Introduction
This document provides important information regarding the Electro Magnetic Compatibility (EMC) of the OMRON Battery-operated / AC Adapter-operated Blood Pressure Monitor, in compliance with IEC60601-1-2:2014 standards.
Specifications
The monitor conforms to the EMC standard IEC60601-1-2:2014. It is crucial to use only specified accessories and cables to avoid increased electromagnetic emissions or decreased immunity, which could lead to improper operation.
Precautions
  • Avoid using the monitor adjacent to or stacked with other devices unless necessary, and verify normal operation if such use is unavoidable.
  • Portable RF communications devices should be kept at least 12 inches (30 cm) away from the monitor to prevent performance degradation.
EMC Environment Guidance
Guidance is provided for the EMC environment in which the monitor should be used, including emission limits and compliance details.
Emission Limits and Compliance
The monitor complies with CISPR 11 Group 1, Class B for conducted and radiated RF emissions, and with IEC 61000-3-3 for voltage fluctuations and flicker.
Immunity Test Levels
  • Electrostatic discharge: IEC 61000-4-2, with levels up to ±15 kV air for enclosure port.
  • Radiated RF electromagnetic fields: IEC 61000-4-3, 10 V/m from 80 MHz to 2.7 GHz.
  • Electrical fast transients/bursts: IEC 61000-4-4, ±2 kV for input AC power port.
  • Surges: IEC 61000-4-5, up to ±1 kV for input AC power port.
  • Conducted disturbances: IEC 61000-4-6, 3 Vrms from 150 kHz to 80 MHz.
  • Rated power frequency magnetic fields: IEC 61000-4-8, 30 A/m at 50 Hz and 60 Hz.
  • Voltage dips and interruptions: IEC 61000-4-11, with specific test levels for input AC power port.
Enclosure Port Immunity
Test specifications for immunity to RF wireless communications devices are detailed, including test frequencies, modulation types, and immunity test levels.
Conclusion
The document concludes with a note that EMC tests have included the AC adapter, ensuring comprehensive compliance with the specified standards.
See more

Catalog excerpts

bp8000m-1

Important information regarding Electro Magnetic Compatibility (EMC) This blood pressure monitor manufactured by OMRON HEALTHCARE Co., Ltd. conforms to IEC60601-1-2:2014 Electro Magnetic Compatibility (EMC) standard. Nevertheless, special precautions need to be observed: • The use of accessories and cables other than those specified or provided by OMRON could result in increased electromagnetic emission or decreased electromagnetic immunity of the monitor and result in improper operation. • During measurement, the use of the monitor adjacent to or stacked with other device should be avoided because it could result in improper operation. In case such use is necessary, the monitor and other device should be observed to verify that they are operating normally. • During measurement, portable RF communications device (including peripherals such as antenna cables and external antennas) should be used no closer than 12 inches (30 cm) to any part of the monitor, including cables specified by OMRON. Otherwise, degradation of the performance of the monitor could result. • Refer to further guidance below regarding the EMC environment in which the monitor should be used. Table 1 - EMISSION Limits and Compliance EMC tests have included the AC adapter.