Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System
14Pages

{{requestButtons}}

Catalog excerpts

Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 1

Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System Ortho Clinical Diagnostics

Open the catalog to page 1
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 2

PLEASE NOTE The information contained herein is based on the experience and knowledge relating to the subject matter gained by Ortho Clinical Diagnostics (OCD) prior to publication. No patent license is granted by this information. OCD reserves the right to change this information without notice, and makes no warranty, expressed or implied, with respect to this information. OCD shall not be liable for any loss or damage, including consequential or special damages, resulting from any use of this information, even if loss or damage is caused by the negligence or other fault of OCD. This...

Open the catalog to page 2
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 3

Electrical Power Requirements Section 1: Electrical Power Requirements AC Power Quality The VITROS® 5,1 FS Chemistry System (hereafter referred to as the 5,1) and the VITROS® 4600 Chemistry System (hereafter referred to as the 4600) are designed to perform satisfactorily with reasonable disturbances and voltage tolerance to the ac power. The quality and reliability of the power source is often dependent on the geographic location of the site. Some sites, where power quality is exceptionally poor, may require special power conditioning equipment to ensure that the tolerances are maintained...

Open the catalog to page 3
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 4

CEE7 “Schuko” plug may be supplied, either straight or right angle.

Open the catalog to page 4
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 5

Electrical Specifications for the Supplied PRINTER Important The following information is only applicable to PRINTERS supplied by OCD. If the PRINTER is being supplied by the customer, refer to the manufacturer’s PRINTER manual for the electrical specifications.

Open the catalog to page 5
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 7

RECEPTACLE Location Specifications Specifications f°r pRINTER pOWER and Specifications for the System only

Open the catalog to page 7
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 8

Acoustic Noise Level Maximum altitude: 2438 m (8000 ft.) This product has been evaluated and is suitable only for Pollution Degree II and Insulation Category II. Additional Requirements for the VITROS® 5,1 and VITROS® 4600 HEAT REJECTION PLENUM If the customer chooses to purchase and use the optional Heat Rejection Plenum, the EXHAUST SYSTEM and DUCT must be installed by the customer. Specifications Air Capacity 7.1 - 9.9 cubic metres (250 - 350 cubic ft.) per minute 259 - 366 metres (850 - 1200 ft.) per minute DUCT 20.5 cm (8 in.) flexible AIR DUCT and attaching hardware Air Flow -...

Open the catalog to page 8
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 9

Networking Specifications Section 7: Networking Specifications The requirements for e-Connectivity network connection are located in the e-Connectivity® Network Connection Specifications and Network Form located on-line at: http://apps.orthoclinical.com/notes/contentpages/techdoc.aspx?view=USSPECS The customer must complete the form and return it to Ortho-Clinical Diagnostics. Important All requirements for lab network connections must be completed before installation of the analyzer(s).

Open the catalog to page 9
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 11

Shipping Specifications Shipping Specifications * Minimum overhead clearance required for unpacking.

Open the catalog to page 11
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 12

Supplied PRINTER Specifications **** Port USB Version (TBD) Printer Language Post Script Installation Specifications for the System only Installation Specifications for the Supplied PRINTER **** **** If the PRINTER is being supplied by the customer, refer to the manufacturer’s PRINTER manual for any specifications.

Open the catalog to page 12
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 13

Specifications for the 5,1 and 4600 HEAT REJECTION PLENUM

Open the catalog to page 13
Site Specifications for the VITROS® 5,1 FS Chemistry System and VITROS® 4600 Chemistry System - 14

Ortho-Clinical Diagnostics Felindre Meadows Pencoed Bridgend CF35 5PZ United Kingdom Ortho-Clinical Diagnostics 100 Indigo Creek Drive Rochester, NY 14626 USA

Open the catalog to page 14