Installation and Site Specification VITROS® ECiQ/ECi Immunodiagnostic Systems Ortho Clinical Diagnostics PART OF THE ^m*OVl«*^O^m*OH FAMILY OF COMPANIES

Open the catalog to page 1
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 2

Export authorized under general license GTDA (General Technical Data Available) IMPORTANT The information contained herein is based on the experience and knowledge relating to the subject matter gained by Ortho-Clinical Diagnostics, Inc. prior to publication. No patent license is granted by the information. Ortho-Clinical Diagnostics, Inc. reserves the right to change this information without notice, and makes no warranty, express or implied, with respect to the information. The company shall not be liable for any loss or damage, including consequential or special damages resulting from the...

Open the catalog to page 2
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 3

Installation and Site Specifications Installation and Site Specifications Requirements Although equipment service representatives uncrate and install the VITROS ECi/ECiQ Immunodiagnostic System at the laboratory site, the site needs to be prepared according to specifications. This section describes general requirements for installing the ECi/ECiQ System at your laboratory, including physical and environmental requirements. Physical Dimensions The following sections provide general component dimensions and a site drawing that illustrates setup and space requirements. Dimensions Table 1...

Open the catalog to page 3
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 4

Installation and Site Specifications Site Drawing The site drawing in figure 1 provides the physical dimensions of the system. The system weighs approximately 800 pounds (366 kilograms). Figure 1. Physical Dimensions The system and printer dimensions are described below: Reference Dimension System Dimensions: G 59.7 cm (23.5 in.) - Left side door clearance Installation and Site Specification VITROS ECi/ECiQ Immunodiagnostic System Version 4.0 from Operator's Guide

Open the catalog to page 4
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 5

Installation and Site Specifications Reference Dimension Printer Stand for Epson LQ/Model 570e and Epson LX/Model 300* Printer Dimensions: Printer Stand with Texas Instruments Omni 800/Model 830e Printer Dimensions: * Standard printer for the system. Printers are ordered in accordance with regional specifications. Safety Requirements Power Requirements The system meets all requirements for bearing the CE marking and safety The system requires a dedicated, single phase, AC power line and one of Line Voltage Frequency Current Draw Input Power North America 120 volt AC 50-60 Hz 12 amps <1.5KVA...

Open the catalog to page 5
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 6

Installation and Site Specifications Printer Line Voltage and Frequency for printers used in North America: Epson LX/Model 300* Texas Instruments Omni 800/Model 830e Line Voltage: 99-132 volt AC Line Voltage: 99-132 volt AC Line Voltage: 120 volt AC *Standard printer for the system. Printers are ordered in accordance with regional specifications. Electrical Outlet (Receptacle Type) The electrical outlet (receptacle type) for the system is as follows: • North America: NEMA 5-20R • Continental Europe: CEE7 “Shuko” Electrical Outlet (Receptacle type) for Printers used in North America: Epson...

Open the catalog to page 6
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 7

Installation and Site Specifications e-Connectivity® Network Connection The following sections provide network connection specifications for e-Connectivity® Interactive System Management and a Site Evaluation Form at the end of this Chapter. Refer to “Power Requirements” on page 3 for additional information. Refer to Chapter 2, "System Overview" for detailed information on e-Connectivity®. Network Connection The following are required for e-Connectivity network connection: IMPORTANT: An auxiliary means of establishing a connection must not be used for the network connection. For example, a...

Open the catalog to page 7
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 8

Installation and Site Specifications • Support IPSec pass through to the Internet I.P. Address of 148.177.0.108. Note: IPSec utilizes port 500 outbound and inbound, port 4500 outbound and inbound. These ports must be open in the local area network’s firewall and allow the specified protocol. • Female RJ45 connector on the network port within 20 feet of the center of the system. Site Evaluation Form Use a copy of the Site Evaluation Form located at the end of this chapter to document the network connection information. Installation and Site Specification VITROS ECi/ECiQ Immunodiagnostic...

Open the catalog to page 8
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 9

Installation and Site Specifications Uninterruptible Power Supply If you experience frequent power fluctuations, you may want to use an uninterruptible power supply. Contact your technical support representative for more information. Entrance to Room The entrance door opening should be a minimum of 76.2 cm (30 in.) wide. Telephone and Broadband Internet Recommendations You may also find it convenient to have a telephone near the system to communicate with technical support personnel during troubleshooting sessions. A broadband Internet connection is required for e-Connectivity® which...

Open the catalog to page 9
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 10

Installation and Site Specifications Environmental Specifications The environmental limits for normal operation of the system are defined below and shown in figure 2. • BTU Output: 4100 BTUs per hour • Operating Temperature: 15–30°C (59–86°F) • Site Relative Humidity: 15–75% RH noncondensing • Altitude: -0.1524/2.439 km (-500/8000 ft.) 30.0°C (86°F) 26.0°C (78°F) Temperature Relative Humidity Figure 2. Environmental Requirements System Relocation The system is mounted on casters to facilitate its relocation within the laboratory. The new site must meet the same space, electrical, and...

Open the catalog to page 10
VITROS® ECi/ECiQ Immunodiagnostic System Site Specifications - 12

Ortho Clinical Diagnostics PART OF THE ^O^mOH<^ofm40H FAMILY OF COMPANIES Ortho-Clinical Diagnostics Johnson & Johnson Fr I BPPI 50-100 Holmers Farm Way High Wycombe United Kingdom 100 Indigo Creek Drive

Open the catalog to page 12

All Ortho Clinical Diagnostics catalogs and technical brochures

  1. ORTHO ProVue

    6 Pages

  2. VITROS® 350

    2 Pages

  3. Pathology

    4 Pages

Archived catalogs