1. Catalogs
  2. Ohio Medical
  3. Push-To-Set TM Intermittent and Continuous Vacuum Regulators

Push-To-Set TM Intermittent and Continuous Vacuum Regulators
1 /12Pages

Push-To-Set TM Intermittent and Continuous Vacuum Regulators

Push-To-Set TM Intermittent and Continuous Vacuum Regulators
1 /12Pages

Catalog excerpts

Push-To-Set TM Intermittent and Continuous Vacuum Regulators-1

Ohio Medical’s Push-To-SetTM Intermittent and Continuous Vacuum Regulators

 Open the catalog to page 1
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-2

Ohio Medical Corporation’s Push-To-SetTM (PTS) family of vacuum regulators come with our durable, time tested, quality design you have come to know. Top mounted on/off switch for ease of use Clear markings show what mode the unit is in Stat Set-Up Intermittent mode starts in the “ON” cycle - No waiting Smaller, sleek design for tight spaces Glow-in-the dark gauge provides visibility in low-light situations Fully adjustable timing cycles preset to 15 seconds On / 8 seconds OFF Quick-To-Max Only 2 turns from 0 to full wall vacuum Patented Push-To-SetTM Built-in feature eliminates the need to occlude...

 Open the catalog to page 2
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-3

now? You K Did Higher than recommen suction levels are frequently set for most recommended common suctioning procedures. Fact: If occlude-to-set is not performed, over-suctioning may occur* Fact: Higher than recommended suction levels may lead to tissue damage/bleeding* Fact: Tissue tears may lead to infection* Tracheal Why PTS dramatically improves setting clinical vacuum settings Patented Push-To-SetTM built-in feature eliminates the need to occlude patient tubing to set accurate vacuum levels Removes a step in setting vacuum level, while ensuring greater patient safety One simple One-handed...

 Open the catalog to page 3
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-4

Push-To-SetTM Adult Intermittent Suction Regulators (ISU) (Analog or Digital) Technical Specications Safety Feature: Push-To-SetTM Automatic occlusion device Quick-To-Max (2 turns to full wall vacuum) One hand to set vacuum level Flow Rate: Intermittent: Preset at 8 L/min per ASTM* Continuous: 0 to 80 L/min (without ttings at full increase) Easily adjustable without cover or gauge removal Silent ON/OFF switching Preset 15 seconds ON / 8 seconds OFF (± 3 seconds) Front case adjustable at bedside Begins in the “ON” mode Vacuum Range: 0 to 200 mmHg and full wall vacuum Gauge Display: Analog: Glow...

 Open the catalog to page 4
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-5

Push-To-SetTM Pediatric/Neonatal Intermittent Suction Regulators (ISU) (Analog or Digital) Technical Specications Safety Feature: Push-To-SetTM Automatic occlusion device Quick-To-Max (2 turns to full wall vacuum) One hand to set vacuum level Positive Pressure Safety Relief Valve Flow Rate: Intermittent: preset at 8 L/min per ASTM* Continuous: 0 to 80 L/min (without ttings at full increase) Easily adjustable without cover or gauge removal Silent ON/OFF switching Preset 15 seconds ON / 8 seconds OFF (± 3 seconds) Front case adjustable at bedside Begins in the “ON” mode Vacuum Range: 0 to 135 mmHg...

 Open the catalog to page 5
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-6

Push-To-SetTM Adult Continuous Vacuum Regulator (CVR) (Analog or Digital) High Three Mode Continuous (Adult) Modes: I (ON), O (OFF), MAX Gauge: High Vacuum (0-760 mmHg) Color Band: Solid orange/variable to orange Regulated Vacuum: 0- Full Vacuum Instantaneous Full Wall Vacuum Mode Standard Three Mode Continuous (Adult) Modes: I (On), O (OFF), MAX Gauge: Standard (0-200 mmHg - Full Vacuum) Color Band: Low, Medium, High & Full Vacuum Regulated Vacuum: 0-Full Wall Vacuum Instantaneous Full Wall Vacuum Mode Standard Two Mode Continuous (Adult) Modes: I (ON), O (OFF) Gauge: Standard (0-200 mmHg-Full...

