Overview: The AI-78 is a gastric-intermittent suction pump designed for digestive drainage. It offers both continuous and intermittent suction with a prefixed vacuum selector set at 90/120 mmHg. The device is equipped with regulatory controls for start/stop cycles and features a quiet, lubrication-free aspiration system, making it ideal for use in intensive care units and hospital rooms.
Technical Specifications:- Chassis: Metallic, fire-painted.
- Engine Pump: Diaphragm type, oil-free.
- Flow Rate: 500 l/h (8.3 l/min).
- Vacuum: Prefixed to 90/120 mmHg (12/16 kPa).
- Programme: Continuous or intermittent.
- Start Cycle: 1 to 120 seconds.
- Stop Cycle: 1 to 90 seconds.
- Collecting Flask: Two 3-liter polycarbonate flasks.
- Voltage: 220 Vac, 50/60 Hz (others available on request).
- Power: 18 VA.
- Dimensions: 24 x 30 x 67 cm.
- Weight: 16 kg.
- Compliance: UNE EN 60601, EN 61000, UNE EN 10079.
Optional Spares:- Bactericidal filter (10/50 units).
- 3L collecting flask (2 units).
- 3L collecting flask with top and connectors.
Additional Features:- Quick connectors and safety valve.
- Control panel with continuous or intermittent selector.
- Bactericide filter included.
Environmental Considerations: The vacuum variation is noted in relation to altitude above sea level at a temperature of 22ºC.
Contact Information: Ordisi, located in Palleja, Barcelona, Spain. Phone: (+34) 93 334 01 12, Fax: (+34) 93 440 25 64, Website: www.ordisi.com, Email:
[email protected],
[email protected].
See more