0 in (178mm)
Instructions for Use:- Connect the gas supply hose to the DISS inlet and a 50 PSI oxygen supply.
- Turn on the oxygen supply.
- Attach suction tubing to the aspirator and secure the collection jar lid.
- Press the green shuttle button to start suction and ensure vacuum creation.
- Keep the collection jar upright to prevent blockage.
- Clear the suction catheter by covering the discharge port if blocked.
- Limit suctioning to five seconds to avoid depleting patient oxygen levels.
- Press the red shuttle button after suctioning to conserve oxygen.
- Repeat as necessary, emptying the collection jar as needed.
Cleaning and Disinfection:- Rinse the aspirator jar with water and suction clean, soapy water through the aspirator.
- Turn off the gas supply, remove the hose, and dispose of the gasket and tubing.
- Wash the jar and lid in soapy water, rinse, and immerse in a cold liquid sterilant.
- Rinse thoroughly, inspect for damage, and replace parts as needed.
- Dry and reassemble, replacing the gasket.
- Test the aspirator for function and return to service center if not functioning.
- Replace suction tubing and catheter as required.
Ordering Information:- 01AS9500: Statvac II with 8 oz jar
- 01AS9502: Statvac II with 12 oz jar
- 17MP3051-cs: Replacement 8 oz jar (Case of 10)
- 17MP3065-cs: Replacement 12 oz jar (Case of 10)
- 17MP3052-cs: Replacement lid gasket (Case of 10)
Contact Information: O-Two Medical Technologies Inc., 7575 Kimbel Street, Mississauga, ON, Canada, L5S 1C8. Phone: +1 905 677 9410, Toll-Free: +1 800 387 3405, Fax: +1 905 677 2035, Email:
[email protected], Web: otwo.com