Additional CX PRO compressor nebuliser information regarding Electro Magnetic Compatibility (EMC) according to IEC60601-1 -2:2001, clause 6.8
4Pages

{{requestButtons}}

Catalog excerpts

Additional CX PRO compressor nebuliser information regarding Electro Magnetic Compatibility (EMC) according to IEC60601-1 -2:2001, clause 6.8 - 1

Additional CXPRO compressor nebuliser information regarding Electro Magnetic Compatibility (EMC) according to IEC60601-1-2:2001, clause 6.8 With the increased number of electronic devices such as PC’s and mobile telephones, medical devices in use may be susceptible to electromagnetic interference from other devices. The EMC (Electro Magnetic Compatibility) standard IEC60601-1-2 defines the levels of immunity to these electromagnetic interferences. From the other hand, medical devices must not interfere with other devices. IEC60601-1-2 also defines the maximum levels of emissions for these medical devices. The OMRON CXPRO conforms to this IEC60601-1-2 standard for immunity and emission. Nevertheless, special precautions need to be observed: The OMRON CXPRO needs to be installed and put into service according to the EMC information below. The OMRON CXPRO is intended for use in the electromagnetic environment specified in the tables below. The customer or the user of the CXPRO should assure that it is used in such environment. In general, although The OMRON CXPRO complies to the EMC standards, it can be affected by portable and mobile RF communications equipment (such as mobile telephones). The OMRON CXPRO should not be used adjacent to or stacked with other equipment. In case adjacent or stacked use is necessary, the CXPRO should be observed to verify normal operation. Guidance and manufacturer’s declaration: Electromagnetic Emissions (IEC 60601-1-2) Emissions test Compliance Electromagnetic environment Harmonic emissions The CXPRO is suitable for use in all Class A IEC 61000-3-2 establishments, including domestic establishments and those directly connected to Voltage fluctuations / the public low-voltage power supply network that Complies flicker emissions supplies buildings used for domestic purposes. IEC 61000-3-3 RF emissions The CXPRO is not suitable for interconnection with Complies CISPR 14-1 other equipment.

Open the catalog to page 1
Additional CX PRO compressor nebuliser information regarding Electro Magnetic Compatibility (EMC) according to IEC60601-1 -2:2001, clause 6.8 - 2

Guidance and manufacturer’s declaration: Electromagnetic Immunity (IEC 60601-1-2) IEC 60601 Compliance Electromagnetic environment – Immunity test Test level level guidance Floor should be wood, concrete, or Electrostatic ceramic tile. If floors are covered with ±6 kV contact ±6 kV contact discharge (ESD) ±8 kV air ±8 kV air synthetic material, the relative humidity IEC 61000-4-2 should be at least 30 %. ±2 kV for ±2 kV for power supply Electrical fast power supply Mains power quality should be lines transient/burst lines that of a typical commercial ±1 kV for ±1 kV for and/or hospital...

Open the catalog to page 2
Additional CX PRO compressor nebuliser information regarding Electro Magnetic Compatibility (EMC) according to IEC60601-1 -2:2001, clause 6.8 - 3

Guidance and manufacturer’s declaration: Electromagnetic Immunity (IEC 60601-1-2) IEC 60601 Compliance Electromagnetic environment – Immunity test Test level level guidance Portable and mobile RF communications equipment should be used no closer to any part of the CXPRO including cables, than the recommended separation distance calculated from the equation appropriate to the frequency of the transmitter. Recommend separation distance d = 1.2 √P d = 1.2 √P 80 MHz to 800 MHz 3 V rms d = 2.3 √P 800 MHz to 2.5 GHz Conducted RF 150 kHz ~ 80 where P is the maximum output power MHz IEC 61000-4-6 3...

Open the catalog to page 3
Additional CX PRO compressor nebuliser information regarding Electro Magnetic Compatibility (EMC) according to IEC60601-1 -2:2001, clause 6.8 - 4

Recommended separation distance between portable and mobile RF communications equipment and the OMRON CXPRO. The OMRON CXPRO is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of he CXPRO can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the CXPRO as recommended below, according to the maximum output power of the communications equipment. Output Power of Transmitter in watt Separation distance according to...

Open the catalog to page 4