1. Catalogs
  2. OES Medical
  3. Lynx 15 Patient Monitor

Lynx 15 Patient Monitor

Lynx 15 Patient Monitor

Lynx 15 Patient Monitor

Product catalog summary
Overview: The document provides detailed technical specifications and features of the Lynx 15 Patient Monitor by OES Medical Limited. It includes information on various monitoring capabilities, standard configurations, optional features, and technical specifications.
SpO2 Monitoring: Utilizes a dual-wavelength optical method with a measuring range of 0% to 100% and accuracy within 2% for SpO2 levels between 70% and 100%. Pulse rate (PR) is measured within a range of 30 to 250 bpm with an accuracy of ±2 bpm or ±2%, whichever is greater.
Cerebral State Monitoring (CSM): EEG sensitivity is ±400µV with a noise level of less than 2µVp-p. Input impedance is greater than 50Mohm. The system updates every second with various filters applied for different parameters like CSI, EMG%, and BS%.
Options and Standard Configuration: Optional features include a touch screen, 2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, 12-lead ECG, and more. The standard configuration includes ECG, respiration, SpO2, PR, NIBP, and temperature monitoring.
Power and Display: The device operates on AC 100V-240V power supply with a built-in lithium battery (11.1V/4400mAh). It features a 15-inch TFT display with a 3-level audible-visible alarm system.
Cardiac Output and IBP: Cardiac output measuring range is 0.2 to 20 L/min with an accuracy of ±0.2 L/min or ±10%. IBP uses a strain gauge transducer with a measuring range of -50 to 300 mmHg and accuracy of ±2% or ±4 mmHg.
ECG and CO2 Monitoring: ECG input dynamic range is ±(0.5mVp~5mVp) with various bandwidths for diagnostic, monitoring, and operation modes. CO2 is measured using an infrared optical method with a range of 0 to 150 mmHg.
RESP, TEMP, and NIBP: RESP measuring range is 0 to 120 rpm with accuracy of ±5% or ±2 rpm. TEMP range is 21.0 to 50.0℃ with accuracy of ±0.2℃. NIBP uses an oscillometric method with various measuring ranges for adults, pediatrics, and neonates.
Features and Connectivity: The Lynx 15 offers user-customized NIBP measuring cycles, versatile clinical calculations, and software upgrades. It supports HL7 protocol for hospital system connectivity and features a 15” display with LED backlight and a 360-degree visible indicator.
See more

Catalog excerpts

Lynx 15 Patient Monitor-1

Input dynamic range: ±(0.5mVp—5mVp) Differential input impedance: >I0MQ 0.5~40Hz (Monitoring) 1~20Hz (Operation) CMRR: >90dB (Diagnostic) >I05dB (Monitoring & Operation) Sensitivity selection: x|/4, x|/2, xI, *2, x4 an d Auto HR accuracy: ±1% or ±2bpm, whichever is greate Pacemaker pulse detection and rejection function Measuring accuracy: ±5% or ±2 rpm, whichever is greater Measuring accuracy: ±0.2 Cfrom 25—45 C Technique: Oscillometric method Typical measurement time : <30 seconds (adult cuff) NIBP measuring range : 40—200mmHg (Pediatric) NIBP measuring range: I0—I50mmHg (Pediatric) NIBP measuring range: 20—I65mmHg (Pediatric) 20—II0mmHg (Neonate) NIBP measuring accuracy: Mean difference: ±5mmHg NIBP measurement mode: Manual, Auto, STAT, Multi-cycle mode Auto measuring intervals: I-480min Technique: Dual-wavelength optical method Measuring accuracy: Arms is not greater than 2% for SpO2 range 70—I00%. PR measuring range: 30—250bpm PR measuring accuracy: ±2bpm or ±2%, whichever is greater Technique: Infrared optical method Sampling mode: Sidestream or Mainstream 4I~70mmHg ±5% of reading 7I~I00mmHg ±8% of reading I0I~I50mmHg ±I0% of reading Flow rate: 50ml/min ±I0 ml/min (Sidestream) CSI and update: 0-I00. filter: 6-42Hz, I sec. update filter: 75-85 Hz, I sec. update. BS%: 0-I00. filter: 2-42 Hz, I sec. update Technique: Strain gauge transducer Measuring accuracy: ±2% or ±4mmHg, whichever is greater ICP, AUXPI, AUXP2 Calibration: zero calibrating Blood temperature measuring: range: 23-43 C, accuracy: ±0.5 C Injecta temperature measuring: range: 0-20 C, accuracy: ±0.5 C Measuring range: 0.2—20 L/min Measuring accuracy: ±0.2 L/min or ±I0%, whichever is greater Built-in lithium battery: II.IV/4400mAh Display: I5 inch TFT display Alarming method: 3 levels audible-visible alarm Networking: Ethernet ECG, Respiration, SpO2, PR, NIBP, Temperature Touch Screen, 2-IBP, EtCO2, Nellcor SpO2, SunTech NIBP, I2-lead ECG, Cardiac Output, Cerebral State Monitoring, Central Monitor Station, Multi-Gas Monitoring Lynx 15 Patient Monitor OES Medical Limited ABC House, Cotswold Dene, Standlake Witney, Oxfordshire, OX29 7QG Tel: +44 (0)1865 301711 Fax: +44 (0)1865 301573 e-mail: [email protected] www.oes-medical.co.uk

 Open the catalog to page 1
Lynx 15 Patient Monitor-2

15” display with LED backlight 9-waveform on screen 360-degree visible indicator with 3-level alarm Patient Monitor Accessory box for standard configuration Integral 3-channel thermal recorder Li-ion battery up to 4 hours continuous monitoring Parameter case for optional parameters Comprehensive calculations for clinical application 15” high resolution display Touch screen optional 9 traces on-screen waveforms and maximal up to 13 User customized NIBP measuring cycles up to 5-phase Data export and software upgrade Versatile clinical calculations for application convenience HL7 protocol, Bed to...

 Open the catalog to page 2