 Open the catalog to page 6
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-7

Push-To-SetTM Pediatric/Neonatal/C.A.S.S. Continuous Vacuum Regulator (CVR)(Analog or Digital) Two Mode Neonatal Continuous Modes: I (ON), O (OFF) Gauge: Low Vacuum ( 0-100 mmHg/Digital) (0-160 mmHg/Analog) Color Band: Low, Medium, High Regulated Vacuum: 0-100 mmHg Two Mode C.A.S.S. Continuous Modes: I (ON), O (OFF) Gauge: Low Vacuum (0-160 mmHg) Color Band: Low, Medium, High Regulated Vacuum: 0-140 mmHg Two Mode Pediadtric Continuous Modes: I (ON), O (OFF) Gauge: Low Vacuum (0-160 mmHg) Color Band: Low, Medium, High Regulated Vacuum: 0-135 mmHg Technical Specications Safety features: Gauge Accuracy:...

 Open the catalog to page 7
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-8

Color Coding Labels Color Code your Push-To-SetTM ISU or CVR Vacuum Regulators. A solution for labeling your regulators to show department ownership or special clinical applications. Can be used on either Analog or Digital Regulators Intermittent Suction Unit 8 Information/products in this catalog subject to change without notice. 8700-0010-107 LIGHT GRAY (5 labels/order) Continuous Vacuum Regulator 8700-0010-110 DARK GRAY (5 labels/order)

 Open the catalog to page 8
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-9

Color Coding Color Code your Push-To-SetTM Digital ISU or CVR Vacuum Regulators. A solution for labeling your regulators to show department ownership or special clinical applications. Color cases are only available with a digital gauge. Additional colors are available for an additional charge. Please contact inside sales at 866-549-6446, Option 1 9 Information/products in this catalog subject to change without notice.

 Open the catalog to page 9
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-10

Ordering Matrix - Intermittent Suction Unit 87 XX - XXXX - 9 XX Push-To-SetTM Series Fittings (patient connection) No Fitting 1/8” NPT Female Tubing Nipple Locking Gland Locking Gland & Overow Safety Trap DISS Male DISS Male & Overow Safety Trap, DISS Wing Nut & Gland Pole Mount Bracket with Locking Gland & Overow Safety Trap Pole Mount Bracket with Tubing Nipple Adapter (vacuum source connection) No Adapter with 1/8” NPT Female 00 Diamond 01 DISS Hand-I-Twist (HIT) 02 DISS Nut & Gland 03 04 Schrader® Chemetron® Rectangle Striker (NCG) 05 Puritan Bennett® 07 O.E.S. Twist 08 O.E.S. Med Star® 09...

 Open the catalog to page 10
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-11

Ordering Matrix - Continuous Vacuum Regulator 87 XX - XXXX - 9 XX Push-To-SetTM Series Fittings (patient connection) No Fitting 1/8” NPT Female Tubing Nipple Locking Gland Locking Gland & Overow Safety Trap DISS Male DISS Male & Overow Safety Trap, DISS Wing Nut & Gland Pole Mount Bracket with Locking Gland & Overow Safety Trap Pole Mount Bracket with Tubing Nipple Adapter (vacuum source connection) No Adapter with 1/8” NPT Female 00 Diamond 01 DISS Hand-I-Twist (HIT) 02 DISS Nut & Gland 03 Schrader 04 Chemetron Rectangle Striker (NCG) 05 Puritan Bennett 07 O.E.S. Twist 08 O.E.S. Med Star 09...

 Open the catalog to page 11
Push-To-Set TM Intermittent and Continuous Vacuum Regulators-12

Suction & Oxygen Therapy Devices ♦ Portable Suction Machines Medical Gas Pipeline Equipment ♦ Vacuum & Air Compressor Packages and Accessories Ask about our clinical excellence program Part 1: Clinical Animation Video “Avoiding the hazards of inadvertent administration of high suction pressures” Part 2: Clinical Video Understanding the Hazards of oversuctioning Both videos can be viewed on our website: www.ohiomedical.com Vacuum Regulators - Quick Reference Chart Type of Regulator Model Type Pediatric Neonatal Push-To-SetTM Intermittent Suction Unit (Digital) Intermittent Suction Unit (Analog)...

 Open the catalog to page 12

Archived catalogs

  1. Care-E-Vac 3

    4  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.