Technical Specifications Overview
Application and Patient Type: Designed for emergency, first aid, and patient transport, suitable for both adult and pediatric patients.
Ventilation and Control: Utilizes IPPV Volumetric and PSV with IPPV Cycle Time. Controlled electronically with a gas supply of oxygen or compressed medical air at pressures between 2.7-6.0 bar.
Flow and Ventilation Modes: Maximum flow rate of 50 l/min with constant and decreasing flow modes in PSV. Ventilation modes include AUT-AST, PSV, Apnea Backup, CPAP, and PEEP.
Measurement and Parameters: Measures airways instant pressure with a breath frequency range of 5-70 bpm and an I:E rate of 1:1.5. Inspiratory period is 40% of the breath cycle with automatic inspiratory pause in PSV mode.
Pressure and Volume: Pressure limit ranges from 0-50 cmH2O, PEEP from 0-20 cmH2O, minute volume from 1-16 L/min, and tidal volume from 15-3000 ml.
User Interface and Alarms: Features a digital manometer, LED battery level indicator, external power failure indicator, and control buttons. Alarm types include battery level, low/high airways pressure, apnea, gas supply, power failure, and obstruction warnings with an intensity of 60 dB.
Power Supply and Consumption: Operates on 12VDC or 100-240VAC with a 12VDC adaptor, consuming 0.02A @220VAC or 0.4A @ 12VDC, with a power usage of 4.5W. Built-in battery allows for 10 hours of continuous operation.
Safety and Environmental Conditions: Includes mechanical limits for airways pressure adjustable from 0 to 50 cmH₂O. Operates in temperatures from -18°C to +60°C and stores between -40°C to +70°C, with humidity levels from 10% to 90% non-condensing.
Dimensions and Weight: Measures 240mm in width, 160mm in length, 110mm in depth, and weighs 2.2 kg.
Compliance and Classification: Complies with standards such as EN 60601-1, EN 794-3+A2, EN 62304, and others. Classified as Class II Type BF according to EN60601Dir.
Accessories: Standard accessories include a carrying bag, feeder, supply cable, disposable PVC patient device, oxygen hose, user manuals, and more. Optional accessories include a trolley, oxygen tube, O2 regulator, bacteria filter, and others.
Additional Features: Attempts to open the airway by applying consecutive extra pressure in case of obstruction.