350/250/250AT Site Specifications
6Pages

{{requestButtons}}

Catalog excerpts

350/250/250AT Site Specifications - 1

^ Ortho-Clinical Diagnostics a company

Open the catalog to page 1
350/250/250AT Site Specifications - 2

Export authorized under general license GTDA (General Technical Data Available) IMPORTANT The information contained herein is based on the experience and knowledge relating to the subject matter gained by Ortho-Clinical Diagnostics, Inc. prior to publication. No patent license is granted by the information. Ortho-Clinical Diagnostics, Inc. reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to the information. The company shall not be liable for any loss or damage, including consequential or special damages resulting from the...

Open the catalog to page 2
350/250/250AT Site Specifications - 3

12 Specifications Environmental Requirements Operating Environment Altitude Sea level to 8000 feet (2400 meters) Temperature and Humidity The environmental limits for normal operation of the analyzer are shown below. 29.4° C (85° F) 25.6° C (78° F) Temperature 108 grains/lb (3.17 grams/ meter) (Grains of moisture per pound of dry air) Operating Characteristics Heat Output 4100 BTUs per hour Noise Output Acoustic Emissions Scale dBC (Europe) VITROS 250/350 Chemistry System - Operator’s Manual

Open the catalog to page 3
350/250/250AT Site Specifications - 4

Requirements at Customer Site VITROS 250/350 Chemistry System - Operator's Manual

Open the catalog to page 4
350/250/250AT Site Specifications - 5

VITROS 250/350 Chemistry System and Printer Specifications Specifications Analyzer Specifications Weight A B C D Installation Specifications for the VITROS 250/350 Chemistry System Reference E F1 G H1 VITROS 250/350 Chemistry System - Operator's Manual With standard printer. Other printers may require additional space. NOTE: The other printer may be located to the left of the analyzer if desired.

Open the catalog to page 5
350/250/250AT Site Specifications - 6

Electrical Requirements Electrical Requirements Receptacle Specifications for the Analyzer Reference Description 61 cm (24 inches) maximum R 120 cm (48 inches) maximum The standard printer plugs into the receptacle on the back of the VITROS 250/350 Chemistry System (lower left corner of the slide processor as you face the back of the analyzer). The printer comes with a double-ended female cord to accomplish this. Electrical Specifications for the Analyzer The VITROS 250/350 Chemistry System will operate on | • 100-120V, 50 or 60 Hz OR • 200-240V, 50 or 60 Hz Your field engineer...

Open the catalog to page 6