Overview: The document describes the H2O2 decontamination technology used in IncuSafe Incubators, specifically models MCO-170AICUVH, MCO-230AICUVH, and MCO-170MUVH. This technology offers a rapid and safe method for decontaminating incubator chambers using vaporized hydrogen peroxide (H2O2).
Decontamination Process:
  • Preparation Time (10-15 minutes): Remove and wipe down interior components, reposition them for in situ decontamination, and set up the optional H2O2 generator.
  • Decontamination Time (Approx. 135 minutes): Activate the H2O2 cycle, warming the chamber to 45°C. H2O2 vapor is generated and circulated, with a UV lamp reducing it to water and oxygen.
  • Finish Time (Approx. 10 minutes): Open the chamber, wipe off remaining liquid, and reposition components to their normal positions.
Mechanism of Action: The process involves converting aqueous H2O2 to vapor using ultrasonics, which circulates throughout the chamber. The vapor breaks down into hydroxyl radicals, initiating a chain reaction that damages contaminants, ultimately resolving into harmless water and oxygen.
Key Benefits: The technology achieves at least a 6 log reduction of major contaminants in under three hours, making it efficient and effective for maintaining sterile conditions in biomedical settings.
Additional Information: For further details, visit the Panasonic Healthcare website.
See more

Catalog excerpts

IncuSafe Incubators H2O2 Decontamination Technology-1

IncuSafe CO2 & Multigas Incubators H2O2 DECONTAMINATION TECHNOLOGY When initiated, the unique high-speed decontamination system uses vaporized H2O2, offering time-saving and documented chamber decontamination with complete safety. DECONTAMINATION CYCLE H2O2 Decontamination How does it work? 1. Hydrogen peroxide (aqueous) is converted to vapour using high frequency ultrasonics. During this process, the fan motor remains active, ensuring H2O2 vapour accesses every point of the chamber and the tubing to and from, and the inside of the CO2 sensor. 2. The H2O2 vapour breaks down into hydroxyl radicals naturally. 3. The hydroxl radicals initiate a chain reaction of electron stealing. STEP 1 Preparation Time: 10 - 15 minutes CONTAMINANT DAMAGE CONTAMINANT CELL 1. Remove all interior components 2. Wipe down the inside of the incubator 3. Reposition interior components to specified locations for in situ decontamination 4. Set up the H2O2 generator (MCO-HP)* *Optional Accessory. H2O2 reagent is required for this process. Decontamination Time: Approx. 135 minutes 1. Press the H2O2 button. The chamber will warm up to 45°C for optimum results 3. H2O2 avapour generation starts 4. Interior fan circulates vapor 5. UV lamp reduces H2O2 to water and oxygen STEP 3 Finish Time: Approx. 10 minutes 2•OH H2O2 separates in two hydroxyl radicals Remaining H2O2 and hydroxyl radicals 4. This unstable internal environment leads to death of contaminants. Remaining hydroxyl radicals and H2O2 are resolved to H2O (aqueous) & O2 (gas). The H2O2 decontamination within IncuSafe Incubators achieves at least a 6 log reduction of major contaminants. The full decontamination process takes less than three hours. High Heat Decontamination •OH + OHH2O2 can obtain an electron from an organic source •OH initiates chain reaction of electron stealing 1. Open chamber door 2. Wipe off remaining liquid with sterile cloth 3. Reposition interior components to normal positions DNA is very susceptible to oxidative damage. Since most bacteria have a single chromosome controlling all their life functions, this kind of effect can be detrimental to their normal function. Prokaryotic organisms often lack repair mechanisms to limit such damage, making them more prone to change. For more information, please visit our website:

 Open the catalog to page 1

All PHC Europe catalogs and technical brochures

  1. ULT Brochure

    52  Pages

  2. VIP Brochure

    16  Pages

  3. MPR-1014R-PE

    2  Pages

  4. MPR-N450FH-PE

    2  Pages

Archived catalogs

  1. MLS

    2  Pages

  2. MIR-162/262

    2  Pages

  3. MIR-H163-H263

    2  Pages

  4. MDF-U731M-PE

    2  Pages

  5. MDF-U5412-PE

    2  Pages

  6. MDF-U3386S-PE

    2  Pages

  7. MDF-U4186S-PE

    2  Pages

  8. MDF-U5386S-PE

    2  Pages

  9. MDF-U700VX-PE

    2  Pages

  10. MDF-U500VX-PE

    2  Pages

*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.