Product Overview:
The pITA Rapid Exchange Neurovascular PTA Balloon Catheter is designed for intracranial stenosis treatment. It features a long soft distal tip with a thin profile of 0.016 inches and a low balloon crossing profile ranging from 0.023 to 0.026 inches. The catheter is compatible with a 0.014-inch guide wire and includes 'zero profile' x-ray markers for enhanced visibility.

Specifications:
  • Total length: 160 cm
  • Balloon diameter sizes: 1.25 mm to 4.0 mm
  • Balloon lengths: 10 mm to 40 mm
  • Nominal pressure: 6 ATM for most sizes, 10 ATM for 1.25 mm and 1.5 mm diameters
  • Rated burst pressure: 14 ATM for most sizes, 18 ATM for 1.25 mm and 1.5 mm diameters
  • Compliance: Non-compliant for 1.25 mm and 1.5 mm, semi-compliant for others

Design Features:
  • Distal shaft diameter: 2.5F / 0.83 mm
  • Proximal shaft diameter: 1.9F / 0.63 mm
  • Embedded x-ray markers for precise placement
  • Central x-ray marker for 1.25 mm and 1.5 mm diameters

Performance Parameters:
The catheter is designed to withstand nominal pressures of 6 ATM and rated burst pressures of 14 ATM for most sizes, with higher pressures for smaller diameters (1.25 mm and 1.5 mm). This ensures reliability and safety during procedures.

Conclusion:
The pITA Rapid Exchange Neurovascular PTA Balloon Catheter offers a range of sizes and pressures suitable for various clinical needs, with specific features enhancing its usability in neurovascular applications.
See more

Catalog excerpts

pITA - Rapid Exchange Neurovascular PTA Balloon Catheter-1

pITA Rapid Exchange Neurovascular PTA Balloon Catheter rapid exchange PTA balloon for intracranial stenosis • Long soft distal tip with 0.016” thin profile • Low balloon crossing profile: 0.023” - 0.026” 0.016” entry profile embedded x-ray markers • Compatible with 0.014” guide wire • “Zero profile” x-ray markers • Total length: 160 cm • Diameter sizes from 1.25 to 4.0 mm Balloon lengths from 10 to 40 mm • Nominal pressure at 6 ATM* Rated burst pressure at 14 ATM* • Non-/semi-compliant** *For pITA with diameter of 1.25 mm and 1.5 mm: Nominal pressure at 10 ATM and rated burst pressure at 18 ATM **For pITA with diameter of 1.25 mm and 1.5 mm: Non-compliant; For all others: Semi-compliant

 Open the catalog to page 1
pITA - Rapid Exchange Neurovascular PTA Balloon Catheter-2

pITA Rapid Exchange Neurovascular PTA Balloon Catheter Catheter Length: 160 cm Distal Marker* Proximal Marker* A Proximal Shaft Diameter: 1.9F / 0.63mm Distal Shaft Diameter: 2.5F / 0.83mm * The pITA with diameter of 1.25 mm x-ray 1.5 mm *The diameter 1.25 mm/1.5 mm balloon catheters have one centrally located and marker only. have only one centrally located x-ray marker only. Rated Burst Pressure [atm] Rated Burst Pressure [atm] phenox GmbH | Lise-Meitner-Allee 31 | D-44801 Bochum, Germany | Tel: +49 (0) 234 - 36 919 - 0 | Fax: +49 (0) 234 - 36 919 - 19 www.phenox.net | [email protected]

 Open the catalog to page 2
*Prices are pre-tax. They exclude delivery charges and customs duties and do not include additional charges for installation or activation options. Prices are indicative only and may vary by country, with changes to the cost of raw materials and exchange rates.