Pacific Image RapidScan One Reader (Non-sterile)
Open the catalog to page 1Product Name Pacific Image RapidScan One Reader (Non-sterile) Use the instrument to objectively read the C or T line of the rapid test. Pacific Image RapidScan One Reader (Non-sterile) uses the principle of reflectance to analyze the C & T line displayed on the rapid test inserted into the instrument, and then determine whether the test item result is positive or negative. Side Effects or Complications 1. This product is not waterproof. Do not get wet. 2. If the product is dropped severely, it will damage the electronic parts. Please use it with care. 2
Open the catalog to page 2Stand-alone operation — Indicator light………..…….…………………………………………..P6 Mobile device connection operation — Connect teaching……………………………..……………………...P8 — Home screen………….…………………..………………………..….P9 System page ………………………………………………………………P11 — User & User-Add……………………..…………………………….P12 — User Management..……………………………………………….P13 — Test Profile……
Open the catalog to page 3Product Introduction
Open the catalog to page 4Appearance Introduction Indicator light Analysis button, switch button (Short press to turn on, long press to turn off) For Ethernet cable and USB port Rapid test insertion port Power supply insertion hole Heat-transfer passage
Open the catalog to page 5Non-exclusive match • We need to communicate and cooperate with the rapid test manufacturer about the reader in advance, and the manufacturer will define the criteria of the image value after the experiment, and then set the cut-off value and the judgment standard of negative and positive. • Therefore, it can be interpreted as long as the above content is set. There is no specific rapid test for this reader. • The rapid test manufacturer which is cooperating with us now for example is like 瑞 凌庚 (幽
Open the catalog to page 6Instructions • There are two ways to use: stand-alone operation & mobile device connection operation (more detailed instructions will be given after page 9.) • For stand-alone operation, you can easily use the button to analyze the reading value, and use the indicator light on the surface to determine whether the sample is positive or negative. • The mobile device connection operation is to transmit a WiFi signal through the reader and connect the mobile device, and then you can log in to the internal screen of the machine through a browser for more advanced operations.
Open the catalog to page 7Software description & working principle • It is necessary for the rapid test manufacturer to use software to set the detection range of the C line and T line, and conduct experiments through the cooperation of both parties; after the experiment, the results are converted into data settings in the reader, and then the threshold for judging negative and positive is defined. • The software itself has the function of analyzing and processing images, and the interpretation time is about 7 to 15 seconds; when the user presses the analysis button, the reader will shoot the image of the rapid...
Open the catalog to page 8Stand-alone operation
Open the catalog to page 9Indicator light Orange light: positive (+) Green light: negative (-) Flashing orange light: invalid result Flashing red light: UV-C disinfection No light: UV-C off Turn on for 3 seconds when the analysis is complete, and turn on for 3 seconds when the rapid test is taken out. Tap once to start the analysis. When it is turned on, the orange light flashes. When it is standing by, the green light shows. When the light flashes green, analysis is in progress. Put in the rapid test. (There is a sensor inside. If the rapid test cannot be detected, the device will not display any result lights.)
Open the catalog to page 10Mobile device connection operation
Open the catalog to page 11Connect Reader to mobile device via Wi-Fi Turn on the Wi-Fi of your mobile device and find "RS_One_xxxxxxx", which is the SSID of RapidScan One. (No password required when connecting to Wi-Fi) If you encounter any difficulties, please contact your provider. Connect to the Web UI through a browser Open the browser (Chrome, IE or Safari), and enter 192.168.0.1 in the URL bar. (This method is called AP Mode, see page 15 for details.) Your browser will connect to the login page. Please enter: User: Admin Password: 0000 (Other users can also be added in the future) Press login to enter the...
Open the catalog to page 12Home Page System For system settings Test Profile Select product information from this section. Information The inspector (the account name used when logging in and cannot be changed) and sample No. (can be changed) Run Test Press the button to get instant results. Set Delay Use the button to delay the image capture time. (See the dialog box at the bottom left.) After using the set delay, the dialog on the left will be displayed; at first, the preset minutes of the file will be displayed, but you can still set 1 to 100 minutes according to your needs. 13
Open the catalog to page 13Home Page Time Date/time of the most recent test Click the picture to enlarge the image. Test Result 1. C line value 2. T line value (T2, T3 line... etc.) 3. Result: value or statement (positive or negative) 4. Image captured by the reader 14
Open the catalog to page 14System Click the gear wheel User Manage user accounts Test Profile Add or delete product information Database Test results are sorted by date/time Settings Management system settings Logout Log out of current account This is the serial number of the software and firmware version.