10 Condition for Transportation and Storage

Open the catalog to page 2
Soft Chamber User Manual - 3

people cannot do without air, especially oxygen. The oxygen percentage in the normal air pressure is 21%. People will suffer a hyporoxic death if oxygen is lower than 12%. 90% of the people in the city are in various degrees of hypoxia. Hyperbaric oxygen therapy improves the quality of life and keeps us more healthy. It is the latest trend in health and beauty. hyperbaric oxygen chamber

Open the catalog to page 3
Soft Chamber User Manual - 4

The following table is a list of symbols and definitions that used with the Hyperbaric Chamber. Type B Applied part Serial number Mains fuse specification AUTHORISED REPRESENTATIVE IN THE EUROPEAN COMMUNITY Refer to instruction manual/booklet Protective earth (ground) HOT SURFACE DO NOTTOUCH Allow to cool before servicing Correct Disposal of This Product (Waste Electrical & Electronic Equipment)Statement: Contact the local authorities to determine the proper method of disposal of equipment and its part Air deflate valve indicator internal/external Caution Caution Keep dry from rain

Open the catalog to page 4
Soft Chamber User Manual - 5

3 Safety Notice In order to ensure safety, please read the following precautions before use warning: Abnormal smell during use, please turn off the power and pull out the plug;Please make sure the socket plug is fully inserted; Please do not disassemble the machine without permission from the supplier; This product needs to stay away from water and fire and sharp objects; Children need to use under adult care; Please do not touch any valves on the chamber while using; Please do not place heavy objects on the machine; Do not use if the wire is loose; Do not pull or twist the power cord; Hand...

Open the catalog to page 5
Soft Chamber User Manual - 6

model specification

Open the catalog to page 6
Soft Chamber User Manual - 8

5 Product assembly and operation Working Scheme Installation 1. Please open the accessory box before installation accessories Black metal frame and connecting rod air tube and oxygen tube anti-roll bolsters internal/external pressure gauge activated carbon filter cotton power cable hexagon wrench spare fuse

Open the catalog to page 8
Soft Chamber User Manual - 9

Connect the air tube on the air compressor to the air inlet valve of the chamber Connect the oxygen tube on oxygen concentrator to the oxygen inlet valve of the chamber. Install internal&external pressure gauges. external pressure gauge Internal pressure gauge (-100-0kPa) (0-100kPa) Note: Please install internal and external pressure gauges before use (the internal pressure gauge should be installed inside the chamber, external pressure gauge should be installed outside the chamber) Install the metal frame (needs to be installed inside the chamber, it is recommended to take all frame pieces...

Open the catalog to page 9
Soft Chamber User Manual - 10

4、Silencer mounted to the air inlet valve inside the chamber Oxygen Mask/ headset mounted to the oxygen inlet valve inside the chamber The oxygen inlet Backup Valve Oxygen mask 5 、Put in mattress and put anti-roll bolsters under the chamber. 6、Installation of Air Cooler Connect the Tubes with Air Compressor, Cooler and Chamber. Please connect air tubes from Air Compressor to the Wet Air In port of Cooler and connect air tubes from Chamber to the Dry Air Out port of Cooler. There is a water drainer on one side of the Air Cooler. Unscrew the black valve when the drainer is full of water.

Open the catalog to page 10
Soft Chamber User Manual - 11

2、Get into the chamber, pull inside and outside zippers and close the air deflate valve (make sure to flatten out the silicone strip between the zippers) Pull the inside and outside zippers Close the external pressure relief valve clockwise Close the internal pressure relief valve anti-clockwise Zippers and Air Deflate Valve can be operated both inside and outside the chamber. 3、Lie down and wear oxygen headset or mask. 4、In about 8 mins, the chamber pressure gauge will slowly rise up to 30kPa, at this time the automatic pressure release valve will exhaust air automatically. (Pressure gauge...

Open the catalog to page 11
Soft Chamber User Manual - 12

5 It is recommended to use 60-80 mins per session. 6、After use, please open the Air Deflate Valve to reduce pressure first. open the external air deflate valve anti-clockwise Open the internal air deflate valve clockwise When the pressure gauge shows 0kPa, open the zippers and get out of the chamber (0-100) (-100-0) After getting out of the chamber, turn off the air compressor, oxygen concentrator and air cooler.

Open the catalog to page 12
Soft Chamber User Manual - 13

Precautions when using When using for the first time, please test the empty chamber first and observe whether the equipment is operating normally. If the equipment is running abnormally, please do not use it. Please contact your dealer promptly. When you need to get out of the chamber, wait until the pressure gauge shows 0kPa before you open the inner and outer zippers. Do not force the zipper to open during use or when pressure relief is not complete. In the event of an emergency or if the air deflate valve fails to operate properly, push the emergency relief valve from inside the chamber...

Open the catalog to page 13
Soft Chamber User Manual - 14

7 maintenance 1. Chamber maintenance and cleaning Clean the chamber: After using a deodorizing and antibacterial spray, or after diluting it twice with alcohol, wipe it and le it dry. Drive the machine to run for 20 minutes. 2. Air compressor maintenance and cleaning It is recommended to replace the activated carbon filter cotton every 6 months. Replace Filter Complete Replacement

Open the catalog to page 14
Soft Chamber User Manual - 15

8 FAQ (1) What should I do to prepare for a treatment? People can go into the chamber fully clothed. It is recommended to wear loose comfortable. Shoes, jewelry, or any sharp objects should be removed before getting into the chamber. It is recommended that women not wear stockings in the chamber, because it may cause feelings of claustrophobia. (2) How will the HBOT treatment feel? The chamber will be filled with compressed air. Once the chamber has expanded to its filled size, it will start to pressurize. While the chamber is pressurizing, your ears will feel the change in pressure inside...

Open the catalog to page 15
Soft Chamber User Manual - 16

(8) Do I need a prescription or supervision to use the chamber? No, you do not need a prescription to use the Portable Hyperbaric Chamber. If you are using the chamber to treat your health condition, we recommend you seek advice from your health practitioner. The Portable Hyperbaric Chamber is easy to use without the need for any supervision or specialised training. Pressure gauges and pressure release valves (9) Are there any contraindications for Hyperbaric Oxygen Therapy? Never use the chamber while you catch a cold or flu symptoms. Never use the chamber if you have rhinitis or otitis...

Open the catalog to page 16