Technical Specifications OverviewOptotek Medical, a member of Lumibird Medical Division, provides technical specifications for their laser systems used in ophthalmic procedures such as capsulotomy, iridotomy, and selective laser trabeculoplasty (SLT). The document outlines the features of the OptoYag&SLT M laser system.
Manufacturer InformationOptotek Medical is based in Slovenia, with headquarters at Lumibird Medical in France. The company is ISO 13485:2016 certified, ensuring compliance with international standards for medical devices.
Laser Specifications- Laser Source: Q-switched Nd:YAG for YAG mode and Frequency doubled Nd:YAG for SLT mode.
- Wavelength: 1064 nm for YAG mode and 532 nm for SLT mode.
- Energy: 0.5 to 10 mJ per pulse for YAG mode and 0.2 to 2.6 mJ per pulse for SLT mode, both continuously variable.
- Pulse Width: 5 ns for both modes.
- Pulse Mode: 1, 2, and 3 pulses per shot for YAG mode; single pulse for SLT mode.
- Spot Size: Less than 10 μm for YAG mode and 400 μm for SLT mode.
- Offset: Adjustable for YAG mode; not applicable for SLT mode.
- Aiming Beam: Red 650 nm with adjustable intensity.
- Repetition Rate: Up to 3 Hz.
- Cooling System: Air cooled.
Slit Lamp Specifications- Magnification: Standard options include 10x, 16x, and 25x, with additional options available.
- Illumination: LED.
System Dimensions and Weight- Size: 73 cm (H) x 60 cm (W) x 48 cm (D).
- Weight: 25.8 kg (56.9 lbs).
Power Requirements- Power Supply: 100 to 240 VAC, 50/60 Hz, 2A MAX.
Optional Accessories- SLT Laser lens, Capsulotomy and Iridotomy laser lenses.
- Motorized lifting column, Footswitch, Five-position magnification changer.
- Beam splitter with co-observation tube.
Other Available Models- OptoYag M: Q-switched Nd:YAG (1064 nm) laser system.
- OptoSLT M: Frequency-Doubled Q-switched Nd:YAG (532 nm) laser system.