Product catalog summary
Benefits
  • Cost Effective: Provides economical monitoring of fixed points at timed intervals, covering larger areas than fixed particle counters with low cost per sample location.
  • Flexible: Allows easy programming of sampling routines and automatic recipe changes upon alarms or external signals, with a large coverage area of up to 125 ft per sample point.
  • Rapid and Accurate Measurements: Features patented flow control to compensate for pressure drops, maintaining constant flow to minimize particle settling and using isokinetic sampling probes to prevent distortion in particle size relationships.
Features
  • Supports multiplexed monitoring of up to 32 locations with a single particle counter.
  • Controlled by Facility Net or Pharmaceutical Net software with programmable recipes and sequences.
  • Includes alarms for low vacuum and automatic pump shut-off for overheating.
Applications
  • Facility monitoring, large minienvironment monitoring, cleanroom verification, and trend analysis.
Compatibility
  • Compatible with Lasair® II, Lasair III particle counters, or IsoAir® PLUS particle sensor.
Specifications
  • Manifold System: AM II-16 supports 10-16 sample ports, AM II-32 supports 22-32 sample ports, suitable for ISO 8 or cleaner environments.
  • Manifold Pump: Requires one pump, with flow rates over 100 ft³/min for AM II-16 and over 150 ft³/min for AM II-32, with automatic shut-off for safety.
  • Power: Various voltage options available, with specific amperage requirements for single and three-phase systems.
  • Manifold Module: Nominal flow rate of 1.5-3 CFM per port, with recommended purge time of 15 seconds and minimal cross talk.
  • Control Box: Uses Ethernet for communication, with RS-232 and Telnet setup modes, and includes automatic alarms for inadequate vacuum.
System Components
  • Includes manifold module, control box, pump, and necessary accessories, excluding particle counter/sensor and software.
Contact Information
  • Headquarters located in Boulder, Colorado, with global offices in various countries for support and service.
See more

Catalog excerpts

AM ll 16/32 Aerosol Manifold for Sequential Monitoring-1

BENEFITS Cost Effective • Most economical monitoring of fixed points at timed intervals • Low cost per sample location • Monitors a much larger area than a fixed particle counter • Ethernet communications to facility monitoring system Flexible • Easily programs a sampling routine, for either simple or complex sequences • Automatically changes sampling recipe upon receiving alarm or external signal • Large coverage area: each sample point can be up to 125 ft away Rapid and Accurate Measurements • Patented flow control ensures the manifold properly Manifold Pump Manifold Module The AM II manifold offers advanced features for sequentially monitoring aerosol contamination at multiple locations with a fixed particle counter. The AM II is ideal for trend analysis and cleanroom verification when continuous monitoring is not needed. compensates for manifold pressure drop, eliminating systematic particle undercounting of up to 15% • Maintains constant flow (nominal 3 CFM) at each sample location to minimize gravitational particle settling • With isokinetic sampling probes, captures particles at same velocity as sample air, preventing distortion in particle size relationships F E AT U R E S • Multiplexed monitoring of up to 32 locations with a CO M PAT I B I L I TY • Lasair® II or Lasair III particle counters, or IsoAir® PLUS particle sensor single particle counter Controlled by Facility Net or Pharmaceutical Net software Programmable recipes and sample point sequences Assembly designed to eliminate contamination from valves Alarm for low vacuum; automatic shut-off for pump overheating A P P L I C AT I O N S • Facility monitoring • Monitoring large minienvironments • Cleanroom verification and trend analysis Without measure

 Open the catalog to page 1
AM ll 16/32 Aerosol Manifold for Sequential Monitoring-2

Manifold System Sample ports Max. 16, Min. 10 Max. 32, Min. 22 Use in locations ISO 8 (Class 100,000) or cleaner 0.1 to 0.3 µm particle counters: Use to simultaneously monitor ISO 3, 4, 5, and 6 (Classes 1, 10, 100, 1000) 0.5 to 5.0 µm particle counters: Use to simultaneously monitor ISO 5, 6, 7, and 8 (Classes 100, 1000, 10,000, 100,000) Operating range Manifold Pump Pumps required Flow rate Safety provision Power Dimensions (l, w, h); weight 1 (single-phase) > 100 ft3/min Automatic pump shutoff 105 – 125 VAC, 60 Hz or 220-240 VAC, 50 Hz 10.9 Amp, single-phase Manifold Module Flow rate Purge...