Catalog excerpts

Finetest AC - 1

Anyang-si, Gyeonggi-do, Korea (431 V, U>) IECIREPI ObelbSJt. Brussels, Belgium IMl Lancing Device :GMMC Digital-ro, Geumcheon-gu Seoul, Korea IECIREPI Global Medical Market Corporation S.L. I'.iu lndiMii.il Gutat de Carlet Medial 20,46240 Carlet, Valenda, Spain E-mail: aricco@gmmc-inccom Full Automatic Coding Test Strip Ejector Operation Manual

Open the catalog to page 1
Finetest AC - 2

Auto-coding Premium

Open the catalog to page 2
Finetest AC - 3

Dear Finetest Auto-coding Premium System Owner TM Thank you for choosing the Finetest Auto-codingTM Premium Blood Glucose Monitoring System. This booklet has important information you must know about the Finetest Auto-codingTM Premium System. Please read it carefully. We understand that self-testing of blood glucose level provides a way to control your diabetes and may give you peace of mind by testing regularly. As a result Finetest Auto-codingTM Premium was developed to provide you with a fast and accurate reading with a convenient and simple process. Our goal at Finetest Auto-codingTM...

Open the catalog to page 3
Finetest AC - 4

Finetest Auto-coding* Premium Blood Glucose Monitoring System KIT Contents ■ ■ -9 infopia co.,ltd. | Intended use of Finetest Auto-coding™ Premium Blood Glucose Test Meter The Finetest Auto-coding™ Premium Blood Glucose Monitoring System is ntended for in vitro diagnostic use only. Finetest Auto-coding™ Premium Blood Glucose Monitoring System is for self-testing. It should not be used for the diagnosis of diabetes or for the testing of The Finetest Auto-coding™ Premium was designed to test a blood glucose (sugar) level and can be used with a capillary whole blood sample. Do not use the...

Open the catalog to page 4
Finetest AC - 5

Reminders Testing during or after meal, physical exercise, or any other event that will impact glucose levels can affect your blood glucose result. Finetest Auto-codingTM Premium Blood Glucose Monitoring The Finetest Auto-codingTM Premium Blood Glucose Monitoring System is intended for use outside the body (in vitro diagnostic use only). The Finetest Auto-codingTM Premium Blood Glucose Monitoring System should be used only for testing glucose and only with fresh capillary whole blood samples. The Finetest Auto-codingTM Premium Blood Glucose Monitoring System should not be used for the...

Open the catalog to page 5
Finetest AC - 6

I Finetest Auto-coding™Premium BlOOd GlUCOSe Meter Test Strip Port: Insert the Finetest Auto-coding Premium Test Strip here. Test Strip Ejector Display Screen Display your test results: Function Button Powers the meter on/off, prompts memory data and acts as an enter button during setup and after tests. Up/Down Arrow Buttons: Adjusts setting, scrolls through memory. -Serial Number - Battery Cover Do not use Element™ Auto-coding blood glucose test meter in a dry environment, especially if synthetic materials are present. Synthetic clothes, carpets, etc., may cause damag'ng static discharges...

Open the catalog to page 6
Finetest AC - 7

Finetest Auto-codingTM Premium Blood Glucose Test Strip Use the Finetest Auto-codingTM Premium Test Strip with the Finetest AutocodingTM Meter and FinetestTM Meter. Apply Blood Target Confirmation Window Storage and Handling Caution 1. Keep out of direct sunlight. 2. Store the Finetest Auto-codingTM Premium Blood Glucose Test Strip vials in a cool, dry place between 2-30°C(36-86°F). Do not freeze. 3. Make a notation of the discard date on the vial label when you first open it. Discard remaining ElementTM Auto-coding blood glucose test strips 3 months after first opening the vial. 4. Close...

Open the catalog to page 7
Finetest AC - 8

Setting your Blood Glucose meter Setting your Blood Glucose meter The Blood Glucose Monitoring System has a wide variety of functions to choose from. From the setup mode you have the ability to turn the activity / meal flags on or off, and you can also set the date/time, designate three unique averages and set up to five daily alarms. Temperature unit Press and release either or buttons to set or . Enter button. Power On Press for 3 seconds at least. User Activity Option After pressing and releasing either or buttons to turn the user Activity option on/off, confirm with Number of days for...

Open the catalog to page 8
Finetest AC - 9

Auto-coding Function Checking the System with Glucose Control Solution How to do the Finetest Auto-codingTM Premium glucose control solution test. Step 1-1 : (refer to page 16) 1) Finetest Auto-codingTM Premium has automatic code recognition function. To check the meter and test strip with glucose control solution, when the symbol, , is displayed, press or button. Then, the symbol, , will be displayed. As the glucose control solution symbol appears, apply the glucose control solution to the test strip. 2) It recognizes automatically the code number. This is a very convenient function...

Open the catalog to page 9
Finetest AC - 10

Performing a Control Solution Test The control solution should be used when there is suspicion that the meter or the strip are not working properly. For example : if strips have been exposed to extreme conditons or if you drop the meter. Performing a Control Solution Test For example If you use a normal control solution, you must use the normal range printed on the side of the strip vial. Check the expiration date before performing a control solution test. Do not used if expired. Once opened, solution expires after 3 months. Note : (In order to perform the more accurate glucose control...

Open the catalog to page 10
Finetest AC - 11

Testing your Blood - Preparation Before testing, be sure to read this section and test strip package insert found in the test strip box carefully. Make sure you have all the necessary items to begin the test : 1) Finetest Auto-codingTM Premium blood glucose test meter. 2) Finetest Auto-codingTM Premium blood glucose test strip. 3) Lancing Device. 4) Sterile Lancet. Caution To reduce risk of infection : Collecting a Drop of Blood The lancing device is a convenient medical tool for collecting capillary blood samples for glucose monitoring or other tests which require one or two drops of...

Open the catalog to page 11
Finetest AC - 12

Collecting a Drop of Blood Collecting a Drop of Blood Disposing of your lancet 1. Unscrew the tip of the lancing device by turning it counter-clockwise while holding the base firmly. 2. With two fingers, pull out the lancet carrier and hold it in place. 3. While holding the lancet carrier, insert a new sterile lancet into the bottom of lancet carrier. 1. Push the needle into the protective cover. 2. Pull the lancet out and discard accordingly. Recommended Available Lancing Device infopia®, Element™ sterile Lancet is used with a Lanzo Lancing device. 12345 5. Put the tip of the lancing...